Synopsis
Magnetic resonance techniques are considered to be
not harmful. The three electromagnetic fields used for MR - static magnetic
field, switched gradient fields, and radio frequency field - interact with
human tissue, but also with other materials exposed to these fields. The physical interactions with human tissue do not cause
irreversible physiological effects, as long as certain limits are not exceeded.
Concerning foreign material (e.g. implants), the
physical effects of the applied fields may cause severe hazards for patients,
staff, and material, if MR examinations are not performed properly.
Objective
Information
about the potential dangers
·
of the static magnetic field B0
(physiological effects: interactions with cells and tissues; physical
effects: interactions with ferromagnetic objects: force and torque;
requirements for MR safe implants and devices),
·
of the low frequency (= audio
frequency) switched gradient fields (physiological effects: induction of current
pulses in nerves, peripheral nerve stimulation, possible cardiac nerve stimulation;
noise; physical effects: induction, effects on implants; requirements for MR safe implants and devices),
·
of the radio frequency field B1
(physiological effects: warming (RF absorption, SAR), burns
caused by current loops with skin-skin contact; physical effects: interaction
with metals and conducting material: induction, heating, sparking, burns at
implant-tissue contact surfaces; requirements for MR safe implants and devices),
·
of the cryo system (loss of
cooling/quench, handling of cryogenic agents).Risks associated with the static field B0
Up to now there
is no scientific evidence that static magnetic fields (B0) within
the range typically used for MR produce permanent bioeffects that could lead to
health problems [1]. Transient effects causing slight indispositions of persons
are possible, usually related to movements in inhomogeneous parts of the field.
However, in spite of the low interaction with body tissue, the static field B0
causes the hazard of most concern: As B0 is commonly produced
by a superconducting coil, it is always switched on. Ferromagnetic objects may
be accelerated towards the magnet. The translational force depends on the spatial gradient of B0 and thus is largest near the entrance of the MR tunnel. The torque acting to align the long axis of a ferromagnetic object with the field lines is largest at the isocenter of the magnet. Persons lying in the scanner or standing
near the bore opening may be injured. Ferromagnetic implants may be dislocated
and damage surrounding tissue. Fatal outcomes of these interactions have been
reported [2,3]. While B0 is strongest in the scanner bore, the field
extends with significant strength several meters around the scanner. A field
strength of 0.5 mT defines the border of the 'controlled access area', which
must be blocked to the general public [4], as impairment of active implants,
e.g. pacemakers, cannot be excluded even in the fringe field of an MR scanner.Risks associated with rapidly switched magnetic fields (gradient fields)
The switched field gradients (Gx, Gy, Gz), necessary to
provide spatial information, are active during scanning only. A set of three small magnetic fields, linearly increasing in three orthogonal directions, is generated, which modifies the static field. Switching
frequencies are in the order of some 100 Hz to several kHz, i.e. in the range of
audible frequencies.
Concerning safety, two effects are of relevance. The first is peripheral nerve
stimulation. Its occurrence depends on gradient steepness and switching time.
The exact function depends on the model applied [5], and people are differently
susceptible to stimulation [6]. Peripheral nerve stimulation is not by itself
dangerous, but it is taken as last noticeable limit before the possible
generation of stimulation in vital nerves, e.g. cardiac nerves, which must be
avoided at any case.
The second effect is noise production. Noise levels of 99 dB(A) may be reached,
sometimes even more, and hearing damage is possible [7]. If implants are
present, gradient switching may induce current pulses, which interfere with the
function of the implant electronics.Risks associated with the pulsed radiofrequency field B1
The radio frequency field B1 has a significant power
only inside or adjacent to the excitation coil. In most cases the body coil is
used for excitation, so that the RF field stretches over a significant portion
of the body. The main concern is heating due to eddy currents, which can be
rather high especially in the presence of metallic implants. While overall warming of the
body is limited to acceptable levels by limiting RF absorption (the RF Specific
Absorption Rate (SAR) must be below 4 W/kg body weight), heat release at
skin-skin contacts in loops formed by arms or legs may cause severe burns at
the contact point. Even second or third degree burns have been reported [8].
In metallic implants the current is higher than in surrounding tissue. At crossover
points of the current into or out of the implant the local current density in
the tissue may be so high that burns are possible. Similar effects may also occur
in wires outside the tissue, but inside the excitation coil. Especially at bad
connections sparking may occur, which in the extreme case may ignite
inflammable material [8]. Sparks can also be generated from carbon rods, as
used for external fixation (Fig.1). The danger of heating hazards is commonly underestimated. Most MR accidents reported in the FDA collection of reports on adverse events (the Manufacturer And User facility Device Experience, MAUDE [10]) refer to burns [11].Risks associated with the cryogenic system
In addition to the electromagnetic fields cryogens
- usually liquid Helium (LHe) - used in superconducting magnets must be
considered. Cryogens pose a risk only in case of a quench, which in most sites never
happens. During an accidental or deliberate quench of the magnetic field
superconductivity is disrupted and the current crosses over to a conducting
copper matrix, where the energy of the current is converted to heat. This
causes the liquid helium to immediately evaporate, which in the end causes a
700-fold volume increase compared to liquid He. This requires the venting off
of typically 700 m3 gaseous He within a couple of seconds. Usually
quench lines are designed to handle this amount of gas. However, imperfectly
maintained quench lines may be blocked. In this case the gas will evaporate
into the scanner room ('in-room quench'), which in most cases is far smaller
than 700 m3, creating a severe overpressure. This has happened a
couple of times, and severe damage to buildings is reported. Therefore, careful
maintenance of the cryo system and the quench lines is mandatory to prevent
the danger of an in-room quench. Safety limits
To cope with MR hazards, limits of fields and field
changes have been determined in international standards, especially for the
specific absorption rate (SAR) to prevent tissue injury due to excessive warming,
and for slew rate and slope of switched gradients to prevent nerve stimulation.
While these limits provide adequate safety against unwanted physiological
reactions, they do not describe adequate limitations for interactions of these
fields with implants, especially active implants. To be able to provide
conditions that allow safe MR examinations also in the presence of implants, in
addition to these limit values based on physiology further scanner parameters
must be controlled and new limiting values for physical interactions with implants must be defined. This is
accomplished by the introduction of a specific implant scanning option (Fixed
Parameter Option: Basic, FPO:B) in the international MR safety standard [4] and
technical specifications for Implants [12].
Conclusion
To prevent accidents and damages, everybody working
with an MR scanner must be informed about possible risks originating from the
electromagnetic fields and the cryogens of the MR system.Acknowledgements
No acknowledgement found.References
[1] Schenck JF (2005) Physical interactions of static magnetic
fields with living tissue. Progress in Biophysics and Molecular Biology
87:185-204
[2] Chen DW (2001) Boy, 6, Dies of skull injury during MRI. The New
York Times, July 31, 2001
[3] Klucznik RP, Carrier DA, Pyka R, Haid RW (1993)
Placement of a ferromagnetic intracerebral aneurysm clip in a magnetic field
with a fatal outcome. Radiology 187:855-85
[4] International Electrotechnical Commission, IEC (2016)
Medical electrical equipment - Part
2-33: Particular requirements for the basic safety and essential performance of
magnetic resonance equipment for medical diagnosis. International Standard IEC 60601-2-33:2010+AMD1:2013+AMD2:2015
CSV Consolidated version: International Electrotechnical Commission,
Geneva, Switzerland
[5] Schaefer DJ (1998) Safety aspects of switched gradient fields.
Magn Reson Imaging Clin N Am 6:731-748
[6] Bourland JD, Nyenhuis JA, Schaefer DJ (1999) Physiologic effects of intense MR gradient
fields. Neuroimaging Clin N Am 9:363-377
[7] McJury M, Shellock F (2000) Auditory noise associated with MR procedures:
A review. J Magn Reson Imaging 12:37-45
[8] Knopp MV, Essig M, Debus J, Zabel HJ, van Kaick G
(1996) Unusual burns of the lower
extremities caused by a closed conducting loop in a patient at MR imaging.
Radiology 200:572-575
[9] Kugel H, Bremer C, Püschel M, Fischbach R, Lenzen
H, Tombach B, Van Aken H, Heindel W (2003) Hazardous situation in the MR bore: Induction
in EEG leads causes fire. Eur Radiol 13:690-694
[10] U.S. Food and Drug Administration (FDA), MAUDE Manufacturer and user facility device
experience (accessed Mar 14, 2017) www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfmaude/TextSearch.cfm
[11] Hardy PT, Weil KM
(2010) A review of thermal mr injuries.
Radiologic Technology 81:606-609
[12] ISO/TC150/SC6/JWG2 and IEC SC62B/JWG1 (2012) Assessment
of the safety of magnetic resonance imaging for patients with an active
implantable medical device, ISO/TS
10974:2012, International Organization for Standardization, Geneva, Switzerland