Synopsis
RF coils (antennas)
for MRI are designed to generate a RF magnetic field inside the patient. Large
body volume coils are optimized for the generation of a homogeneous RF magnetic
field. Local surface coils are designed to provide high signal to noise ratio.
Different designs and related physical aspects are discussed.
Introduction
In nuclear magnetic
resonance imaging (MRI) radio frequency (RF) signals establish a bi-directional
link between the MRI system and the patient [1].
The
clinical MRI systems of today operate at static magnetic field strengths Bo
between 0.2 Tesla and 3 Tesla with proton resonance frequencies between 8 MHz
and 130 MHz. RF coils for MRI are designed to generate a magnetic
field inside the patient. Only the negative (for reception) and positive (for
transmission) rotating magnetic field components contribute to the imaging
process. All other fields, like ~B0-directed magnetic fields or electric
fields have to be avoided as much as possible since they reduce the coils
sensitivity. Particularly the electric fields can be dangerous because these
are responsible for absorption of energy in the patient, resulting in heating
up the tissue. Energy absorption in transmit coils corresponds to noise
generation in receive coils, thus also receive coils have to be designed such,
that they generate low electric fields in case they would be used for
transmission. The RF-frequency increases with the static magnetic field
strength B0. At least, if the coil dimensions become comparable to
the wavelength inside the patient, propagation effects have to be included in
field simulations. In order to select the correct coil, the characteristics of
the volume of interest (VOI) is crucial. For whole body scans the standard
design of the birdcage coil is a convenient choice, whereas a surface coil is
more suitable for small regional scans, the head coil is preferred for taking
images of the brain.Electromagnetic field of the RF Coil
The
efficiency of RF coils for MRI systems in terms of optimal spin excitation and
maximised SNR is strongly influenced by the RF electromagnetic field
interaction with the human body as a load to the coil [2-6]. The degree of coil
loading is subject to the large variation in sizes of the patient population
ranging from babies to male or female adults with slim, normal or corpulent
stature and body weights from 3 kg up to about 120 kg. For MRI systems with
B0 > 0.5
T the Larmor frequency is > 21 MHz.
In this range the onset of dielectric effects associated with the high value of
the permittivity εr ~ 60 to 100 of
the body tissues is observed. As result, the wavelength in the body is
comparable with the body dimensions and, at higher frequencies the increasing
effect of the B1 eigenfield modifies the B1 field generated by the RF coil
inside of the body [1]. Furthermore, eddy currents due to the conductivity σ ~ 1 S/m of body
tissues reduce the penetration of the B1 field into the body at higher
frequencies (Fig1a). Assuming an average relative dielectric constant of εr =
50, the wavelength is about 33cm for 3T-systems or 14cm for 7T systems (Table
1). Predictions on RF coil performance must therefore be based on studies of
the coil with patient loads. Mimicking this condition by phantoms provides at
best an incomplete evidence of the real performance. A theoretical treatment of the
complicated RF boundary problems as given by the patient loaded RF coils
requires the application of numerical methods [7]. The method of moments offers
a fast simulation technique that includes propagation effects. It is based on
the electric and/or magnetic field integral equations (EFIE, MFIE). The
structure (but not the free space) is separated into sections small compared to
wavelength (and skin depth in case of dielectric bodys as-well). The (electric
and/or magnetic) currents on the structure are discretized in so-called base
functions and the total current can be described by a linear combination of
these base functions. By using given boundary conditions, a set of linear
equations is compiled to solve for the unknown base function weighting factors.
After solving this with standard iterative or direct solving methods, the electric
and magnetic fields is calculated by integral equations.
A patient model can be
incorporated by equivalent currents on the surface or inside the tissue. Fig. 1c shows the magnetic field inside a homogeneous human head model at 7T calculated
by the Method of Moments [8]. Due to resonance effects in the patient model,
the field is much less homogeneous compared to the unloaded case [12, 13]. The FDTD method is based on a finite difference approximations in space and
time that solves Maxwell’s time dependent curl equations and widely applied for
EM simulations in MRI [9,10]. If the FDTD simulation is fed by an impulse
response, the result after Fourier transformation yields information on the RF
properties of the model over a wide frequency range. The smallest detail of the
simulated RF coil model geometry typically dictates the requirements of the
equal-distance cubical grid size, resulting in very long computation times.
This makes the method less suitable for RF coil design, but suitable for the
verification process. Therefore, the method is widely used for characterization
of the local absorption (SAR) inside a high resolution dielectric model of the
human body (see Fig.1d). The finite difference method divides the object into (interleaved)
cubical cells, thus it cannot accurately model arbitrary geometries. This
disadvantage is overcome by the finite element method, which divides the region
of interest into irregular tetrahedron cells. Since the size and orientation of
these cells can be arbitrary, the finite element method can accurately model
arbitrary geometries [11]. RF Coil Characteristics
RF coils, as first part of the RF chain, have a severe
impact on the final signal-to-noise (SNR) level. The equivalent measure of the
SNR for transmit coils is the B1 power
sensitivity SP. The power sensitivity is defined by the amount of power P0 required to
produce a certain field strength at a reference point, typically at the isocenter of the unloaded
coil [2],
$$S_{PO}=\frac{\hat{B}_1}{\sqrt{P_0}}$$
The power sensitivity of the loaded coil
is
$$S_{PI}=\frac{\hat{B}_1}{\sqrt{P_P+P_O}}=S_{P0}*\sqrt{\frac{Q_l}{Q_o}}$$
with the power P0 dissipated in the
resistance R0 of the unloaded coil and the power PP
dissipated in the equivalent resistance RP of the patient or
phantom. The power sensitivity SPI also be expressed in terms of the unloaded and loaded Q-factors.
The Q‑factor $$$Q=\frac{\omega L}{R}$$$ is an inverse measure of
the RF losses in the coil with ωL being the reactive and R the
resistive part of the impedance of the coil. The losses of the unloaded
coil, modeled by a series resonant circuit, R0, are caused by the equivalent
series resistance of the capacitors and by the resistance of the conductor
structure. The RF losses induced in the patient add a further component RP
to the resistive part of the coil impedance. The resistance R0
can be minimized by a careful design of the coil; however, the resistance RP,
determined by the patient, can (e.g.) be reduced by restricting the FOV seen by
the coil. The load factor$$$LF =\frac{Q_0}{Q_l}$$$ is the ratio between the quality factor QO of the unloaded
coil and the quality factor Ql of the loaded coil. The power
efficiency η of the coil can be expressed using the load factor, the
powers or the resistances
$$\eta=\frac{P_{P}}{P_{0}+P_{P}} =\frac{LF-1}{LF}=\frac{R_P}{R_P+R_0}$$
The power efficiency of the coil increases
with the load factor LF. The
load factor is
sufficiently high, if the loading effects dominate and the resistance of the
coil structure is negligible. Then, the noise results from the presence of the
patient and the RF coil itself does not notably increase the noise level.
Applying the analogy for transmission and reception, the transmitted power is
predominantly dissipated in the patient and not in the RF coil. The load factor
of an MRI coil can be influenced by the design of the coil and the placement of
the coils with respect to the RF screen and the load.
Within the patient load exposed to the
B1 field of the RF coil, the principal RF loss and thus for MRI
irreducible contribution arises from the electric fields which are magnetically
induced in the different conductive body tissues. However, the interaction of
the unavoidable electric fields in the RF coil structure with the patient load
cause an additional loss contribution, which can be minimized by increasing the
self-resonance frequency of the coil with additional series capacitors in the coil
conductors. The sign and the value of the relative frequency shift, $$$\triangle f_l/f_0 =(f_l - f_0)/f_0$$$, observed when the RF coil is loaded by the
patient, characterise the contribution from the residual RF electric fields in
the coil to the RF absorption in the patient in terms of unwanted dielectric
losses [2]. The dielectric loss contribution is negligible when Δfl/f0 is positive, i.e. the
resonance frequency of the RF coil rises slightly when the patient is inserted.
Since the RF loss induced by the transversal B1
field in the patient is irreducible in MRI, high efficiency of the RF coil
requires a high value of the ratio Q0/Ql . Obvious goals
are a high Q0 together with a low value of Ql, which
requires strong coupling to the patient load using low impedance RF coils.
Typical values for the Q factor range in MR coil design from QL =
100 (loaded) to QO=400 (unloaded), e.g. measured on a phantom for
the first and in air for the latter.
RF Coil Measurements
The RF loss contributions of the RF
coil and of the patient load can be found from measured values of the resonance
frequencies and the Q‑factors for the empty coil, f0, Q0,
and for the coil with patient load, fl, Ql.
The quality factor of a resonant coil
can be measured using a single pickup-coil connected to a network analyzer
[14,15,16]. Fig. 2a shows the schematics of a pickup coil, coupling to a series
resonant circuit. The network analyzer can be short-calibrated for M= 0 to
compensate for the pickup-coil impedance jωLPU. Assuming a low coupling M
and a not too low quality factor Q, the reflection at
the so called 3dB-frequencies provides
the quality factor Q=ω0/Δω. The
3dB-frequencies can be found using the logarithmic display on a network analyzer
(Fig.2c). If the 0dB line is placed at the top of the display and the scaling is
chosen such that the resonance peak just hits the bottom, than the curve
displayed intersects the center line at the 3dB-frequencies as shown in Fig.2b. With the power absorbed in the resonator $$$P=\frac{1}{2} R |I_{coil}|^2$$$
the power sensitivity is given by $$$S_P =B_1 /\sqrt{P}$$$. The power sensitivity can be obtained
from a transmission measurement. The sensitivity SP obtained by a transmission
measurement (Impedance of cable Z0, area of pick up coil APU, Inductivity of
pick up coil LPU) is given by:
$$S_p =\frac{\sqrt{2 Z_o}}{\omega A_{PU}}|S_{21}|\sqrt{\frac{1+(\frac{\omega L_{PU}}{Z_o})^2}{1-|S_{11}|^2}}$$
If the coil to be tested already has a feeding port, it can be connected to
the network analyzer (Fig.7). Alternatively, the coil can be fed by a separate
pickup-coil. The second port of the network analyzer is connected to a
receiving pickup-coil which is placed at that point where SP is to
be determined. The network analyzer radiates a power Pnwa,out into the cable
connected to its port 1. Neglecting losses in the pickup-coil 2, the power $$$P_{coil} = P_{nwa,out} (1-|S_{11}|^2)$$$ reaches the MRI-coil. Always cable
losses have to be calibrated out. If there are significant losses in the
pickup-coil, these also have to be considered.
Volume Coils
Ideal volume coils for MRI use with a human body, head or
limb have cylindrical symmetry, sample access from one or both sides and should
provide a uniform transverse magnetic B1 field. A uniform
transverse magnetic field can be generated by an axial current flow with cosine
current distribution around a cylinder wall. For practical applications, the
continuous wall current sheet can be approximated by N equiangular spaced
discrete current rods. This cylindrical N‑conductor structure forms a
transmission line which propagates transverse electromagnetic (TEM) modes r,
with $$$1\leq r \leq N/2$$$ for the unshielded structure and $$$0\leq r \leq N/2$$$in the presence of an external coaxial shield,
where the current distribution in the rods numbered $$$i =1,2,..,(N-1),N$$$ is given by:
$$I_i,_r = I_0 cos [ (i-1)r2\pi/N]$$
The r = 1 mode provides the required
homogeneous transverse magnetic field for applications in RF coils, where
a short section of the N-conductor TEM transmission line [17,18,19] is terminated at both ends with suitable
reactances to establish resonance. Principally, there are two methods of
terminating the TEM section, which differ in the paths of the RF currents
outside the N‑conductor section.
Hayes et.al. [20] introduced the birdcage structure
with ring sections interconnecting all
rods at both ends as termination of the TEM line section and with tuning
capacitors applied in series to the rods (low-pass) or to the ring sections
(high-pass) or to both positions (band-pass). Here the rod currents accumulate
in the ring to high values depending on N, and the inductances of the ring
sections strongly influences the separation of the resonance modes. Tropp [21]
presented an extended theory of the birdcage resonator and of the involved
mutual inductances in the unshielded and shielded TEM line sections.
Depending to the
placement of the tuning capacitors in series to rods, to rings, or to both, the
birdcage structure [22] is called low-pass,
high-pass, or band-pass (Fig.3a). The mutual
coupling between the conducting rods and rings and the selection of the
capacitor values result in multiple resonant modes.
According to the
current distribution, each of the modes generates different distributions of
the electromagnetic field. The desired cosine-shaped
current distribution is achieved by adjustment of the resonance frequency of
the mode generating the homogeneous field. The whole body transmit coil
is located inside a cylindrical RF screen, to provide a defined RF environment.
Without the RF screen, the coil strongly couples with the gradient coil and
reduces the performance of both coils. The cylindrical RF shield is transparent
(high impedance) for the gradients to prevent eddy currents and impermeable
(low impedance) for RF to reduce losses during transmission and interfering
signals during reception.
Matching the Coil
MR transmit coils are
matched to the impedance of the RF chain (50Ω) in order to avoid power reflection, and thus, to decrease
power efficiency [23, 24]. Generally, an RF coil exhibits a complex impedance,
so that a matching network with at least two degrees of freedom is required. Preferably,
the matching is achieved with capacitors, as capacitors are available over a
wide range in a compact shape and do not generate magnetic stray fields by themselves. Driving the coil in quadrature
The desired circularly
polarized RF field is achieved by driving the body coil in quadrature mode. The
quadrature drive of a cylindrical body coil requires the existence of two input
ports, which have to be driven by two input RF signals with 90o phase
difference [25,26,27]. A quadrature hybrid splits the Tx input signal at port1 (Fig.3b) to the
ports 2 and 3. During reception, the receive signals are combined and appear at
port 4. The ports have to be connected with respect to the static magnetic
field B0, so that the rotation direction is in the same direction as
the Lamor precession. A quadrature combiner can have several tenths of a dB of insertion loss,
lowering the overall noise figure of the receiver correspondingly. Hybrids for lower frequencies can be realized
easily using lumped element components. In the quadrature hybrid, the reflected
waves are cancelled, if the mismatched load is symmetric.Multi Channel Transmit and Receive Coils
At field strengths higher than 3T, wave propagation effects (sometimes
called “dielectric resonances”) represent a primary cause of RF field
inhomogeneities in body MR imaging. RF shimming techniques, based on multi-channel
RF transmit technology [28-31], bear the potential of compensating these RF
field inhomogeneities. Furthermore, parallel transmission systems allow advanced
imaging techniques, in particular the acceleration of multi-dimensional
RF-pulses [32,33]. The transmit coil consists of a degenerate birgcage [34,35]
loops [36] or a degenerate TEM coil [37,38], in which the individual coil elements
are decoupled from another. Each individual decoupled coil element is driven by
an individual RF source. A method to protect the RF power amplifiers against
reflected and coupled power from the coil elements is to locate circulators
between coil and amplifiers. Each channel can be controlled individually in
amplitude and phase in order to create spatio-temporal variations of the B1-
field, which in sum with the propagation effects cause the desired RF field.Receive Coils
During the acquisition, the induced voltage,
caused by rotating spins, is detected by an RF receive coil and is stored in
memory. After acquisition has finished, linear transformation (Fourier
transformation) produces the final MR image from the sampled voltage values.
Thus, the data of the final image is proportional to the detected voltage,
which itself is a superposition of signal from spin and noise. The SNR relates
the voltages of noise and spin [40-43]:
$$\frac{S}{N} (r) =\frac{\omega V\mid M * \frac{B(r)}{I}\mid}{\sqrt{8kT\triangle f (R_A +R_L )}}$$
The spin signal (numerator) is proportional to
the frequency w, the considered Volume V, the nuclear magnetization M, and the magnetic
flux density B in the point r in space produced by the current I. The resistors
RA (antenna loss) and RL (load loss), the bandwidth Df and
the temperature T dictate the noise (denominator). The noise of a well-designed
MR coil must not be dominated by the antenna loss.The signal-to-noise ratio attainable with a MR receive
coil is inversely proportional to the effective coil losses. It has been shown
that these resistive losses are proportional to the patient volume illuminated
by the coil [44-45]. The losses of the unloaded coil, R0, are caused by the resistance
of the conductor structure and the equivalent series resistance of the
capacitors. The RF losses induced in the patient add a further component RP
to the resistive part of the coil impedance. The resistance R0
can be minimized by careful design of the coil; however, the resistance RP,
determined by the patient, can only be reduced by restricting the FOV. Hence,
for dominating patient losses, the SNR is nearly inversely proportional to this
patient volume. Depending on the application, the region-of-interest (ROI)
encompasses only a fraction of the total patient volume seen by the coil.
Consequently, the noise induced is larger than the value corresponding to the
ROI if a body sized coil is used. Therefore, dedicated coils were introduced to
improve the SNR (see Fig.4a). Receive coils, designed for certain parts of the
human body are called surface coils due to their placement directly on the
patient. A single receive coil as depicted in Fig.5a consists of the conductor
structure, which is optimized for the dedicated imaging application.
Distributed series capacitors along the conductor provide equal current
distribution and reduce dielectric losses.
Preamplifiers are necessary to boost the extremely
weak signal. The preamplifier should be in close proximity of the coil so that
additional components do not diminish the performance. The best noise figure is
achieved if the matched coil presents the so-called optimum noise impedance to the preamplifier. This impedance is realized by transforming the
impedance of the coil via a low loss matching network to the optimum noise
impedance of the preamplifier [47,48].
Coil Detuning
The RF receive coil elements are optimized to resonate
at the MR frequency. During excitation of the nuclear spins, a strong RF field
is applied in the same space as where the receive coil resides. Without any
counter measures the current in the receive coil elements would rise to
unacceptable high levels
To overcome this problem, the receive coil elements
are switched to detune state during the transmission of the RF excitation pulse
[49,50]. Circuits that prevent or reduced current flow in the coil element as a
result of an induced RF voltage are shown in Fig 5b. If the MR system is in
receive mode, the detuning circuit prevents inter-element or inter-coil
coupling. If the MR system is in transmit mode, the detune circuit must prevent
distortion of the B1 excitation field. The detune circuit is an important component
for patient safety. It prevents exceeding the local SAR limits.Coil Cable and RF Trap
Typically a coaxial cable connects the preamplifier to
the rest of the RF chain. The coil elements are magnetically and electrically
coupled with the transmit coil. The coil as a whole or the cable on its own can
act like an antenna. If this is the case, energy is withdrawn from the transmit
coil and the intended homogeneity of the transmit coil is compromised. The
common mode current on the cable itself is a source of magnetic field
surrounding the cable. This locally changes the magnitude of the magnetic field
and hence the flip angle of the excitation. The shield of the cable has to be
of high impedance for currents induced by the transmit field to avoid imaging
artifacts and heating. Among other solutions, parallel resonant baluns with
high impedances are used to minimize the RF currents on the shield [51, 52]. Decoupling of RF Coils
It is common practice to describe two coupling coils using an equivalent
circuit diagram. In this diagram the inductive coupling is represented by a
mutual inductance on a common path. It is obvious, that for decoupling this
shared inductance has to be compensated by an additional capacitor [53,54]. The
value of the capacitor must yield the same impedance as the inductance does (Fig5c).
Since this condition can be satisfied only for a single frequency, capacitive
decoupling is always narrowband. Another
way to reduce coupling is the idea to arrange adjacent coils in such a smart
manner, that the integral flux through the surface of a secondary coil caused
by the primary coil vanishes (Fig5d). This can be done very simple for next neighbor
coils overlapping them [55]. But the complexity increases using an array, which
usually comprehends next-neighbor and non-next-neighbor coils. On the other
hand, the coupling between non-next-neighbor coils is much less critical, therefore
it is needful to check individually, if decoupling is mandatory. Receive Coil Arrays
The application of arrays [55] of several elements overcomes
the limitation of the restricted FOV of a single surface coil and provides optimized
imaging performance over large body areas (Fig.4b). The individual
signals detected by the coil elements are fed to individual receiver channels. During
reconstruction, the individual signals of these coils are combined, and a higher
SNR is available for a large imaging volume. Arrays can be designed to be tight fitting to the body and effectively
receiving from volumes smaller than a typical body volume coil. In case
noise signals are coupled from one coil to the other, the SNR of the combined
image is reduced. To prevent that such correlated noise diminishes image
quality, the individual coil elements have to be mutually decoupled. Minimization
of the coupling and noise correlation can be achieved by different strategies
[55-59]. Next neighbor coil elements can be mutually decoupled using a defined
element-to-element overlap, so that the mutual inductance is cancelled. However,
this technique restricts the shape and the positioning of the elements, which
limits the optimization of the coil for imaging parameters or body regions.
Furthermore, the decoupling of the second next neighbors in an array of coils
is challenging. Preamplifier decoupling provides effective decoupling of a coil
from its neigbours. For optimum performance, the
network shown in Fig.5a transforms the coil impedance ZCoil
to 50 Ω to offer noise match at the input to the amplifier
[60] and inversely transforms the amplifier input impedance ZAmp
to a quasi open circuit Zin at the terminals of the coil. This
method reduces coil currents, and thus, unwanted coupling effects between
neighboring antennas. Ideally, the individual coils maintain their
individual SNR as if operating as a single element. In advanced methods [61- 63],
multi-element surface coils are used to encode the spatial information. These
methods utilize the spatial dependence of the B1 field maps of the coils to accelerate the
imaging process.Double Resonant Coils
In
addition to proton imaging (which is the most abundant atom), other nuclei
(such as 31P or 19F) are used as well, to quantify metabolic concentrations in
neural or muscular tissues. Double tuned coils are used for imaging and
spectroscopy at high fields. However, this requires dedicated MRI transmit and
receive coils resonant at both frequencies and maintaining a good performance
for both frequencys. 19F MR imaging has a high potential for the detection and
direct quantification of fluor-labeled tracers and drugs in the field of
molecular imaging [64, 67]. The combination with 1H imaging provides the
associated anatomical information. Thus far, the integration of separate 19F and
1H RF coils has typically been used to support this application. The challenge
is to provide virtually the same sensitivity profile for the two frequencies.
This allows the correction of B1 inhomogeneities in the 19F image via the 1H
sensitivity profile.
TEM-
and Birdcage designs exit using separate feeding ports for both nuclei [68-71].Acknowledgements
The author wish to acknowledge Christian Findeklee, Oliver Lips and Peter Vernickel for valuable assistance and helpful discussion.References
[1] Röschmann P. RF coil (Antennes)
for MRI, Syllabus contribution, 16th ESMRMB 1999
[2] P. Roeschmann. Radiofrequency Penetration and Absorption in the Human
Body:
Limitations to High-Field Whole-Body Nuclear Magnetic
Resonance Imaging.
Med. Phys., 14(6):922–931, 1986.
[3] J. T. Vaughan, M. Garwood, C.M. Collins, W. Liu, L. DeLaBarre, G.
Adriany,
P. Andersen, H. Merkle, R. Goebel, M. B. Smith, and K.
Ugurbil. 7 T vs. 4
T: RF Power, Homogeneity, and Signal-to-Noise
Comparison in Head Images.
Magnetic Resonance in Medicine, 46:24–30, 2001.
[4] T. S. Ibrahim, R. Lee, A. M. Abduljalil, B. A. Baertlein, and P.-M. L.
Robitaille.
Dielectric Resonances and B1 Field Inhomogeneity in
UHFMRI Computational
Analysis and Experimental Findings. Magnetic Resonance
Imaging, 19:219–
226, 2001.
[5] Q. X. Yang, J. Wang, X. Zhang, C. M. Collins, M. B. Smith, H. Liu,
X.-H. Zhu,
J. T. Vaughan, K. Ugurbil, and W. Chen. Analysis of
Wave Behavior in Lossy
Dielectric Samples and High Field. Magnetic Resonance
in Medicine, 47:982–
989, 2002.
[6] A. Kangarlu, B. A. Baertlein, R. Lee, T. Ibrahim, L. Yang, A. M.
Abduljalil, and
P.-M. L. Robitaille. Dielectric Resonance Phenomena in
Ultra High Field MRI.
Journal of Computer Assisted Tomography,
23(6):821–831, 1999.
[7] Jianming Jin: Electromagnetic
Analysis and Design in Magnetic Resonance Imaging; CRC Press 1997, ISBN
0-8493-9693-X
[8] [CON] CONCEPT II: Method of
Moments program of the Technical University of Hamburg-Harburg
[9] TS Ibrahim et al Phys. Med. Biol. 46(2001) 609-619
[10] Karl S. Kunz, R.J. Luebbers, The FDTD Method for electromagnetics ISBN
0-8493-8657-8
[11] ANSYS HFFS Simulator
[12] Leussler C. et al, ESMRMB
Shielded Endcap Bandpass Resonator for Ultra High Field Imaging at 7 Tesla EPOS
389, 2004
[13] DeMeester G., Shoulder
Shield for 7.0T T/R Brain Coil , et al,
Proc. Intl. Soc. Mag. Reson. Med. 14 (2006) p. 2610
[14] Hoult,D.I., The NMR Receiver: A Description and Analysis
of Design, Progr. NMR Spectr.,12, pp.41-77, 1978
[15] Ginefri, J.C. et al Quick measurement of nuclear magnetic
Resonance Coil Sensitivity with a single
loop probe, Rev. Sci. Instrum. 70,4730-4731,1999
[16] Chen, C.N. et al The field dependance of NMR Imaging.I.
Laboratory Assessment of Signal-to Noise Ratio and Power deposition, Magn. Reson. Med.,3, 722-729, 1986
[17]
Röschmann P. US patent 5160890, 1991.
[18] Bridges
J. Cavity resonator . U.S. Patent 4751464, 1988.
[19] Chingas GC, Zhang N. Design
Strategy for TEM high Field Resonators. Proceedings of the 4th Annual Meeting of ISMRM, New York, USA
1996: p.1426.
[20] Hayes CE, Edelstein WA, Schenk JF, Mueller OM, Eash
M. An Efficient, Highly Homogeneous RF coil for whole Body NMR Imaging at 1.5T.
J Mag Reson 1985;63:622-628.
[21] Tropp J. Mutual Inductance
in the Bird-Cage. J Magn Reson 1997;126-1:9-17.
[22] Röschmann P. Analysis of Mode
Spectra in Cylindrical N-Conductor Transmission Line Resonators with expansion
to Low-, High- and Band-pass Birdcage Structures. Proc. Intl. Soc. Mag. Reson.
Med.1995;3: p. 1000.
[23] P.L. Kuhns et al, “Inductive
Coupling and Tuning in NMR Probes; Applications,” J. Magn. Reson. 78, 69-76
(1988)
[24] ARRL Handbook of Radio
Communications 86th edn. Wilson, M.J.(ed), American Radio Relay League, Inc.
Newington, USA, 2009
[25] Jin J.
Elektromagnetic Analysis and Design in MRI, CRC Press, New York, 1999.
[26]
Mispelter J, Lupu Mihaela L, Briguet A. NMR probeheads, Imperial College press,
2006.
[27] Sorgenfrei B, Edelstein W.
Optimizing MRI signal-to-noise ratio for quadrature unmatched RF coils. Magn
Reson Med 1996; vol. 36:104-10.
[28] Ullmann P, Junge S, Wick M,
Seifert F, Ruhm W, Hennig J. Experimental analysis of parallel excitation using
dedicated coil setups and simultaneous RF transmission on multiple channels. Magn.
Reson. Med. 2005;54:994-1001.
[29] Zhu Y, Watkins R, Giaquinto
R, Hardy C, Kenwood G, Mathias S, Valent T, Denzin M, Hopkins J, Peterson W,
Mock B. Parallel excitation on an eight transmit-channel MRI system. Proc.
Intl. Soc. Mag. Reson. Med. 2005;13: p.14.
[30] Setsompop K, Wald L,
Alagappan V, Gagoski B, Hebrank F, Fontius U, Schmitt F, Adalsteinsson E.
Parallel RF transmission with eight channels at 3 Tesla. Magn. Reson. Med. 2006;56:1163-1171.
[31] Graesslin I, Vernickel P,
Schmidt J, Findeklee C, Röschmann P, Leussler C, Haaker P, Laudan H, Luedeke
KM, Scholz J, Buller S, Keupp J, Börnert P, Dingemans H, Mens G, Vissers G,
Blom K,Swennen N, van der Heijden J, Mollevanger L, Harvey P, Katscher U. Whole
body 3T MRI system with eight parallel RF transmission channels. Proc. Intl. Soc. Mag. Reson. Med. 2006;14: p.129.
[32] Katscher U, Börnert P,
Leussler C, van den Brink J. Transmit SENSE. Magn. Reson. Med. 2003;49:144-150.
[33] Zhu Y. Parallel excitation. Magn. Reson. Med. 2004; 51:
775-784.
[34] Leussler C, Stimma J,
Röschmann P. The
Bandpass Birdcage Resonator Modified as a Coil Array for Simultaneous MR
Acquisition. ISMRM 1997; p. 176.
[35] Alagappan V, Nistler J,
Adalsteinsson E, Setsompop K, Fontius U, Zelinski A, Vester M, Wiggins GC,
Hebrank F, Renz W, Schmitt F, Wald LL. Degenerate Mode Band-Pass Birdcage Coil
for Accelerated Parallel Exitation. Magn Reson. Med. 2007;57:1148-1158.
[36] Weyers
DJ. Proc. Intl. Soc. Mag. Reson. Med. 2002; p. 901.
[37] Leussler, C. Wo patent 02/095435, 2002.
[38] Vernickel P, Röschmann P, Findeklee C, Lüdeke KM,
Leussler C, Overweg J, Katscher U, Grässlin I, Schünemann K. Eight-Channel
Transmit/Receive Body MRI Coil at 3T. Magn Reson. Med.
2007;58:381-389.
[39] Vester M. Proc. Intl. Soc. Mag. Reson. Med. 2006; p. 2024.
[40] D. I.
Hoult. The Principle of Reciprocity in Signal Strength Calculations – A
Mathematical
Guide. Concepts in Magnetic Resonance, 12(4):173–187, 2000.[41] Schnell W et al
Ultimate signal-to-noise-ratio of surface and body antennas for magnetic
resonance imaging, IEEE Transactions on Antennas and
Propagation (Volume: 48, Issue: 3, Mar
2000 )
[42] C. M.
Collins and M. B. Smith. Calculations of B1 Distribution, SNR, and SAR
for
a Surface Coil Adjacent to an Anatomically-Accurate Human Body Model.
Magnetic
Resonance in Medicine, 45:692–699, 2001.
[43] W. A.
Edelstein, G. H. Glover, C. J. Hardy, and R. W. Redington. The Intrinsic
Signal-to-Noise
Ratio in NMR Imaging. Magnetic Resonance in Medicine,
3:604–618,
1986.
[44] Hoult DI, Richards RE. The signal-to-noise ratio of
the nuclear magnetic resonance Experiment. J Magn Res.1976;24:71-85.
[45]
Röschmann P, Tischler R. Surface coil proton imaging at 2 T 1986;Radiology
161:251-255.
[46] Wang J, Reykowski A, Dickas J. Calculation of the
Signal-to-Noise Ratio for Simple Surface Coils and Arrays of Coils. IEEE Trans.
Biomed. Eng.1995;Vol. 42, No. 9:908-17.
[47] Reykowski
A, Wright SM, Porter JR. Design of Matching Networks for Low Noise
Preamplifiers.
Magn Reson Med 1995;33:848-852.
[48] Findeklee
C. Proc. Intl. Soc. Mag. Reson. Med. 2009;17:507
[49] Edelstein et al., Electronic decoupling of
Surface-Coil Receivers for NMR Imaging and Spectroscopy1986, J. Magn.
Reson.,67, 156-161.
[50] Joel Mispelter, Mihaela Lupu, Andre Briguet, NMR
Probeheads, Imperial College Press ISBN 978-1-84816-662-2, 2015
[51] R.
Dean Straw. The ARRL Handbook for Radio Communications. ARRL-the
National
Association for Amateur Radio, 2006.
[52] Seeber et al., Floating Shield Current Current
Suppression Trap, Concepts in Magn. Reson., 21B, 26-31, 2004
[53] Wang
J., Proc. ISMRM 4:1434 (1996)
[54] Jevtic
J., Proc. ISMRM 9:17 (2001)
[55] Roemer,
et al, The NMR Phased Array. MRM 1990;16:192-225
[56] Leussler C. and Holz D. (1991) Improvement of SNR at
low field strength using mutually decoupled coils for simultaneous NMR imaging.
Proc. 11th SMRM 1991;p.734.
[57] Reykowski A. Theory and Design of Synthesis Array
Coils for MRI, Texas A&M Univ., Dissertation, 1996.
[58] Hutchinson M, Raff U. Fast MRI data acquisition
using multiple detectors. Magn Reson Med 1998;6:87–91.
[59] Kwiat D, Einav S. A decoupled coil detector array
for fast image acquisition in magnetic resonance
imaging. Med Phys 1991;18:251–265.
[60] Reykowski A, Wright SM, Porter. JR Design of
Matching Networks for Low Noise Preamplifiers, Magn Reson Med 1995;33:848-852.
[61] Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P.
SENSE: sensitivity encoding for fast MRI. Magn Reson Med 1999;42:952-962.
[62] Weiger M, Pruessmann K, Leussler C, Roeschmann P,
Boesiger P. Specific coil design for Sense: a six element cardiac array. MRM
2001;45:495.
[63] Sodickson DK , Manning WJ. Simultaneous acquisition
of spatial harmonics (SMASH). Magn Reson Med 1997;38:591-603.
[64] Morawski
A et al., MRM 52:1255 (2004)
[65] Gonen
O et al., MRM 37:164(1997)
[66] Krems
B et al., Magn Reson B 108:155(1995)
[67] Mazurkewitz
et al EPOS 821 page 370-371 ESMRMB (2006)
[68] X. Zhang et al.: A circular-polarized double-tuned
(31P and 1H) TEM coil for human head MRI/MRS at 7T, Proc. Intl. Soc. Mag.
Reson. Med. 11 234 (2003)
[69] S. B. King et al.: A Comparasion of Double Tuned
Birdcage and Spiral Birdcage RF Coils, Proc. Intl. Soc. Mag. Reson. Med. 11
(2004) 1559
[70] W. Dürr, S. Rauch: A Dual-Frequency Circularly
Polarizing Whole Body MR Antenna for 69/170 MHz , Magnetic Resonance In
Medicine 19 (1991) 446-455
[71] Findeklee et al, Efficient Design of a novel Double
Tuned Quadrature Headcoil
for
Simultaneous 1H and 31P MRI/MRS at 7T p.
891 ISMRM 2005