As the main magnetic field strength increases, the corresponding RF wavelength is shortened. This leads to pronounced wave effects in the transmit field, causing inhomogeneous excitation. Multi-channel arrays provide additional degrees of freedom to mitigate such effects and to manipulate (or to tailor) RF transmission. Roughly these can be divided in 3 types, namely local arrays, remote circumferential arrays and travelling wave arrays. Examples of these arrays are presented in this educational talk.
Arrays for imaging of the torso and large fields of view can roughly be divided into three groups: local arrays (9-15), remote circumferential arrays (16, 17) and travelling wave antennas/arrays (18, 19), though there is a smooth transition between these groups.
Local arrays achieve a high transmit efficiency and if used as a transmit/receive array they also provide a high SNR at least in the periphery. On the downside these arrays can take up a lot of space within the small bore diameter of about 60 cm due to the necessary thickness of the elements. This makes it difficult to include subjects with a larger body physique and reduces patient comfort.
Remote circumferential arrays or integrated arrays have the advantage of leaving more space for the patient inside the bore, but they are less power efficient and exhibit a lower SNR at least in the periphery when used as transmit/receive arrays. Furthermore, integrating such arrays is time consuming and quite intricate because a lot of the system has to be disassembled. Figure 1 shows an example of such an integrated 32-channel body coil. The freed space inside the bore can be used to accommodate subjects with a larger body physique, or to include other equipment like dedicated local receive arrays. When installing remote circumferential coil arrays, it may be necessary to also include extra peripheral components like controllers for T/R-switches and detuning as well as switches for the receive chain to ensure other experiments with local transmit coils are not influenced and that not too many system resources (PIN-diode currents, receive channels etc.) are used up since the coil arrays cannot be easily removed.
Travelling wave concepts are easy to implement, but their transmit efficiency can be very low. When imaging the center of the body, half of the body tissue in one z-direction has to be penetrated to reach the FOV. Concerning space inside the bore, this concept shows the same advantages as the remote circumferential arrays.
Results & Discussion
The existence of travelling wave concepts shows that an important issue to consider in all types of arrays is wave propagation. At higher field strengths the ratio of the size of the transmit elements to the wavelength increases, leading to potentially higher radiation, especially since waveguide modes can propagate along the bore. Concepts known from lower field strengths like estimating the SNR efficiency from the ratio of loaded Q to unloaded Q are not valid anymore since the losses in the “unloaded” case include radiation losses that are not necessarily equal to the radiation losses in the loaded case where more of the power is radiated towards the sample. On the other hand, the radiation has to be taken into account since it has an impact on field distribution inside the bore and power can be radiated out of the bore, potentially causing unwanted interaction with magnet room peripheral components. Furthermore one fundamental difference between transmit arrays and receive arrays is the most important figure of merit of each case. While for receive elements it is $$$\frac{B_1^{-}(\overrightarrow{r})}{\sqrt{P}}$$$ as a measure of SNR at the point of interest, for transmit arrays the more important value is $$$\frac{B_1^{+}(\overrightarrow{r})}{\sqrt{SAR_{most critical}}}$$$ since in the transmit case the limit for the most critical SAR aspect is a hard limit while more transmitter power is just a matter of amplifier costs.1. Hoult DI, Phil D. Sensitivity and power deposition in a high-field imaging experiment. Journal of magnetic resonance imaging : JMRI. 2000 Jul;12(1):46-67. PubMed PMID: 10931564.
2. Van de Moortele PF, Akgun C, Adriany G, Moeller S, Ritter J, Collins CM, et al. B(1) destructive interferences and spatial phase patterns at 7 T with a head transceiver array coil. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 2005 Dec;54(6):1503-18. PubMed PMID: 16270333.
3. Vaughan JT, Garwood M, Collins CM, Liu W, DelaBarre L, Adriany G, et al. 7T vs. 4T: RF power, homogeneity, and signal-to-noise comparison in head images. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 2001 Jul;46(1):24-30. PubMed PMID: 11443707.
4. Collins CM, Liu W, Swift BJ, Smith MB. Combination of optimized transmit arrays and some receive array reconstruction methods can yield homogeneous images at very high frequencies. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 2005 Dec;54(6):1327-32. PubMed PMID: 16270331.
5. Mao W, Smith MB, Collins CM. Exploring the limits of RF shimming for high-field MRI of the human head. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 2006 Oct;56(4):918-22. PubMed PMID: 16958070. Pubmed Central PMCID: 4040521.
6. Orzada S, Maderwald S, Poser BA, Bitz AK, Quick HH, Ladd ME. RF excitation using time interleaved acquisition of modes (TIAMO) to address B1 inhomogeneity in high-field MRI. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 2010 Aug;64(2):327-33. PubMed PMID: 20574991.
7. Grissom W, Yip CY, Zhang Z, Stenger VA, Fessler JA, Noll DC. Spatial domain method for the design of RF pulses in multicoil parallel excitation. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 2006 Sep;56(3):620-9. PubMed PMID: 16894579.
8. Katscher U, Bornert P, Leussler C, van den Brink JS. Transmit SENSE. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 2003 Jan;49(1):144-50. PubMed PMID: 12509830.
9. Graessl A, Renz W, Hezel F, Dieringer MA, Winter L, Oezerdem C, et al. Modular 32-channel transceiver coil array for cardiac MRI at 7.0T. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 2014 Jul;72(1):276-90. PubMed PMID: 23904404.
10. Metzger GJ, van de Moortele PF, Akgun C, Snyder CJ, Moeller S, Strupp J, et al. Performance of external and internal coil configurations for prostate investigations at 7 T. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 2010 Dec;64(6):1625-39. PubMed PMID: 20740657. Pubmed Central PMCID: 2991410.
11. Oezerdem C, Winter L, Graessl A, Paul K, Els A, Weinberger O, et al. 16-channel bow tie antenna transceiver array for cardiac MR at 7.0 tesla. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 2015 Jul 17. PubMed PMID: 26183320.
12. Orzada S, Quick HH, Ladd ME, Bahr A, Bolz T, Yazdanbakhsh P, et al. A flexible 8-channel transmit/receive body coil for 7 T human imaging. In: Proceedings of the 17th scientific meeting, International Society for Magnetic Resonance in Medicine, Honolulu, p 2999. 2009.
13. Raaijmakers AJ, Italiaander M, Voogt IJ, Luijten PR, Hoogduin JM, Klomp DW, et al. The fractionated dipole antenna: A new antenna for body imaging at 7 Tesla. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 2015 May 2. PubMed PMID: 25939890.
14. Snyder CJ, Delabarre L, Moeller S, Tian J, Akgun C, Van de Moortele PF, et al. Comparison between eight- and sixteen-channel TEM transceive arrays for body imaging at 7 T. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 2012 Apr;67(4):954-64. PubMed PMID: 22102483. Pubmed Central PMCID: 3290686.
15. Winter L, Ozerdem C, Hoffmann W, Santoro D, Muller A, Waiczies H, et al. Design and evaluation of a hybrid radiofrequency applicator for magnetic resonance imaging and RF induced hyperthermia: electromagnetic field simulations up to 14.0 Tesla and proof-of-concept at 7.0 Tesla. PloS one. 2013;8(4):e61661. PubMed PMID: 23613896. Pubmed Central PMCID: 3632575.
16. Vaughan JT, Snyder C, Delabarre L, Tian J, Adriany G, Andersen P, et al. Clinical Imaging at 7T with a 16 Channel Whole Body Coil and 32 Receive Channels. In: Proceedings of the 17th scientific meeting, International Society for Magnetic Resonance in Medicine, Honolulu, p 392. 2009.
17. Orzada S, Bitz AK, Gratz M, Johst S, Völker MN, Kraff O, et al. An integrated 8-channel Tx/Rx body coil for 7 Tesla whole body MRI. in: Proceedings of the 23rd scientific meeting, International Society for Magnetic Resonance in Medicine, Toronto, p 630. 2015.
18. Andreychenko A, Bluemink JJ, Raaijmakers AJ, Lagendijk JJ, Luijten PR, van den Berg CA. Improved RF performance of travelling wave MR with a high permittivity dielectric lining of the bore. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 2013 Sep;70(3):885-94. PubMed PMID: 23044511.
19. Brunner DO, Paska J, Froehlich J, Pruessmann KP. Traveling-wave RF shimming and parallel MRI. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 2011 Jul;66(1):290-300. PubMed PMID: 21695729.