Synopsis
The native accuracy of
gradient and shim systems is too low in order to drive MRI sequences. To avoid
corresponding image artefacts the gradient chains are feed-backed, pre-distorted
and post-corrected based on accurate characterizations or direct measurements of
the field evolution in the scanner. In this talk, the underlying principles of
the encountered distortions and frequently applied correction methods will be
discussed.
Introduction
The spatio-temporal
dynamics of the magnetic field inside the scanner generates and encodes the NMR
signals. In order to interpret and reconstruct images from the acquired raw
signals the exact evolution of the magnetic field inside the scanner has to be accurately
known. But the employed hardware cannot natively provide the accuracy in the field
dynamics that is ideally required spatially and temporally. Spatially, the
applicable equations of magnetostatics set limits on realizable field patterns.
This leads to deviations of the field pattern generated by a gradient or shim
coil from the targeted ideal distribution. A gradient coil would ideally produce
a spatially linearly increasing field strength along a given axis. In reality
there will be a spatial non-linearity of the field but the resulting image
distortions can mostly be handled in post-processing by unwarping the images to
remove the geometrical distortions [1].
Realizing the desired temporal
evolution of the magnetic field in the scanner is more complex and furthermore
subject to very tight tolerances. Stray induction into conductive surfaces,
mechanical interactions and externally induced fields cause delays, distortions
and fluctuations that heavily impair the resulting image quality.
Deviations from the
intended field evolution corrupt the acquisition in various ways; errors in the
k-space trajectory of the read-out cause ghosting, warping and shading in the
reconstructed images. Equally unwanted field transients affect the slice
selection by exciting incorrect volumes leading to inconsistent data sets, loss
of steady-state conditions or signal strength.
Furthermore, currently
developed advanced acquisition methods heavily push the need for accuracy of
the sequence execution for speeding up the acquisition, deducing quantitative
information or by requiring a high consistency between several acquisitions for
later data fusion. Obviously deriving quantitative information requires a high
degree of fidelity from the system but equally highly efficient encoding
schemes turn out to be more susceptible to errors in the underlying encoding.
For example EPI, UTE and spiral acquisitions are very susceptible to minute
delays in the gradient encoding while magnitude images from normally employed
Cartesian single line spin-warp read-outs are very robust against gradient
delays. Similarly, fast sequences react very sensitive to steady-state
violations due to their long echo-trains.
Finally, the trend
towards higher field strength sets even higher needs for stronger and faster
switching gradients in order to provide on the one hand robustness against
susceptibility artefacts by fast high-bandwidth read-outs and on the other to
turn the increased baseline SNR into increased speed and resolution.
In order to obtain the
required accuracy in the field evolution, MRI scanners require therefore very
accurate compensation and correction schemes in order to overcome the remnant
shortcomings of the employed hardware. The basic principle behind this approach
is that calibration, compensation and correction schemes can turn the precision
and accuracy achieved in measurements of the magnetic field evolution into an
enhanced fidelity of the acquisition and reconstruction. This in turn implies
that the accuracy by which the field dynamics can be measured sets the ultimate
boundary for the achievable requirements set by the experiments. To achieve
this, three basic types of schemes are employed:
·
Feedforward
or pre-emphasis corrections pre-distort the requested waveform sent to the
input based on a system characterization such that the output fits the demand
best.
·
Feedback
approaches measure the deviations of the observable concurrently at the output
of the system and correct the input in real-time to compensate for the
deviations.
·
Post-correction
schemes correct the value of the obtained measurement in the acquired data for
the effect of known deviations present during the experiment.
Figure 1 depicts
the application of the three approaches on the example of a typical gradient
chain.
The main advantage of feedforward
and feedback approaches is that they improve the fidelity of the physical
execution of the experiment. The experiment can therefore be executed under
almost ideal conditions. This is important because not all deviations can be
post-corrected or inflict penalties in the conditioning or numerical complexity
of the reconstruction problem. Generally, the effect must be encoded in the
acquired data in order to be post-corrected out of the obtained data. For
example an accurate execution of slice selection gradients is required in order
to provide signal from the correct slice location, keep steady state conditions
and avoid signal loss by through-slice dephasing all of which can typically not
be post-corrected on the acquired data. For post-correcting errors of the
read-out trajectory it has to be ensured that the signal is sufficiently
encoded in order to render the reconstruction problem well-conditioned. There
is however a bigger leeway since some of the encoding can be deferred to
parallel imaging schemes. Finally, there are classes of acquisition schemes
that encode all properties of the spin system such as stochastic resonance
approaches [2] and MR fingerprinting [3]. In these cases, the acquisition can be
post-corrected even from heavily distorted sequence executions as long as the
randomization criteria and the bandwidth coverage remain fulfilled. However, it
is then required to encode the entire scope of parameters, which might come
with an overhead in the acquisition at least if some of the fitted values are
not used in the end.
The major downside of feedforward
and feedback approaches is that they require additional driving power to play
out the compensating waveform. This overhead has to be accounted for and limits
the maximum available slew rates and amplitudes. Furthermore, feedforward
approaches critically depend on the exact knowledge and stability of the system
behavior to pre-distort the input waveform to the system since it needs to
predict the field evolution based on the request. In contrast, feedback approaches
are insensitive to small changes of the system but require an accurate
concurrent measurement of the regulated entity with low delay and high
bandwidth in order to keep the system stable. Magnetic field measurements are therefore
typically not applied for feedback control of fast switching encoding gradients
but can be well applied for correction of slow field drifts [4]. However, gradient amplifiers are equipped
with current sensors regulating the output current. This feedback highly linearizes
and equalizes the gradient chain with respect to the current in the leads. But the
field evolution produced by the coil can deviate from that of the driving
current in its terminals requiring additional correction in order to
effectively ensure a precise image encoding.
As mentioned above the
major advantage of post-correction schemes is that they do not require
additional driving power and do typically not interfere with the execution of
the sequence but the conditioning of the reconstruction problem can worsen
increasing noise in the result. In turn, post correction schemes can correct
for field distortions where no actuation is available for compensation, e.g.
higher spatial orders or outside the bandwidth gradient and shim channels in
particular can operate in. Examples of such approaches that are regularly
applied are correction of gradient induced B0 fluctuations which are
typically implemented by shifting the frequency the acquired signals are
demodulated with and/or the phase of the acquisition as for instance in EPI
phase corrections [5]. Furthermore, the entire reconstruction can be
based on predicted [6] or even concurrently monitored k-space
trajectories [7] and even entirely monitored sequences [8] taking all field evolution into account.
In summary, feedforward,
feedback and post-processing corrections complete each other and are therefore
almost exclusively applied in combination.Encoding field generation and distortion effects
As shown in
Fig. 1 the required gradient fields are generated by first converting the
requested digital waveform into a low-power analogue signal, which is then sent
to the gradient power amplifier stage. The gradient amplifier then drives the
current through one gradient coil or a segment of it which produces the
required magnetic field. Since the temporal evolution of the gradient and shim
fields depend on the driven current waveform, very high fidelity waveforms have
to be generated in the first place using high dynamic range waveform
synthesizers. Due to the high power requirements, very power efficient
topologies have to be employed in the gradient amplifier. Such topologies do not
provide the needed accuracy and linearity. Therefore the gain stage is tightly
controlled by a feedback mechanism. A current sensor, typically of a flux-gate
type, dynamically measures the output of the amplifier and deviations are filtered
and fed back to the input. This linearizes the amplifier over a large dynamic
range and stabilizes it against various sources of drifts in the gain stage. Furthermore,
the feedback renders the driven current independent of the load impedance at
the output. This effectively compensates for the effect of the coil inductance
and resistance within the feedback bandwidth. As a consequence the current
driven through the terminals of the coils correspond very precisely to the
requested waveform.
By this feedback the
field evolution in the coil is controlled to the degree the field inside the
coil depends on the current through the terminal of the coil. There are however
effects that let the field generated by the coil deviate from the current
running through the terminals of the coil.
Primarily current
paths driven by induction caused by gradient switching do not run through the
terminals and produce additional field. By Lenz’s law, these currents oppose
the slewing of the field in the coil and therefore introduce a delay and slows
down the ramping of the field. Furthermore, the field pattern is distorted from
the field of these additional current paths. Currents induced in other gradient
or shim coils produce a secondary field with predominantly the spatial pattern
of the coupled channel and appear like an unwanted activation of that channel. These
couplings to other gradient or shim channels are therefore called cross terms.
However, ramping a
magnetic field in bulk conductors causes locally flowing electric currents, the
so called eddy currents. The quasi-stationary field ($$$B$$$) generated by
impressed and induced currents ($$$J$$$) in conductive materials ($$$\sigma,
\mu$$$) is governed by a diffusion equation:
$$\sigma \mu \frac{\partial
B}{\partial t}= \nabla^2 B=-\nabla \times J.$$
The equation links
the temporal dynamics with the material properties and its geometry which shows
that the eddy current effects and lifetime scale with the conductance of the
material and its size. The larger
the current path and the higher the bulk conductivity the longer the energy can
be stored in the eddy and consequently the longer its time constant. Although,
MRI scanners and the components present in the bore are designed to exhibit
minimal contiguous conductor surfaces, eddies in the conductors of gradient
coils, RF coils , RF screens, cryostat walls and cold structures critically
distort the field evolution and have orders of magnitude different time scales
and intensities. Eddies that produce the same field pattern as the generating
coil cause a delay and low passing of the gradient pulses. Eddies that produce
other field patterns by their current paths cause apparent cross-terms.
In addition to
induction of additional currents, the current carrying conductors are subject
to strong Lorentz’s forces. Thereby the coils are deformed and consequently change
their field pattern and efficiency. This is prominently visible at frequencies
of mechanical resonances of the gradient coil former. If such a resonance is
excited by the requested waveform the coil mechanically oscillates and produces
and oscillatory field response. Such electromechanical oscillations can have
very long decay times (several 10 to 100 ms) and occur typically above 1 kHz. Furthermore,
the frequency and quality factor of the mechanical resonances are found to be
highly dependent on the temperature of the gradient and shim coil due to the temperature
coefficients of the mechanical properties of the employed formers.
However, besides their
temperature dependence the above mentioned effects are typically governed by
linear equations. Although the field response of the coil is temporally and
spatially distorted the output remains a linear response of the current driven
through the coil.
But besides inductive
and mechanical couplings also thermal effects can cause field distortions. Particularly
thermal coupling from the gradient coil to passive shim irons can cause field
drifts. This coupling is not a linear response to the driven sequence. The
heating induced in the gradient coils depends on the root-mean-square current
but also on the driven slews but equally on the system cooling and ambient
factors. These effects are therefore very hard to characterize up-front and are
typically tackled by run-time measurements for instance by re-determining the NMR
center frequency of the sample or by use of external references [9] for monitoring the field evolution.
Finally, there can be
induced field dynamics in the scanner which is not driven by the scanner at
all. Such fields can originate from close by installations such as elevators,
moving cars, power lines, train tracks etc. But also the subject can induce
field variations by breathing or limb motion [10] which moves susceptible material in the bore. Correspondingly,
these effects cannot be calibrated nor predicted and need to be either
suppressed or post-corrected based on concurrent measurements of the field
evolution [11].Measurements
Irrespective if feedforward,
feedback or post-correction schemes are applied the field evolution has to be
measured in the first place. Based on these measurements the system can then be
characterized in order to be calibrated and corrected. Ideally, the magnetic field
evolution in the scanner is captured with high temporal and spectral resolution
and a spatial resolution that allows separating the different involved cross
terms. Basic calibration schemes however have only to determine the delay of
the individual channels and the most important decay times of the involved eddy
currents.
The latter can be very
efficiently implemented using NMR based measurements on phantoms [12-15]. Thereby simple blip encodings can be employed
to study the temporal alignment and phase evolution of the echoes representing
the individual delays of the gradients. Even entire k-space trajectories and
higher spatial orders can be acquired
using spatial-spectrally encoding the phase evolution driven by the
trajectory [16]. The main advantage of these methods is that
they do not require additional hardware besides the phantom. However, for
resolving spatial-spectral field dynamics the methods have to acquire several
shots with very high consistently and the additionally required additional gradient
lobes must not affect the measurement. Therefore, the MRI system has to be
stable and repeatable over all these shots and fluctuations and drifts can
often not be quantified and hamper the calibration. Furthermore, they require a
very well shimmed setup to characterize long term effects and also the gradient
system has to operate accurately enough to perform correct excitations and
encodings of the trajectory. This poses a certain bootstrapping problem in
particular at installation of the scanner. Typically this is circumvented by
first shimming the magnet using an external field mapping device.
Alternatively external
field measurement devices can be employed to measure the gradient dynamics. A
comparably simple method is based on pick-up loops which sensitize the
induction caused by the gradient ramping [17]. While they can very well detect channel
delays, they often do not deliver enough sensitivity to measure long term eddy
currents, field drifts or to map an entire k-space trajectory. This is because
they measure the induction and in order to determine the phase accrual induced
to the spins their signal has to temporally integrated twice, which has very
unfavorable noise propagation properties.
Recently NMR based
field probe systems [7, 11, 18] are increasingly used. The NMR signal offers a
very high sensitivity to detect the relevant magnetic field component as well as
its time integral even with high bandwidth. By employing an array of such
sensors the spatial-temporal field evolution can be measured in a single shot
and even during the actual experiment. Therefore, transient effects, drifts and
even subject induced effects can be acquired. Furthermore, the actual field
evolution during the experiment can be captured and applied in the post
processing step. On the downside the method requires additional involved
hardware.Synchronization
When interpreting the
calibration data acquired with the MR scanner as well as with external devices
the problem of synchronicity has to be concerned. While the NMR sequence is
acting on the spins and the spins produce their signal in the physical time
frame, the time frames the MRI scanner’s sequence programs are operating in is
typically a different digital time raster (see Fig. 2). Since the
different subsystems such as RF transmit, receive, gradient and shim chains
exhibit different delays relative to the physical time frame there can be discrepancies
between the different time rasters. A particular problem is that the delay of
the RF chains can be critically dependent on the loading condition of the RF
coils. Furthermore, the physical time frame is only observable via the receive
chain. Therefore the system is typically calibrated to a roughly compensated
discrete receiver time raster. In turn, a system which has an apparently
synchronized gradient and receiver system can still exhibit systematic delays
between the transmitter and the gradient chain as well as the receiver. Gradient system characterization
The goal of a system
characterization is to predict the output of the system dependent on its given inputs.
This knowledge can then be used for calibration, feedback adjustments,
post-correction and sequence optimization. Fields induced by sources not under
control of the scanner or that are not dependent on accessible variables cannot
be considered for a calibration and represent therefore a constant source of
potential confounders of the calibrated system’s output but also for the
calibration procedure itself. However for calibration purposes the MRI system has
to be considered as time invariant and it remains to other means that other
confounders are minimal.
Additionally, the
gradient and shim systems can be to a very large degree considered as linear
systems, i.e. the field output is considered linearly dependent on the demand
sent to the system [19]. Ensuring this linearity however is a critical
effort in the hardware design and is ensured by very involved feedback driving
circuits as mentioned above.
As a consequence of
the linearity and time invariance, the field output ($$$B(r,t) = \mathcal{F}^{-1}(b(r,\omega))$$$) is
given as convolution of the demanded waveform ($$$ D(t) = \mathcal{F}^{-1}(d(\omega))$$$)
and the impulse response function ($$$IR(r,t) = \mathcal{F}^{-1}(H(r,\omega))$$$) of the system:
$$B(r,t) = D(t) \otimes
IR(r,t),$$
or analogously in frequency domain:
$$b(r,\omega) =
d(\omega) \cdot H(r,\omega).$$
The calibration
sequences characterizing the system have therefore to provide the data basis to
determine the transfer function within the required bandwidth. This is achieved
by recording the field output for a set of known inputs which cover the required
frequency span. The transfer function is then found by a deconvolution of the
output with the input signal:
$$H(r,\omega) = \frac {b(r,\omega)}{d(\omega)}.$$
In principle various test functions
($$$d(\omega)$$$ can
be employed as input such as blips or sweeps as long as the system is not
driven to substantial non-linearity arising from amplitude or slew-rate
limitations. Figure 3 shows an example for the response of a y gradient coil to
a trapezoidal blip.
The impulse response and equally the transfer functions contain all the
information on delay, eddy currents oscillations etc. as depicted in Fig. 4.
As seen, the delay causes the impulse response function of the native system to
peak after the demand and the transfer function exhibits a corresponding phase
gradient. The eddy current terms low pass the response which causes the
transfer function to be below the requested amplitude at higher frequencies.
Correspondingly, the calculated step response creeps slowly to the targeted
value. Finally, the oscillations visible in the transfer function as clear dips
cause the field fluctuate around the target value ones excited by a step in the
targeted field.
In order to
characterize the spatial behavior of the dynamic fields it is recommendable to
decompose the spatial degrees of freedom into a set of spatial basis functions
$$$B_i(r), i = 0, …n$$$:
$$B(r,t)=\sum_{i=0…n}
B_i(r)\cdot J_i(t),$$
$$b(r,t)=\sum_{i=0…n}B_i(r)\cdot
j_i(\omega).$$
The components
$$$J_i(t)=\mathcal{F}^{-1}(j_i(\omega))$$$ represent
the time evolution of the field pattern. Choosing the spatial field pattern of the individual gradient and shim
coils as such a spatial basis is often very insightful since the coupling terms
are directly visible in the activation of the corresponding components
($$$J_i(t)$$$). Consequently, the impulse responses and the transfer function can
be decomposed in the same spatial basis and with respect to every possible
input waveform ($$$D_p(t)$$$) which yields:
$$B(r,t) =
\sum_{i=0…n} B_i(r) \cdot J_i(t) = \sum_{i=0…n, p=0…n} B_i(r) D_p(t) \otimes
IR_{i,p}(t)$$,
$$b(r,t)= \sum_{i=0…n, p=0…n} B_i(r) d_p(\omega) \cdot H_{i,p}(\omega).$$
The system’s
response is then characterized by a matrix of transfer/response functions. The
response of the channel to itself is characterized in the $$$ IR_{Iii}(t)$$$
functions, the cross-term couplings are represented in the off-diagonal
functions.Feedforward compensation
The basic idea of feedforward
compensations is to pre-distort the requested waveform such that the field
evolution will fit the request within the targeted bandwidth. For instance a
delay can be compensated by playing out the waveform earlier such that it will
arrive in time. Each frequency component is scaled and phased based on its
characterized response such that it will be equalized at the output, which is
called pre-emphasis. Since eddy currents and other band-limiting effects lower
the efficiency of the gradient system at higher frequency, these components
have to be scaled up in the demand sent to the amplifier, which results in a
required overdrive in the pre-distorted waveform. The amplifier has
consequently to produce higher output voltages or currents in order to speed up
the ramping of the fields and to suppress cross terms. This in turn limits
effective slewing and peak values of fields that can be requested when
designing the not pre-emphasized waveform. Therefore it is important to limit
the targeted response of each channel to the minimum required bandwidth by
using a windowing function ($$$W(\omega)$$$)
attenuating the request for high frequency components. Then the pre-distorted
waveform ($$$D^C(t) = \mathcal{F}^{-1}(d^C(\omega)$$$) can be calculated by a
deconvolution operation as also
depicted in Fig. 4:
$$d^C(\omega) =
\frac {d(\omega) \cdot W(\omega)}{H(\omega)} = d(\omega) \cdot H^+(\omega),$$
$$D^C(t) = D(t)
\otimes IR^+(t),$$
where
$$$H^+(\omega)=\frac{W(\omega)}{H(\omega)}$$$, and $$$IR^+(t) =
\mathcal{F}^{-1} (H^+)$$$.
Analogously cross-terms
can be suppressed by inverting the transfer function matrix for each frequency
(see Fig. 5):
$$H^+(\omega) _{i,p} =
W(\omega) \left ( H(\omega)^ {-1} \right
)_{i,p} \forall \omega.$$
Also here other
channels have to allocate overhead power in order to compensate for the
coupling. Therefore the inversion of the transfer matrix has to be carefully
regularized by use of an appropriate pseudo-inverse ensuring a well-conditioned
inverse at every frequency.
In many cases the
calculation of the pre-distorted waveforms is simplified and numerically
optimized by translation to time domain filters. Due to the involved long integration
times this is often implement as infinite impulse response filters. For this,
the impulse response of the system has to be decomposed into the individual
elements of the filter.
A further major
consideration of feed-forward approaches is that the convolution kernel
obtained for the pre-emphasis is not causal. This means that it has to be known
in advance what waveform will be required in order to send out the signal in
time. This is for typical sequence execution no problem since the waveforms are
known well ahead of time. However, it can become a consideration limiting the
reaction time in real-time applications for triggering, prospective motion
correction or field feedback.Feedback compensation
Feedback approaches
offer the advantage that the behavior of the native system can vary to a
certain degree maintaining the overall output if the measurement of the
feedback variable is not affected, the system remains stable and does not
overshoot out of the regulation. However, in order keep the feedback stable,
the output value has to be measured with very low delay compared to the
bandwidth the feedback is supposed to operate in. Therefore, feedback of fast gradient
waveforms is typically only applied to the output current of the amplifier,
which can be measured very fast and with high accuracy. This linearizes the
gradient amplifier and compensates drifts and gain changes of its output stage.
Furthermore, the low-pass frequency response of the gradient coil with regards
to its input current is largely compensated by this feedback circuit. Additionally,
the feedback loop of the amplifier produces a very high impedance at its output
which largely hinders net currents to flow due to induction from other
channels. However, as mentioned above the controlled value is the current
present at the terminals of the amplifier, whose dynamics can differ from that
of the field inside the coil. Nevertheless, low frequency drifts can be well
compensated using either short frequency navigator acquisitions or independent
field monitoring devices [4]. Since many of these drift effects are not
captured by the feedforward calibration of the system, the feedforward and
feedback approaches complement each other very well. Post-correction
Post-correction of the
acquired data compensating for individual effects of the gradient chain
distortion is widely implemented at different stages of the signal processing on
current MRI scanners. E.g. the NMR frequency shifts induced cross-terms to the
uniform field for instance caused by long lived eddy currents as generated by
diffusion sensitizing gradients is demodulated out of the acquired data by the eddy
current correction system. The phase inconsistency of odd- and even-echo tops
is typically corrected for in EPI scans [5] by shifting the phase of the acquired signals.
Gradient chain delays are often compensated for in reconstruction UTE sequences
etc. However, in principle the full system characterization can be employed to
predict the field evolution during a scan in order to obtain the k-space
trajectory of given scan [6]. Furthermore, concurrent field measurements
can be employed [7, 20, 21] to obtain the effective gradient encoding of a
scan as it was present during the acquisition. These measurements then not only
include the predictable field evolution based on the demanded sequence but also
externally, temperature and subject induced effects which can then be handled
by the reconstruction.Acknowledgements
No acknowledgement found.References
1. F. Schmitt, Correction of Geometrical Distortions in
MR-images, in Computer Assisted
Radiology / Computergestützte Radiologie, H. Lemke, et al., Editors. 1985,
Springer Berlin Heidelberg. p. 15-23.
2. R.R. Ernst, Magnetic resonance with stochastic excitation. Journal of Magnetic
Resonance (1969), 1970. 3(1): p.
10-27 DOI: http://dx.doi.org/10.1016/0022-2364(70)90004-1.
3. D. Ma, V. Gulani, N. Seiberlich, K.
Liu, J.L. Sunshine, J.L. Duerk, and M.A. Griswold, Magnetic resonance fingerprinting. Nature, 2013. 495(7440): p. 187-192 DOI: http://www.nature.com/nature/journal/v495/n7440/abs/nature11971.html#supplementary-information.
4. Y. Duerst, B.J. Wilm, B.E. Dietrich,
S.J. Vannesjo, C. Barmet, T. Schmid, D.O. Brunner, and K.P. Pruessmann, Real-time feedback for spatiotemporal field
stabilization in MR systems. Magn Reson Med: p. n/a-n/a.
5. M.H. Buonocore and L. Gao, Ghost artifact reduction for echo planar
imaging using image phase correction. Magn Reson Med, 1997. 38(1): p. 89-100 DOI:
10.1002/mrm.1910380114.
6. S.J. Vannesjo, N.N. Graedel, L.
Kasper, S. Gross, J. Busch, M. Haeberlin, C. Barmet, and K.P. Pruessmann, Image reconstruction using a gradient
impulse response model for trajectory prediction. Magn Reson Med, 2015: p.
n/a-n/a DOI: 10.1002/mrm.25841.
7. C. Barmet, N.D. Zanche, and K.P.
Pruessmann, Spatiotemporal magnetic field
monitoring for MR. Magn Reson Med, 2008. 60(1): p. 187-197 DOI: 10.1002/mrm.21603.
8. D.O. Brunner, B.E. Dietrich, M.
Çavusoglu, B.J. Wilm, T. Schmid, S. Gross, C. Barmet, and K.P. Pruessmann, Concurrent recording of RF pulses and
gradient fields – comprehensive field monitoring for MRI. NMR Biomed, 2015:
p. n/a-n/a DOI: 10.1002/nbm.3359.
9. B.J. Wilm, Y. Duerst, B.E. Dietrich,
M. Wyss, S.J. Vannesjo, T. Schmid, D.O. Brunner, C. Barmet, and K.P.
Pruessmann, Feedback field control
improves linewidths in in vivo magnetic resonance spectroscopy. Magn Reson
Med, 2014. 71(5): p. 1657-1662 DOI:
10.1002/mrm.24836.
10. S.J. Vannesjo, B.J. Wilm, Y. Duerst,
S. Gross, D.O. Brunner, B.E. Dietrich, T. Schmid, C. Barmet, and K.P.
Pruessmann, Retrospective correction of
physiological field fluctuations in high-field brain MRI using concurrent field
monitoring. Magn Reson Med, 2014. 73(5):
p. 1833-1843 DOI: 10.1002/mrm.25303.
11. C. Barmet, N. De Zanche, B.J. Wilm,
and K.P. Pruessmann, A transmit/receive
system for magnetic field monitoring of in vivo MRI. Magn Reson Med, 2009. 62(1): p. 269-276 DOI:
10.1002/mrm.21996.
12. N.P. Davies and P. Jezzard, Calibration of gradient propagation delays
for accurate two-dimensional radiofrequency pulses. Magn Reson Med, 2005. 53(1): p. 231-236 DOI:
10.1002/mrm.20308.
13. P. Jezzard, A.S. Barnett, and C.
Pierpaoli, Characterization of and
correction for eddy current artifacts in echo planar diffusion imaging.
Magn Reson Med, 1998. 39(5): p.
801-812 DOI: 10.1002/mrm.1910390518.
14. G.F. Mason, T. Harshbarger, H.P.
Hetherington, Y. Zhang, G.M. Pohost, and D.B. Twieg, A Method to measure arbitrary k-space trajectories for rapid MR
imaging. Magn Reson Med, 1997. 38(3):
p. 492-496 DOI: 10.1002/mrm.1910380318.
15. V.J. Schmithorst and B.J. Dardzinski, Automatic gradient preemphasis adjustment: A
15-minute journey to improved diffusion-weighted echo-planar imaging. Magn
Reson Med, 2002. 47(1): p. 208-212
DOI: 10.1002/mrm.10022.
16. J.H. Duyn, Y. Yang, J.A. Frank, and
J.W. van der Veen, Simple Correction
Method for k-Space Trajectory Deviations in MRI. Magn Reson, 1998. 132(1): p. 150-153 DOI: http://dx.doi.org/10.1006/jmre.1998.1396.
17. M.A. Morich, D.A. Lampman, W.R.
Dannels, and F.T.D. Goldie, Exact
temporal eddy current compensation in magnetic resonance imaging systems.
IEEE Transactions on Medical Imaging, 1988. 7(3): p. 247-254 DOI: 10.1109/42.7789.
18. B. Dietrich, D.O. Brunner, B.J. Wilm,
C. Barmet, S. Gross, L. Kasper, M. Haeberlin, T. Schmid, S.J. Vannesjo, and
K.P. Pruessmann, A field camera for MR
sequence monitoring and system analysis. Magn Reson Med, 2015 DOI:
10.1002/mrm.25770.
19. S.J. Vannesjo, M. Haeberlin, L.
Kasper, M. Pavan, B.J. Wilm, C. Barmet, and K.P. Pruessmann, Gradient system characterization by impulse
response measurements with a dynamic field camera. Magn Reson Med, 2012. 69(2): p. 583-593.
20. B.J. Wilm, C. Barmet, S. Gross, L.
Kasper, J. Vannesjo, M. Haeberlin, B. Dietrich, D.O. Brunner, T. Schmid, and
K.P. Pruessmann. High-Resolution
Single-Shot Spiral Imaging using Magnetic Field Monitoring and its Application
to Diffusion Weighted MRI. in Proc
Intl Soc Magn Reson Med. 2015. Toronto, Canada. p. 955.
21. B.J. Wilm, C. Barmet, M. Pavan, and
K.P. Pruessmann, Higher order
reconstruction for MRI in the presence of spatiotemporal field perturbations.
Magn Reson Med, 2011. 65(6): p.
1690-1701 DOI: 10.1002/mrm.22767.