Synopsis
This lecture will describe some of the safety
and image artifact issues related to performing MRI in patients with an
implantable device (e.g., pacemaker, ICD, CRT).
Target audience
Basic science and
clinical imagers involved with cardiovascular MRI, particularly those with interests in heart failure and cardiac rhythm disorders.Objectives: Upon completion of this educational presentation, the participants should be able to:
-
Understand some of the safety aspects for performing
MRI safely in patients with an implantable defibrillator
- Understand some of the challenges for obtaining
high image quality from patients with an implantable defibrillator
Purpose/Introduction:
Heart
failure (HF) is a major healthcare problem in the United States (1) and worldwide. Both echocardiography
and MRI play an important role in diagnosis,
prognosis and risk stratification of patients with HF (2). MRI is arguably more versatile and
capable of a comprehensive evaluation, including function of both left and
right ventricles, perfusion, edema, and scar/fibrosis. Despite these
advantages, MRI’s utility may be limited in many HF patients with left
ventricular ejection ≤ 35% who also have a prophylactic implantable
defibrillator. One could argue that these HF patients tend to be sicker and
thus are the ones who would derive more benefit from MRI, since they are more
likely to have structural heart disease that could be characterized with MRI. Despite
the fact that MRI can be performed safely in most patients with a defibrillator
at 1.5T (3-6), many patients who would derive benefit from MRI do not
undergo MRI, largely due to image artifacts arising from the transformer of a
device (e.g., pacemaker, implantable cardioverter
defibrillator [ICD], cardiac
resynchronization therapy implantable cardioverter defibrillator [CRT-D]).
Clinical Indications:
It is estimated that approximately 50% of
patients with a defibrillator will be indicated for a clinical MRI exam during
the lifetime of the device. Indications for MRI in patients with an implantable
device: brain (40%), spine (22%), heart (16%), abdomen or pelvis (13%), and
extremity (9%).Safety Concerns:
- Force
and torque: The amount of ferromagnetic materials on a
device is minimal, thus this is not considered to be a safety issue at 1.5T (7,
8).
- Device
Functionality: Changes in device settings (e.g., reed-switch)
may occur due to static magnetic field (B0), leading to asynchronous pacing in
pacemakers and inhibition of shock delivery in ICDs. These need to be monitored
carefully (4).
- Gradient
Fields: Switching of magnetic field gradient could in theory induce currents
to produce direct stimulation of the heart (9).
- Radiofrequency
Fields: RF induced intracardiac lead tip heating has been demonstrated in
vitro (10) and in vivo (11).
- These safety concerns can be mitigated – but not
eliminated - by following an established device protocol at 1.5T (5, 6).
Image Artifact Concerns:
Both
intracardiac leads and the transformer of a device act as sources of image
artifacts. Intracardiac leads typically appear as signal void, but for the most
part produce benign image artifacts. The transformer, on the other hand, causes
significant center frequency shift (~kHz range) and generates significant image
artifacts and signal dephasing.
- Standard
cardiac pulse sequences using b-SSFP readout are sensitive to off-resonance
effects, so they do not work in patients with a device.
- FLASH
based sequences are able to suppress image artifacts, but two factors need to be
considered, including geometric distortion and dephasing.
- Spin-echo
based pulse sequences are able to suppress image artifacts and achieve
insensitivity to dephasing, but one must be careful to use low RF deposition.
- Wideband
cardiac pulse sequences have been introduced recently to suppress image
artifacts for these applications: LGE (12, 13), T1 mapping (14, 15), and perfusion (16).
Results:
Figure
1 shows representative T1 maps, perfusion, and LGE images of a
patient with a subcutaneous ICD (S-ICD), which
is considerably larger than both ICD and CRT-D. Compared with standard
acquisitions, wideband acquisitions suppressed image artifacts induced by S-ICD.Discussion:
As more academic centers become comfortable with
performing cardiovascular MRI safely in patients with an implantable
defibrillator, more investigations are needed to improve image quality and
identify clinical utility in HF and cardiac rhythm disorders.Acknowledgements
No acknowledgement found.References
-
Heidenreich PA,
Trogdon JG, Khavjou OA, et al. Forecasting
the future of cardiovascular disease in the United States: a policy statement
from the American Heart Association. Circulation. 2011;123(8):933-44.
- Slaughter RE, Mottram PM. What Should Be the Principle Imaging
Test in Heart Failure—CMR or Echocardiography? JACC: Cardiovascular Imaging.
2010;3(7):776-82.
- Nazarian S, Halperin HR. How to perform magnetic resonance
imaging on patients with implantable cardiac arrhythmia devices. Heart Rhythm.
2009;6(1):138-43.
- Nazarian S, Hansford R, Roguin A, et
al. A prospective evaluation of a
protocol for magnetic resonance imaging of patients with implanted cardiac
devices. Ann Intern Med. 2011;155(7):415-24.
- Nazarian S, Roguin A, Zviman MM, et al. Clinical utility and safety of a
protocol for noncardiac and cardiac magnetic resonance imaging of patients with
permanent pacemakers and implantable-cardioverter defibrillators at 1.5 tesla.
Circulation. 2006;114(12):1277-84.
- Sommer T, Naehle CP, Yang A, et al. Strategy for safe performance of
extrathoracic magnetic resonance imaging at 1.5 tesla in the presence of
cardiac pacemakers in non-pacemaker-dependent patients: a prospective study
with 115 examinations. Circulation. 2006;114(12):1285-92.
- Schmiedel A, Hackenbroch M, Yang A, et
al. [Magnetic resonance imaging of
the brain in patients with cardiac pacemakers. Experimental and clinical
investigations at 1.5 Tesla]. Rofo. 2005;177(5):731-44.
- Luechinger R, Duru F, Scheidegger MB,
et al. Force and torque effects of a
1.5-Tesla MRI scanner on cardiac pacemakers and ICDs. Pacing Clin
Electrophysiol. 2001;24(2):199-205.
- Levine GN, Gomes AS, Arai AE, et al. Safety of magnetic resonance imaging
in patients with cardiovascular devices: an American Heart Association
scientific statement from the Committee on Diagnostic and Interventional
Cardiac Catheterization, Council on Clinical Cardiology, and the Council on
Cardiovascular Radiology and Intervention: endorsed by the American College of
Cardiology Foundation, the North American Society for Cardiac Imaging, and the
Society for Cardiovascular Magnetic Resonance. Circulation.
2007;116(24):2878-91.
- Achenbach S,
Moshage W, Diem B, et al. Effects of
magnetic resonance imaging on cardiac pacemakers and electrodes. Am Heart J.
1997;134(3):467-73.
- Luechinger R,
Zeijlemaker VA, Pedersen EM, et al. In
vivo heating of pacemaker leads during magnetic resonance imaging. Eur Heart J.
2005;26(4):376-83; discussion 25-7.
- Rashid S,
Rapacchi S, Vaseghi M, et al. Improved
late gadolinium enhancement MR imaging for patients with implanted cardiac
devices. Radiology. 2014;270(1):269-74.
- Ranjan R,
McGann CJ, Jeong EK, et al. Wideband
late gadolinium enhanced magnetic resonance imaging for imaging myocardial scar
without image artefacts induced by implantable cardioverter-defibrillator: a
feasibility study at 3 T. Europace. 2015;17(3):483-8.
- Hong K, Jeong
EK, Kim D. Wideband
arrhythmia-insensitive-rapid (AIR) cardiac T1 mapping pulse sequence for
suppressing image artifacts induced by ICD. In: Proceedings of the 23rd Annual
Meeting of ISMRM, Toronto, Ontario, Canada 2015. Abstract No. 6747.
- Shao J, Rashid
S, Renella P, et al. Myocardial T1
mapping for patients with implanted cardiac devices using wideband inversion
recovery spoiled gradient echo readout. Magn Reson Med. 2016.
- Kim D, Lee D,
Wilcox J, et al. Wideband cardiovascular MRI for imaging patients with
intracardiac device implantation. 103rd Annual Meeting at RSNA. Abstract No.
16018349.