Synopsis
The spin echo pulse sequence is one of the most important
pulse sequences in MRI. Fast spin echo imaging is routinely used due to its
robustness to susceptibility variations and local field inhomogeneities, as
well as for its ability to produce excellent T1, T2 and PD images. The aim of
this lecture is to describe the basic physical principles governing spin echo
imaging and to illustrate the effect of key imaging parameters, such as TE, TR
and ETL (echo train length) on image contrast.
Aim:
The aim of this lecture is to illustrate the basic physics underlying
spin echo (SE) and fast spin echo (FSE) imaging. Physical principles will be
explained in simple terms, favoring intuition over mathematical formulations, and
several examples, from common clinical applications will be presented to directly
illustrate the effect of changing important imaging parameters, such as TE, TR
and ETL (echo train length) on image quality, image contrast and scan time. Specific objectives:
-
Understand
how spin echoes are formed and how they can be used to generate different image
contrasts. SE pulse sequences always start with a 90 degrees excitation
pulse that tips the magnetization in the transverse plane. Static magnetic
field inhomogeneities cause spins to dephase, resulting in the transverse
magnetization to decay with time constant T2*. A 180 degree RF pulse applied at
time TE/2, where TE is the desired echo time, effectively reverses this
dephasing process due to static magnetic field inhomogeneities, until all the
spins will come back into phase at t=TE, when a spin echo is formed. It’s
important to understand that while the dephasing due to static magnetic field
inhomogeneities is rephased at the time of the spin echo, the effect of
non-static field inhomogeneities due to spin-spin interactions are not
rephased, which results in T2 weighting of the detected signal. Figure 1 shows
a simple spin echo pulse sequence.
- Understand
strengths and pitfalls of SE-based pulse sequences and how they compare to
gradient-echo-based techniques. From the simple description above follows
that SE pulse sequences are insensitive to magnetic field inhomogeneities. This
is one of the definite advantages of SE imaging. However, several spin echoes
are necessary to form an image as each echo corresponds to a single line in
k-space and the pulse sequence in Figure 1 needs to be repeated for each phase
encoding step. Depending on the desired contrast, a relatively long time might
be needed for the longitudinal magnetization to recover (according to T1
relaxation) before the next SE experiment can start. This is why SE
acquisitions, unlike gradient-echo-based ones, tend to be quite long. Other important
differences between SE- and gradient-echo imaging, including flow-related
effects, RF energy deposition and TE/TRs achievable with each technique, will
be discussed during this lecture.
- Understand
the basic physics underlying fast spin echo imaging and some of the most common
clinical applications. In a Fast Spin Echo (FSE) pulse sequence, the
excitation pulse is followed by a train of refocusing pulses, so that multiple
spin echoes are generated. Each echo is phase encoded differently, so that
several lines of k-space are acquired following each excitation pulse, significantly
reducing scan time. FSE, also known as Turbo Spin Echo (TSE) or RARE (Rapid
Acquisition with Relaxation Enhancement), is one of the most commonly used
pulse sequences, because of its robustness to susceptibility variations and
field inhomogeneities, as well as for the diverse image contrasts that can be generated.
Figure 2 shows a simplified FSE pulse sequence with ETL=5. Note that in this
case the effective echo time (Teff) is defined as the time when the center of
k-space, which contains most of the energy, is acquired.
- Learn to
recognize common artifacts encountered in fast spin echo imaging and simple mitigation
strategies to improve image quality. FSE imaging techniques have replaced
conventional SE protocols in many applications, due to the shorter duration of
the acquisition. However, long echo trains also mean increased blurring due to
T2-induced decay along the echo train and increased RF power deposition. Figure
2 shows the typical blurring encountered in long-echo-train FSE with respect to
SE imaging. Differences in contrast between SE and FSE due to magnetization
transfer effects and other mechanisms will be briefly discussed.
Summary:
By the end of this lecture, participants should understand
the basic physical principles underlying SE imaging. In particular, they should
appreciate the role played by different imaging parameters to obtain the
desired image contrast as well as to mitigate common artifacts encountered in SE
and FSE imaging.
Acknowledgements
No acknowledgement found.References
[1] Jung BA , Weigel M, Spin echo magnetic resonance
imaging. JMRI 2013, 37:805-817. This is a brief, excellent review of the
material that will be covered during this lecture.
[2] Haacke EM,Brown RW,Thompson MR,Venkatesan R, Magnetic
resonance imaging: physical principles and sequence design. New York: John
Wiley and Sons; 1999.
Classical book on MRI physics.