Liangjie Lin1,2,3, Yanqin Lin1, Dan Tian1, Hongyi Yang4, Zhiliang Wei2,3, Peter B. Barker2,3, Kai Zhong4, and Zhong Chen1
1Department of Electronic Sciense, Xiamen University, Xiamen, People's Republic of China, 2Russell H. Morgan Department of Radiology and Radiological Science, Johns Hopkins University, Baltimore, MD, United States, 3F. M. Kirby Center for Functional Brain Imaging, Kennedy Krieger Institute, Baltimore, MD, United States, 4High Magnetic Field Laboratory, CAS Center for Excellence in Brain Science, Chinese Academy of Sciences, Hefei, People's Republic of China
Synopsis
Conventional localized 1H MRS pulse
sequences, such as PRESS and STEAM, generally suffer from J coupling modulations which can aggravate attenuation of multiplet
resonances during echo times. Here, the “perfect echo” module combined with an optimized
localization scheme is utilized for in-phase single-voxel in vivo MRS at 9.4 T.
The relative signal intensities of multiplet to singlet resonances acquired at
short and moderate echo times increase substantially in comparison with those at
PRESS spectra. Therefore, direct MRS quantification of many important
metabolites, such as glutamine, glutamate, γ-aminobutyrate, aspartate, and
myo-inositol, may be improved.
Purpose
MR spectroscopy, as a noninvasive technique for
in vivo metabolic and biochemical detections, has now been widely utilized for
preclinical and clinical studies. Currently, Point RESolved Spectroscopy
sequence (PRESS) and Stimulated Echo Acquisition Mode (STEAM) are the two
commonly used single voxel 1H MRS pulse sequences. However, they both
suffer from scalar coupling (J) modulations during echo times, thus hampering direct quantification of
multiplet resonances. In the Carr-Purcell-Meiboom-Gill (CPMG) experiments1, J modulations within echo time can be
suppressed through repeating 180° pulses with very short interpulse intervals (τ
≪ 1/∆ϑ, ∆ϑ is the
chemical-shift difference between coupled resonances). Additionally, the
“perfect echo”2, achieved by placing a 90° pulse in the midpoint of two spin echoes, can also refocus evolution of scalar coupling with longer interpulse intervals (τ ≪ 1/J).
Therefore, the radio frequency (RF) power deposition can be greatly reduced by
the “perfect echo” in comparison with CPMG under identical echo times. The “perfect
echo” has been previously utilized to the acquisition of magnetic resonance
spectroscopic imaging3. In this study, the “perfect echo” module combined
with an optimized localization scheme is introduced to achieve In-Phase Point
RESolved Spectroscopy (IP-PRESS) for in vivo detection.Methods
The proposed pulse sequences (IP-PRESS), modified
from conventional PRESS sequence, are shown in Fig. 1. Commonly, J values within metabolites are smaller
than 15 Hz. As a result, the achievable maximum TE in the single echo
IP-PRESS (Fig. 1(b)) can be about 30 ms. For longer echo times, the “perfect
echo” need to be repeated for satisfactory suppression of J modulation, and the double echo IP-PRESS is presented in
Fig. 1(c) for example. A healthy rat was scanned to demonstrate the in vivo
performance of IP-PRESS. The voxel size was 5.0×5.0×7.0 mm3. Average
number were 128 for PRESS and 256 for IP-PRESS with TR = 2.5 s. For PRESS, the
minimum TE (13 ms) was achieved by using TE1 = 7 ms and TE2
= 6 ms. Other TEs were achieved by setting TE1 = 10 ms, with TE2 values of 10, 20, 30, and 50 ms, respectively. For IP-PRESS, short TEs (14, 20,
and 30 ms) were achieved by single echo IP-PRESS, and moderate TEs (40, and 60
ms) were achieved by double echo IP-PRESS. Experiments were performed on 9.4
T/400 mm animal scanners (Agilent Technologies, Inc., Santa Clara, CA, USA),
using volume RF coils. The variable power and optimized relaxation delays
(VAPOR) module provided by vendor was used to suppress the strong water signal.Results and discussion
Figure 2 presents PRESS and IP-PRESS rat brain
spectra acquired at 9.4 T. The blue and green shaded regions in Fig. 2(a)
mainly comprise of multiplet resonances. Due to the
evolution of scalar coupling, intensities of these multiplet resonances in
PRESS spectra suffer from significant reduction even at the shortest echo time
(13 ms), and decay nearly to zero at 60 ms. By refocusing the scalar coupling
evolution, relative signal intensities of multiplet resonances to those of
singlet resonances are greatly improved in IP-PRESS spectra and the multiplet
signals can still be recognizable at 60 ms. Therefore, recognition and analysis
of multiplet signals can be more feasible in IP-PRESS spectra. The signal
intensity and baseline performance of IP-PRESS spectrum at 30 ms are favorable,
and the metabolite assignment on this spectrum is shown in Fig. 2(b). A
limitation of IP-PRESS pulse sequences lies in that only half of the magnetization
from designated voxel can be retained. Additionally, the RF power deposition will also increase in either single or multi echo IP-PRESS in comparison with PRESS.Conclusion
We have developed a new pulse sequence for
in-phase single-voxel MRS. In vivo rate brain spectra acquired at short and
moderate echo times at 9.4 T demonstrate that relative intensities of scalar
coupling resonances to singlet resonances are significantly enhanced. Therefore,
the proposed pulse sequence may provide an approach for better quantification
and fitting of scalar coupled peaks for proton MRS in brain.
Acknowledgements
This work was partially supported by the
National Natural Science Foundation of China under Grants 11105114 and
11375147, and the Natural Science Foundation of Fujian Province of China under
Grant 2014J05012, and the State Scholarship Fund of China (No. 201606310173).References
1.
Hennig J, Thiel T, Speck O. Improved sensitivity to overlapping multiplet
signals in in vivo proton spectroscopy using a multiecho volume selective
(CPRESS) experiment. Magn Reson Med 1997;37(6):816-820.
2.
Lin YQ, Lin LJ, Wei ZL, Zhong JH, Chen Z. Localized one-dimensional single voxel
magnetic resonance spectroscopy without J
coupling modulations. Magn Reson Med DOI 10.1002/mrm.26066.
3.
Pan J.W., Avdievich N., Hetherington H.P. J-refocused
coherence transfer spectroscopic imaging at 7 T in human brain. Magn Reson Med 2010;
64(5):1237–1246.