Gokce Hale Hatay1, Muhammed Yildirim1,2, and Esin Ozturk-Isik1
1Institute of Biomedical Engineering, Bogazici University, Istanbul, Turkey, 2Advanced Diagnostic Imaging, Philips Healthcare, Best, Netherlands
Synopsis
This study aims at investigating the effects of compressed sensing
data acquisition and reconstruction factors for accelerated phosphorus MR
spectroscopic imaging (31P-MRSI). Simulated 31P MRSI
datasets containing healthy and tumor regions were created based on the
metabolite information of brain tumor patient 31P-MRSI acquired at
3T. k-space data were randomly undersampled with three different reduction
factors while preserving the central portion for different noise levels,
reduced datasets were reconstructed using compressed sensing by combining eleven
different total variation and L1-norm penalties. Findings showed that data
acquisition pattern and reconstruction parameters have a significant effect on
the resultant 31P-MRSI spectral quality.
Purpose
Phosphorus MR spectroscopic imaging (31P-MRSI) provides
important information about energy metabolism, oxygen state and pH for brain
tumors. However, MR sensitivity of
phosphorus is 15 times less than that of proton and it also has a lower
gyromagnetic ratio (17.24 MHz/T), so larger voxels and averaging several
acquisitions are necessary for adequate signal to noise ratio (SNR) for
31P-MRSI. Previous studies have reported the feasibility of compressed sensing1 to accelerate 31P-MRSI with low SNR penalty2. In this study, we
investigate the impact of factors governing compressed sensing data acquisition
and reconstruction, such as the undersampling pattern, matrix size, noise and
the regularization parameters on the performance of accelerated 31P-MRSI.Methods
Phosphorus MR spectra of seven patients diagnosed with brain
tumors (2 Non-Hodgkin’s lymphoma (NHL), 1 grade I oligodendroglioma, 1
metastasis, 1 grade II oligodendroglioma, 2 grade II astrocytoma, average
age=47.3±12.8 years) that were scanned on a clinical Philips 3T scanner were retrospectively
quantified to estimate the metabolite peak level differences of healthy brain
and brain tumor3. The amplitudes of
phosphocreatine (PCr), glycerophosphorylcholine (GPC),
glycerophosphorylethanolamine (GPE),
inorganic phosphate (Pi), phosphorylcholine (PC), phosphorylethanolamine
(PE), γ-adenosinetriphosphate (ATP), α-ATP and β-ATP peaks of the brain tumor
spectrum were set to be 0.59, 0.93,
0.75, 0.62, 0.82, 0.81, 0.75, 0.72 and 0.72 times the peak amplitudes of the simulated
healthy brain spectrum, respectively. Two dimensional 8x8, 16x16 and 32x32 31P-MRSI
datasets containing a tumor voxel at the corner of top left region, and healthy
spectra at the rest of the array were simulated in MATLAB (The Mathworks Inc., Natick,
MA). Four noise levels that had a standard deviation equal to the 8%, 30%, 40%
and 50% of the maximum signal intensity of the healthy signal were added to the
spectra. Six different random undersampling patterns that have total reduction
factors of 4, 5 and 10, which fully samples a high (6.25% of the total array
size) or low (1.56% for 16x16, 32x32 and 4.69% for 8x8) portion of the k-space center,
were generated for each matrix size. k-space data were undersampled with these
patterns. Each dataset was reconstructed
by using a 3D compressed sensing reconstruction algorithm2 modified from the SparseMRI
package1 with a combination of eleven different total variation (α) and L1-norm (λ) penalties. Root mean square error (RMSE) of each spectra was calculated.
Results
The effective spatial resolution
factor increased with reduction factor and sampling more at the k-space center
for each matrix size (Table 1).The exclusion of L1-norm and total variation penalties from the
inverse problem resulted in a less distinguishable tumor and healthy spectra (Figure
1b) than using a low value of 0.001 for both penalties (Figure 1c). Increasing
the λ penalty resulted in a baseline removal of the spectra and loss of peak
intensities (Figure 1d), and increasing the α penalty resulted in a smearing of
the tumor voxel intensity on the PCr frequency image and the loss of the tumor
region contrast (Figure 1e). The results indicated that using lower values for
both penalties resulted in a narrow point spread function and a more
accurate definition of the tumor voxel location based on its spectrum. Figure 2
shows the RMSE values of the compressed sensing reconstructed datasets with six
different k-space undersampling patterns and four different noise levels, for
8x8, 16x16 and 32x32 matrix sizes. The RMSE values increased with the noise
level for all the datasets. On the other hand, slightly less RMSE values were
observed in high k-space central sampling patterns for all reduction factors.
Matrix sizes did not highly affect the RMSE values. Also it was observed that
higher L1-norm penalties resulted in higher RMSE values due to the denoising
effect. Discussion
Phosphorus MR spectroscopic imaging provides valuable information
about brain tumors, but 31P-MRSI has acquisition time and signal
intensity limitations that prevent the widespread usage in clinical settings.
Compressed sensing reconstruction enables shorter acquisition time and enhances
the signal to noise ratio. The center of the k-space holds the high signal
intensities, so sampling more at the central k-space resulted in better
spectral quality, and lower RMSE values than sampling a lower portion of the
central k-space. The simulation results showed that total variation and L1-norm
penalties affected the definition of tumor location. Conclusion
The results of this study showed that the signal quality of resultant 31P-MRSI was related with the choice of the undersampling pattern, noise level and regularization parameters of compressed sensing. Compressed sensing accelerated 31P-MRSI could be used in the clinical settings for imaging brain tumors with the appropriate set of the undersampling and reconstruction parameters. Acknowledgements
This
study was supported by TUBITAK Career Development Grant 112E036, EU Marie Curie
IRG grant 256528, and Philips Healthcare through a research agreement. References
1. Lustig, M., D. Donoho, and J.M.
Pauly, Sparse MRI: The application of
compressed sensing for rapid MR imaging. Magn Reson Med, 2007. 58(6): p. 1182-95.
2. Hatay, G., et al. Comparison of 2D Iterative Frame Based and
3D Direct Compressed Sensing Reconstruction for Accelerated Phosphorus MR
Spectroscopic Imaging of Human Brain. in In Proceedings of the 22nd Annual Meeting of ISMRM. 2014. Milan,
Italy.
3. Citak Er, F.,
et al., Classification of Phosphorus
Magnetic Resonance Spectroscopic Imaging of Brain Tumors Using Support Vector
Machine and Logistic Regression at 3T. Conf Proc IEEE Eng Med Biol Soc,
2014. 2014: p. 2392-5.