Lucian A. B. Purvis1, Ladislav Valkovic1, Matthew D. Robson1, and Christopher T. Rodgers1
1OCMR, Department of Cardiovascular Medicine, University of Oxford, Oxford, United Kingdom
Synopsis
Calculation of in vivo concentrations requires knowledge of
the B1 field. A common solution to this problem has been to use
field maps measured in phantoms, but this becomes increasingly difficult at
high field. The size of the effect of material and B0 field strength
determining B1 in the liver using phosphorus (31P)
phantoms was investigated at 1.5, 3, and 7T using CST simulations. The effect
of concentration differences at 7T was demonstrated using 15 and 30mM phosphate
phantoms. At 1.5T, using phosphate phantoms with concentrations between 5-40mM
give an error of less than 3%. This increases to 10% at 3T, and 20-114% at 7T.
Purpose
The calculation of absolute concentrations in vivo requires
a reference sample of known concentration1 and the
signals from the scanned subject and the reference must be corrected for the
difference in B1. A common solution to this problem has been to use
field maps measured beforehand in phantoms2, 3.
However, this becomes increasingly difficult at higher field strengths due to
increased B1 inhomogeneity, making the material properties ever more
important4.
In this work, the impact of different material properties on
31P-MRS of the liver is investigated at 1.5, 3, and 7T to estimate
the caution necessary when using phantoms for B1 inhomogeneity
correction. B1+ fields were used, as they can be acquired
in phantoms without prior knowledge of the full B1 field or use of
adiabatic pulses. Methods
CST: CST Studio Suite 2016 (CST AG, Darmstadt, Germany) was used
to simulate B1+ fields of a 10cm loop coil centred above
a 300×300×150mm3 phantom and two voxel models, Laura and Gustav. The
phantom materials used were the CST models for liver and muscle, and materials
with conductivities of 0.02, 0.15, 0.5 and 2 Sm-1 to simulate
various phosphate concentrations. The
fields were simulated at 25.9, 49.9, and 120.3MHz. The coil was tuned and matched at each
frequency, and the B1+ field was sampled along a line
through the centre of the coil and phantom (see Figure 1). The B1+
profiles were normalized to a “reference fiducial” 10mm behind the face of the
coil and averaged across the depth of the liver (see Figure 2).
Phantom:Two K2HPO4(aq) phantoms (15mM and
30mM) with static conductivities (i.e. ω = 0) 0.30 and 0.54Sm-1,
respectively5, were used.
The conductivity is expected to be higher at 120.3MHz due to the Debye-Falkenhagen
effect.
All data were acquired on a Siemens Magnetom 7T system using
a single 10cm diameter 1H loop and a custom-made 10cm 31P
loop. Coil location and loading was calculated using a phenylphosphoric acid
fiducials6. The T1
for each phantom was determined using non-localized inversion-recovery signals.
Ten 31P 3D GRE images (32×16×18, 15.6×15.6×15mm3) were
acquired with varying voltages. Transmit field maps were calculated by fitting
saturation curves to each voxel, and simulating the B1+
required for the given flip angle.
Results and discussion
The CST B1+ ratios to the pure liver
phantom are given for each material at several depths at 7T in Figure 2, and
averaged over liver depths in Figure 3. They show that B0 field effects
are important when considering the makeup of a phantom. The higher the field,
the more closely the phantom must match in vivo tissue.
At 1.5T, using a reference phantom with a conductivity of
between 0.15 and 0.7Sm-1 would give an error of less than 3%. This
increases to 10% at 3T, and 20-114% at 7T. These conductivities are equivalent
to approximate K2HPO4(aq) concentrations of 5-40mM. The
phantom field maps (see Figure 5) show that even concentrations as close as 15
and 30mM have significant differences in B1+. For less than 10% error at 7T, K2HPO4(aq)
concentrations of 20-30mM should be used. At 3T, the difference between human voxel models is
insignificant (<0.2%). At 7T, it becomes more important with differences of
3.5%. Simply matching the phantom conductivity to liver tissue would give 1.5%
error at 3T and 10% error at 7T, but it is possible to more closely match the
in vivo values with slightly lower conductivities.
An example of the phosphate images in given in Figure 4. The
amplitudes of a single pixel were plotted against voltage to show what was
fitted. These results are a simplification of the problem that are
meant to demonstrate the magnitude of the errors resulting from B0
field strength and phantom composition. This is shown by the field maps of the
two phantoms at 7T, where the asymmetry due to conductivity can be seen. This effect twists the field in opposite directions for B1+ and
B1-. However, as the magnitude of the change is
expected to be the same, comparing B1+ can give an
estimate of the error in B1- as well. Conclusion
In this study, the effect of phantom choice on B1
has been demonstrated in simulation at various field strengths. At 1.5T, a wide
range of phantom conductivities (and therefore concentrations) can be used. At
3T, the errors are increased, but are still less than 10% for phosphate
concentrations between 5-40mM. At 7T, even the differences between subjects
start to be important, but if care is taken with phantom selection, errors can
remain below 5%. Acknowledgements
Funded
by a Sir Henry Dale Fellowship from the Wellcome Trust and the Royal Society
(Grant Number 098436/Z/12/Z). LABP has a DPhil studentship from the Medical
Research Council (UK). References
1. Bottomley PA. NMR Spectroscopy of the Human Heart. In:
Harris RK, Wasylishen RE, editors. Encyclopaedia of Magnetic Resonance. Chichester:
John Wiley; 2009.
2. Buchli R, et al. Magn Reson Med. 1994
3. Chmelik M, et al. Magn Reson Med. 2008
4. Vaidya MV, et al. Concepts Magn Reson Part B Magn Reson Eng. 2016
5. Pethybridge AD, et al. J Solution Chem. 2006
6. Rodgers CT, et al. Magn
Reson Med. 2014