Jan Fritz1, Wesley D Gilson2, Christoph Forman3, Esther Ratihel3, Mathias Nittka3, and Allan Belzberg1
1The Johns Hopkins University School of Medicine, Baltimore, MD, United States, 2Siemens Healthcare USA, 3Siemens Healthcare GmbH
Synopsis
Interventional MR imaging at 3 Tesla benefits from high signal
and affords visualization and subsequent targeting of submillimeter structures,
but needle artifacts may be exaggerated. Optimized fast gradient echo- and
turbo spin echo-based pulse sequences minimize in-plane signal displacement,
but through-plane artifacts remain. Compressed Sensing Slice-Encoding Metal Artifact
Correction (SEMAC) MRI has the ability to minimize through-plane displacement,
and thus holds promise to improve the accuracy of device localization. We demonstrate
the clinical feasibility of Compressed Sensing SEMAC TSE for interventional MR
imaging at 3 Tesla and visualization of the needle artifact with high accuracy.
Introduction
Interventional MR imaging at 3
Tesla benefits from high SNR and CNR and affords visualization and subsequent
targeting of submillimeter structures such as small nerves [1]. However, a
limitation of 3 Tesla can be the exaggeration of needle artifacts, which may
impair the ability to precisely localize the needle’s tip relative to the
target. Optimized fast gradient echo and turbo spin echo-(TSE)-based pulse
sequences minimize in-plane signal displacement, but through-plane artifacts
remain. Slice-encoding metal artifact correction (SEMAC) MRI has the ability to
minimize through-plane displacement [2], and thus holds promise to improve the
accuracy of device localization. However, the additional encoding steps for
through-plane metal artifact correction result in longer acquisition times. Compressed
sensing-(CS)-based data sampling strategies exploit the intrinsic sparsity of SEMAC
imaging and can reduce the acquisition time by up to 60% [3].Purpose
To evaluate the feasibility and
accuracy of compressed sensing Slice Encoding for Metal Artifact Correction (SEMAC)
TSE for the visualization and targeting of MR-conditional needles during
interventional MR imaging at 3 Tesla. Methods
Institutional review board
approval was obtained, and all subjects gave informed consent. A 3T MR imaging
system (MAGNETOM Skyra, Siemens Healthcare, Erlangen, Germany) was used in
conjunction with a 4-channel receive-only flexible surface coil and the embedded spine
array coil. We evaluated four commercially available MR-conditional needles of
14G (Invivo, Coax Needle), 16G (Invivo, Coax Needle), 20G (Cook, MReye Needle)
and 22G (Cook, MREye Needle). Using an acrylic glass needle guide phantom, all
needles were inserted into a pork shoulder orthogonal to the external static
magnetic field, which mirrors our clinical practice. Eight pulse sequences
including SEMAC TSE and a CS-SEMAC TSE prototype (Figure 1) were tested. Both SEMAC-based
pulse sequences used 11 encoding steps. Conventional SEMAC used a 3-fold 1D GRAPPA
acceleration, whereas 8-fold CS-SEMAC acceleration was achieved through incoherent
undersampling of the 2D-phase encoding matrix and non-linear, SENSE-type
reconstruction with L1-norm-based regularization [3] (Figure
1). Length
and width of the needle artifacts were quantified using a full-width-half-maximum algorithm [4]. Five MR-guided percutaneous injection procedures were subsequently
performed comparing TSE with a receiver bandwidth of 400 Hz/pixel and CS-SEMAC
using the 20G (Cook, MReye) needle. The needle visualization was visually
assessed by the interventionalist for both pulse sequences and subsequently
ranked in accordance with operator preference. Results
The needle was visualized by all pulse
sequences with sufficient contrast to the surrounding muscle tissue (Figure 2).
Artefact widths and lengths showed a marked dependency on the pulse sequence
type and design (Figure 3). The widths and lengths of the needle artifact were largest
on gradient echo images, and smallest and most accurate on SEMAC images. There
was no significant difference between SEMAC and CS-SEMAC measurements. The
needle tip location errors as indicated by the length of the needle artifacts
was highest for the TSE pulse sequence with 100 Hz/pixel bandwidth and lowest
for SEMAC and CS-SEMAC. All MR-guided procedures were successful with no complications.
CS-SEMAC MR images better visualized the needle compared to 400 Hz/pixel
bandwidth TSE in 4 out of 5 cases (80%) and similar in the other case. CS-SEMAC
images additionally demonstrated a more favorable contrast of the dark needle
artifact to surrounding muscle, ligament, tendons and fibrous tissues.Discussion
Our study demonstrates the
feasibility of an 8-fold accelerated CS-SEMAC TSE pulse sequence for
interventional MR imaging at 3 Tesla. CS-SEMAC TSE displays the needle artifact
with the best accuracy and with a needle tip error of less than 1 mm. During
interventional MR imaging procedures, the needle may be better visualized on
CS-SEMAC images due to a more favorable contrast difference between the signal
intensity of the needle and surrounding tissues. Due to the 2D phase-encoding scheme required for artifact reduction, the acquisition time of CS-SEMAC was
about 1 min. CS-SEMAC may be most helpful for situations where the needle tip needs to be visualized with highest
certainty, rather than routine use. Conclusion
CS-SEMAC TSE is feasible for interventional MR imaging at 3
Tesla and visualizes the needle artifact with high accuracy, which may be helpful
for situations where the needle tip needs to be visualized with highest
certainty.Acknowledgements
The authors would like
to acknowledge Jens Wetzl and Michael Zenge for their work on the image
reconstruction framework.References
1. Fritz J, Dellon AL, Williams EH, Belzberg AJ, Carrino JA. 3-Tesla High-Field Magnetic Resonance Neurography for Guiding Nerve Blocks and Its Role in Pain Management. Magn Reson Imaging Clin N Am. 2015 Nov;23(4):533-45.
2. Lu W, Pauly KB, Gold GE, Pauly JM, Hargreaves BA. SEMAC: Slice Encoding for Metal Artifact Correction in MRI. Magn Reson Med. 2009 Jul; 62(1): 66–76.
3. Fritz J, Ahlawat S, Demehri S, Thawait GK, Raithel E, Gilson WD, Nittka M. Compressed Sensing SEMAC: 8-fold Accelerated High Resolution Metal Artifact Reduction MRI of Cobalt-Chromium Knee Arthroplasty Implants. Invest Radiol. 2016 Oct;51(10):666-76.
4. Thomas C, Wojtczyk H, Rempp H, Clasen S, Horger M, von Lassberg C, Fritz J, Claussen CD, Pereira PL. Carbon fibre and nitinol needles for MRI-guided interventions: first in vitro and in vivo application. Eur J Radiol. 2011 Sep;79(3):353-8.