Alex Bhogal1, Carrie Wismans1, Christiaan Vinckers2, Peter R Luijten1, Dennis WJ Klomp1, and Jannie P Wijnen1
1Radiology, University Medical Center Utrecht, Utrecht, Netherlands, 2Psychiatry, University Medical Center Utrecht, Utrecht, Netherlands
Synopsis
In this
work we attempt to overcome MRSI limitations associated with extra-cranial lipid signal leakage and low SNR at high resolution. We use a dedicated crusher coil
for lipid signal removal, in combination with a high density receive array and
an 7T MR scanner for boosted SNR.
Introduction
Neurodegenerative
diseases and mental disorders are thought to cause changes in metabolic
processes related to neuronal functioning in the cerebral cortex. Elevated
glutamate-glutamine levels have been reported in un-medicated schizophrenia
patients(1), while changes in metabolite levels
of N-acetyl aspartate, creatine and myo-inositol have been associated with Alzheimer’s
related neurodegeneration(2). Cortical metabolite concentrations
can be detected using proton spectroscopy (H1-MRS), however,
limitations relating to metabolite signal contamination from extracranial
lipids and partial volume effects relating to the large voxel sizes have
hindered the progress of magnetic resonance spectroscopic imaging (MRSI) as a routine clinical application.
In this work we attempt to overcome these limitations through the use of a
crusher coil for lipid signal removal, in combination with a high density
receive array and an 7T MR scanner for
boosted SNR. Methods
Two
datasets were acquired from a healthy 35 year old male subject with voxel dimensions
of 3x3x10mm3 and 3x3x5mm3, respectively. MRSI data were
acquired using a Philips 7T scanner (Best, The Netherlands) and dual channel volume transmit coil (NOVA
medical, Wilmington MA, USA) in combination with a high density, 32 channel receive array(3) (MR Coils, Zaltbommel, NL). Extracranial lipid
signal contamination was reduced using a Crusher Coil(4) (MR Coils) which was pulsed
directly after each RF excitation for approximately 0.8ms; for setup see fig. 1.
A pulse acquire sequence was used with a CHESS water suppression. For scan
parameters see figure legends. An un-suppressed water acquisition (at half the
spatial resolution) was used for eddy-current correction and for removal of residual
water signal and water side bands. The water spectrum was fitted to the water
suppressed spectrum and the fit was subtracted and followed by SVD filtering.
The resulting spectra were phased and fitted using LCModel (5). Results
Example,
single voxel spectra derived from the lower resolution acquisition are shown in
figure 2. Average SNR and FWHM for this
selection were 6.83±0.98 and 12.75±1.73Hz, respectively. Figure 3 shows a 3x3
grid of adjacent voxels selected from a
predominantly GM region atop the cortex. Average SNR and FWHM for this
selection were 3.56±0.53 and 16.89±2.58Hz, respectively. These results are in line with
expectations since halving voxel volume should yield a comparable reduction in
signal. Results from fitting metabolite peaks to an example spectra using
LCModel, as well as their representative GM location are show figure 4. Discussion
Using a crusher coil to suppress extra-cranial lipid signal and a high density receive array, we are able to acquire high resolution MRSI datasets of
cortical tissue. Our work raises the potential for high resolution spectroscopy
of the cortex at 7T; previous studies have reported voxel dimensions of 3x3x10mm3 (4) and 3.4x3.4x8mm3 (6). Importantly, we are able to
suppress high lipid signal during the aquisition and acquire data within a reasonable scan time on the
order of 12 minutes. Apart from reduced partial volume effects, high resolution MRSI has intrinsically reduced lipid contamination. The spatial extent of the
point spread function side lobes for each voxel is lower. In combination with
external lipid signal crushing, cortical metabolite spectra are therefore less
susceptible to corruption from signal leakages originating at extra-cranial
lipid regions. Several issues require further investigation; particularly the
optimal current amplitude and duration for lipid signal removal considering the
limited penetration depth of the surface receive coils. Conclusion
Initial results are promising and we have shown that it is possible to acquire good quality in voxels smaller than customary for MRSI imaging. Refinement of this methodology brings the potential to further increase MRSI resolution to gain insight into the mechanisms of cortical metabolic disorders and fine scale neuro-chemical responses.Acknowledgements
No acknowledgement found.References
1. Poels EMP, et al., Imaging
glutamate in schizophrenia: review of findings and implications for drug
discovery. Mol Psychiatry 2014;19(1):20-29.
2. Hancu I, et al., 1H MR spectroscopy using TE averaged PRESS: a more sensitive technique
to detect neurodegeneration associated with Alzheimer's disease. Magnetic
resonance in medicine : official journal of the Society of Magnetic Resonance
in Medicine / Society of Magnetic Resonance in Medicine 2005;53(4):777-782.
3. Petridou N, et al., NMR in
biomedicine 2013;26(1):65-73.
4. Boer VO, et al., Lipid suppression for brain MRI and MRSI by means of a dedicated
crusher coil. Magnetic resonance in medicine
2015;73(6):2062-2068.
5. Provencher SW. Estimation of
metabolite concentrations from localized in vivo proton NMR spectra. Magnetic
resonance in medicine 1993;30(6):672-679.
6. Strasser
B, et al., (2 + 1)D-CAIPIRINHA accelerated MR spectroscopic imaging of the brain at 7T.
Magnetic resonance in medicine Aug 22. doi: 10.1002/mrm.26386