A voxel-based implementation of the Bloch-Siegert phase shift method within a semi-LASER based localization sequence that shows a high robustness to determine transmit gain (TG) for the same volume that is excited for the spectroscopy experiment is demonstrated at 7T. Phantom and in-vivo measurements show higher robustness over a large range of initial TG settings and voxel locations resulting in a faster and more reliable calibration procedure to achieve good voxel selection and spectrum quality and avoid additional calibration steps.
With the high B1 field inhomogeneity at ultra-high field the adjustment of the transmit gain (TG) on a slice might lead to miscalibration of TG in smaller target volumes. This can result in poor excitation slice profile and out-of-voxel signal suppression in single-voxel MRS. To correct for this miscalibration additional calibration steps for excitation, refocusing and water suppression pulses in a semi-LASER can be used.
The phase based Bloch-Siegert (BS) shift method has already been successfully used for TG calibration [1] and can be included in various types of imaging and spectroscopy sequences. BS-PRESS [2] and BS-STEAM [3] determine the optimal TG in the same volume that is excited for the spectroscopy experiment from the average of the BS phase shift of the first high SNR data points of the FID and have already been successfully implemented at 7T.
This study will present an implementation of the Bloch-Siegert phase shift method within the semi-LASER [4] based localization sequence that shows a high robustness to determine TG for the same volume that is excited for the spectroscopy experiment in-vivo at 7T to reduce the amount of additional calibration steps.
A semi-LASER sequence with crusher scheme described in [4] including two off-resonant Bloch-Siegert pulses (fermi, 4 ms, ± 4 kHz relative to water, KBS = 73.68 rad/G2) (Figure 1) was implemented. Experiments were performed both in a MRS phantom (MRS HD Sphere, GE Healthcare, diameter = 18 cm) and in-vivo in a whole body MR950 7T scanner (GE Healthcare) with standard head coil (Nova Medical, 2-channel Tx, 32-channel Rx) with TE/TR = 40ms/4s, 2 averages/scheme).
Data processing was implemented in MATLAB (MathWorks) as an automatic process on the scanner after scan execution. TGpredicted was calculated from the mean of the Bloch-Siegert phase shift [2] using data up to the point of the FID where magnitude signal dropped to 50% of the initial signal. TGstarting was first set by the standard slice-based multi-echo sequence in the automated prescan (APS) process [5]. In order to test the robustness of the technique, TGpredicted was calculated with TGstarting set manually over a larger range around optimal TG. Additional calibration steps acquiring water signal over a range of flip angles for excitation, refocusing and water suppression pulses were performed for TGstarting and TGpredicted. Semi-LASER spectra (TE/TR = 30ms/4s, 32 averages) were acquired with TGpredicted.
1. Sacolick et al. Magn Reson Med 2010; 63: 1315–1322
2. Toncelli et al. Proceedings of ISMRM Workshop on UHF MRI 2013; 34
3. Toncelli et al. Proceedings of ISMRM 2015; 4695
4. Öz et al. Magn Reson Med 2011; 65: 901–910
5. King et al. Proceedings of ISMRM 1995; 699