The ACRIN 6657-extension trial, the first multi-center trial using magnetic resonance spectroscopy (MRS) in breast cancer, has completed and its initial results have been recently published. This study reports on the quality of the MRS data and identifies technical and logistic factors that contributed to a lower-than-anticipated data yield.
Nine sites, using 1.5 T and 3 T systems from multiple vendors, participated in the trial. Details of the trial design and MRS measurement procedures are presented in Ref [2]. Briefly, patients received MRI/MRS scans at baseline (TP1), 1-4 days after first treatment (TP2), between chemotherapy regimens (TP3), and prior to surgery (TP4). Single-voxel MRS was acquired with and without water suppression to measure the tCho and T2-corrected water resonances, as previously described [3]. Spectra were processed and fit to estimate peak amplitudes. Detection of a tCho peak was determined using criteria based on SNR, Cramer-Rao bounds, and fit quality. The tCho concentration was calculated in molal units (mmol/kg-water) using water as an internal reference. Raw and processed data were subjectively scored for quality on a good/fair/poor scale, and data scored as poor were removed for the primary analyses.
A quality control (QC) program was used to help control MRS data quality [4]. Prior to enrolling patients, each site was required to qualify by submitting MRS measurements with acceptable accuracy and spectral quality from a trial-specific spectroscopy phantom (shown in Figure 1). Qualification was on a per-system basis, where a system was defined as the unique combination of MR scanner, breast coil, system software, and pulse sequence. Change in any of these required a new system qualification. Phantom QC measurements were acquired biweekly for each system throughout patient acquisition.
Overall Data Attrition: Figure 2 graphically shows the breakdown of data attrition for the primary analyses. 119 subjects were enrolled in the trial. Seven subjects withdrew before completing the study, 3 were ineligible due to development of metastases, 2 received a non-compliant treatment regimen, and 5 were imaged at incorrect timepoints. MR spectroscopy data at either the first or second time-points was not acquired (n=20) or lost (n=15) in 35 of the remaining 102 cases, leaving 67 cases with MRS data available at both TP1 and TP2. Of these, 16 had poor quality MRS, and 18 had no measurable tCho at either TP1 or TP2, leaving 33 cases with analyzable MRS.
Undetectable Pre-treatment tCho: To determine which factors contributed to spectra with unmeasurable tCho, 14 potential explanatory factors reflecting technical and biological factors were identified and independently evaluated using logistic regression models with chi-square tests (shown in table 1). Only four factors were associated with a higher likelihood of tCho detection: low lipid fraction (p<0.001), small voxel size (p=0.04), narrow water line width (p=0.002), and higher field strength (p=0.006).
Spectral Quality: A total of 165 in vivo MRS datasets from 91 patients were submitted for TP1 and TP2. The subjective quality scores were good (47/165, 28%), fair (85/165, 52%), and poor (33, 20%). The poor cases were attributed to system calibration failure (9), a single unstable system (12), poor shim/baseline (3), inconsistency between water and metabolite scans (5), incomplete acquisition (3), and patient motion (1).
Phantom Reproducibility: Repeated phantom measurements of acceptable quality were available for 15 systems (scanner/coil/software/sequence) used in the trial. Figure 3 shows the mean [tCho] and within-system coefficient of variation (wCV = standard deviation / mean) by system, site, vendor, and field strength. The mean [tCho] measured across systems was 0.85 +/- 0.11 mmol/kg-water, which is lower than the prepared concentration of 1.0 mmol/kg. The wCV (lower plot) ranged from 8.0 - 31%, which indicates high variability both within and between systems.
1. Bolan PJ, Kim E, Herman BA, et al. Magnetic Resonance Spectroscopy of Breast Cancer for Assessing Early Treatment Response: Results from the ACRIN 6657 MRS Trial. In: Proceedings of the 102nd RSNA, Chicago; 2016. p. 16002277.
2. Bolan PJ, Kim E, Herman BA, et al. Magnetic Resonance Spectroscopy of Breast Cancer for Assessing Early Treatment Response: Results from the ACRIN 6657 MRS Trial. Jounal Magn. Reson. Imaging 2016.
3. Meisamy S, Bolan PJ, Baker EH, et al. Neoadjuvant chemotherapy of locally advanced breast cancer: predicting response with in vivo (1)H MR spectroscopy--a pilot study at 4 T. Radiology 2004;233:424–431. doi: 10.1148/radiol.2332031285.
4. Bolan PJ, Garwood M, Rosen MA, Levering A, Blume JD, Gimpel J, Esserman LJ, Hylton NM. Design of Quality Control Measures for a Multi-Site Clinical Trial of Breast MRS - ACRIN 6657. In: Proceedings of the 16th Annual Meeting ISMRM. Toronto; 2008. p. 1588.