Tanvir Baig1, Bhumi Bhusal1, Pallab Bhattacharyya2, Stephen Jones2, and Michael Martens1
1Physics, Case Western Reserve University, Cleveland, OH, United States, 2Cleveland Clinic Foundation, Cleveland, OH, United States
Synopsis
A SEEG electrode contains a number of contact points and as
many number of insulated wires each following a contact. Here, we studied the
coupling among the contact wires of SEEG electrode inserted into phantom, by
using temperature measurements at different contact points, and varying the
lengths of contact wires. From the experiment, we saw that, changing the length
of one contact wire will result in change in temperature rise to other contacts
also, even if wire lengths corresponding to those contacts are unchanged. This
shows significant coupling among the wires at the RF frequency of 3T MRI
system.
Purpose
The RF heating from in-situ
electrodes during MRI scans is a safety concern, and there have been many
published measurements of RF heating near single wire implants, guidewires, and
electrodes1-5. However, no measurements have been reported on the
impact of RF coupling among the multiple contacts and contact wires in stereoelectroencephalography
(SEEG) electrodes used for epilepsy studies. SEEG electrodes, as shown in Figure 1,
have multiple sensing surfaces (contacts), and the same number of insulated
wires (contact wires) connecting them to readout instrumentation. The wires are
insulated from each other, but may couple with each other in addition to
coupling with the RF field. The effect
of this coupling is studied with measurements of RF heating near SEEG
electrodes that are penetrating a phantom. Temperature measurements were performed
during a RF transmit sequence in a 3 T MRI system using a head-only,
transmit and receive RF coil. Methods
For the experimental measurements, an 8-contact SEEG (model
no 2101-08-091, PMT Corporation) electrode is inserted into an ASTM Phantom in filled with gelled saline and instrumented
with temperature sensors (model m3300, Luxtron Santa Clara, CA USA). The whole
set up is placed inside a 3 T Siemens Prisma MRI system with head Tx/Rx coil as shown in Figure 2. The SEEG has 8 contact points, each connected
to its own insulated lead wire (contact wire) and total length of 35 cm between
to ends of the electrode. For this experiment,
9 cm of the electrode is inserted straight into the phantom. The SEEG and wire
leads are placed parallel to, but offset from, the main axis, and enter the
phantom 4 cm below the top and 2 cm inside from the lateral surface. For
measurements, different configurations were created by connecting 0.5 mm
diameter wire of different lengths to the different ends of the lead wires.
This method allowed for measurements where the length of the wire attached to
each contact point could be independently controlled. During the RF sequence, the temperature rise was
measured at the contact points 1, 4 and 8 (1 is the most distal contact point,
4 is the middle one and 8 is the most proximal one as referenced to the entry
point of electrode into phantom) of the electrode by placing the thermal sensors
at the respective positions and applying TSE sequence with a head SAR of 2.8 W/Kg,
whole body SAR of 0.2 W/Kg, time averaged RF Power of 8.3 W, TR = 6470 ms, TE =
71 ms, Turbo Factor = 15 and Echo trains per slice = 18 (Figure 3 shows Schematic of experiment).
To find the maximum heating (resonant) and minimum heating (antiresonant)
lengths of the wire, we changed the length of the contact wire 1, keeping other
wires at fixed lengths (35 cm the electrode length). The resonant length was
found as 65cm and antiresonant length as 125 cm. Then, temperature measurements
were made by setting wire lengths at different resonant and antiresonant combinations,
as shown in table (Figure 4). Results
The temperature measurements for different configurations
are presented in table (Figure 4). The measurements demonstrate the RF coupling
among the wires of SEEG electrode. If any one of the wires is extended to the
resonant length (65 cm), then the temperature at each of the three contacts is
increased, by at least a factor of 3 compared to just the SEEG electrodes alone.
Also, adding a wire at the antiresonant length (125 cm) will reduce the maximum
temperature rise in all 3, but the temperature may still be higher than the
SEEG electrode alone, if there is another wire at the resonant length. Discussions
The results presented in the table (figure 4) indicate that
the temperature rise at the particular tip changes on changing other wire
lengths, even if the corresponding wire length is constant. So there is
significant effect of one contact wire length, on the temperature rise at the contact
(tip) of another wire, indicating significant amount of coupling among the
wires. The temperature rise is maximum at the contact '1' in almost all cases,
because this will act as the real tip due to the coupling effect. Acknowledgements
This research was
supported in part by a Fellowship from the Cleveland Clinic Foundation Epilepsy
Center.References
1. Young CJ, Karmarkar P, and McVeigh ET. Minimizing RF Heating of Conducting Wires in
MRI. Magn Reson Med 2007; 58:1028-1034.
2. Armenean C, Perrin E, Armenean M, Beuf M, Pilleul F, Saint-Jalmes H. RF-induced
Temperature Elevation Along Metallic Wires in Clinical Magnetic Resonance
Imaging: Influence of Diameter and Length. Magn Reson Med 2004; 52:1200-1206.
3. Konings MK, Bartels LW, Smits HFM,
Bakker CJG. Heating Around Intravascular Guidewires by Resonating RF Waves. J
Magn Reson Imaging 2000; 12:79-85.
4. Nitz WR, Oppelt A, Renz W, Manke C,
Lenhart M, Link J. On the Heating of Linear conductive structures as Guide
Wires and Catheters in Interventional MRI. J Magn Reson Imaging 2001;
13:105-114.
5.
Carmichael
DW, Thornton JS, Rodionov R, Thornton R, McEvoy A, Allen PJ, Lemieux L. Safety
of Localizing Epilepsy Monitoring Intracranial Electroencephalograph Electrodes
Using MRI: Radiofrequency-Induced Heating. J Magn Reson Imaging 2008;
28:1233-1244.