Bryant T. Svedin1, Allison Payne1, and Dennis L. Parker1
1Radiology and Imaging Sciences, University of Utah, Salt Lake City, UT, United States
Synopsis
A
multi-echo pseudo-golden angle stack of stars sequence for use in free-breathing interventional procedures is evaluated in vivo
with 5 healthy volunteers for use in MR thermometry in the breast. High spatial
and temporal resolution (1.3 mm3, 1.43 s) is achieved through
k-space filtering. PRF temperature, T2*, ρ (signal magnitude at TE = 0),
respiration correction and fat/water separation are simultaneously measured.
Use of a pseudo-golden angle increment allows for the removal of phase (and
therefore PRF temperature) artifacts due to changing k-space sampling between
reconstructed time points. k-Space sampling based phase reference library greatly
improves temperature standard deviation compared to a single baseline reference.
Purpose
Radial acquisitions offer several unique
advantages for proton resonance frequency (PRF) shift thermometry. Frequently sampling
the k-space center provides motion robust images, as well as the ability to
correct for respiration induced off resonance1. Arbitrarily high
spatial and temporal resolution can be achieved by acquiring successive radial
spokes separated by the golden angle and applying a sliding k-space filter,
similar to the k-space weighted image contrast (KWIC) filter2, to
the reconstruction3. This work investigates a pseudo-golden angle 3D
multi-echo stack of stars acquisition to simultaneously measure PRF shift
temperature, T2* and ρ (signal magnitude at TE = 0), correct respiration-induced
off resonance, and provide water/fat separation with high spatial and temporal
resolution in human breast from multiple healthy volunteers.Methods: Experiment
A 3D stack of stars spoiled GRE sequence was
modified to acquire multiple echo contrasts using a bipolar readout and a
pseudo-golden angle increment. The angle used, based on the ratio of two
Fibanacci numbers α = (1 – 233/377)*360 ≈ 137.5066, will repeat the k-space
trajectory after 377 views. Experiments were performed in 5 healthy volunteers laying
prone in a Siemens 3T Prisma scanner with a breast-specific MRgFUS system4
to assess the effectiveness of this sequence and reconstruction technique in
both coronal and sagittal orientations (1.3x1.3x1.3 mm, FOV = 166x166x20.8 mm, Matrix Size = 128x128x16, Flip Angle = 10, TR
= 11 ms, 6 Echoes, TE = 2.46/3.75/5.04/6.33/7.62/8.91 ms, Partial Fourier: 5/8
in slice direction).Methods: Reconstruction
The slope of the phase vs TE at the
center of k-space was used for respiration correction1. Data was then
reconstructed using a sliding KWIC filter with 13 innermost lines, with each
successive ring using the next higher Fibonacci number until 377 lines in the
outermost ring. The sliding window was advanced 13 views between each reconstruction
time point providing a temporal resolution of 1.43 seconds. The k-space sampling
pattern was repeated after 29 reconstructed time points.
Water and fat images were produced using
the three point Dixon method with the first three echoes. T2*/ρ maps were
calculated using linear regression of the log of the magnitude images vs TE, weighted
by the magnitude M(TEj), as shown in Equation 1.
$$\chi^2=\sum_{j=1}^nM_j (ln(M_j)-(a+bTE_j))^2,\quad a=ln(\rho),\quad b = \frac{-1}{T_2^*}\quad [1]$$Phase data (ψ) from each echo was combined using
linear regression of the phases vs TE, weighted by the magnitude squared, as
shown in Equation 2.
$$\chi^2=\sum_{j=1}^nM_j^2 (\psi_j - (a+bTE_j))^2, \quad a=\psi_0, \quad b = \beta \quad [2]$$
PRF temperatures were calculated using four
different methods. Method 1: The first image was used as the single reference phase
for each echo independently. Method 2: A multi-baseline library was used where
the reference image had the same k-space sampling pattern as the current time
point for each echo independently. Method 3: The combined echo phase was used
with the first image as the reference. Method 4: The combined echo phase was
used with the k-space sampling pattern multi-baseline library.
Results
Figure 1 shows the central slice of the water
and fat images produced from the sequence for each of the 5 volunteers. Figure
2 shows an example of the PRF temperature standard deviation in aqueous tissue
for the 4 methods for volunteer 2. Figure 3 displays the mean PRF precision
within the breast vs TE for the four methods of calculating PRF temperature
change for volunteer 3. Table 1 lists the mean PRF precision within aqueous
tissue in the breast for all 5 volunteers for each of the four PRF calculation
methods. Figure 4 displays the precision of the initial magnitude ρ
as a percent change in adipose tissue.Discussion and Conclusions
This method simultaneously creates separate water
and fat images, and measures PRF temperature, T2* and ρ with a large field of
view (166x166x20.8 mm), high spatial resolution (1.3 mm3) and high
temporal resolution (1.43 s). The PRF precision within all 5 volunteers varied
from 0.4 to 1.2 °C using the combined phase and baseline library for both
coronal and sagittal orientations. The T2* and ρ measurements could be used as
another possible measure of temperature change, especially in adipose tissue
which does not exhibit a PRF shift with temperature5,6. All reconstructed
images contain residual artifacts from regridding and KWIC undersampling which
will change between reconstructed time images as the KWIC filter is advanced.
For this reason, a pseudo golden angle increment was chosen to cause the
residual artifacts to repeat and thus be removable in temperature difference
measurements by using a baseline library. The method shown here provides
promising results for this sequence and reconstruction method for use in free
breathing interventional treatments.Acknowledgements
Funding
Sources: NIH R01
EB013433, CA 172787References
1. Svedin BT, Payne A, & Parker
DL. Respiration artifact correction in three-dimensional proton resonance
frequency MR thermometry using phase navigators. Magn Reson Med.
2015;doi:10.1002/mrm.25860
2. Song HK, & Dougherty L.
k-space weighted image contrast (KWIC) for contrast manipulation in projection
reconstruction MRI. Magn Reson Med, 2000;44(6), 825-832.
3. Winkelmann S, Schaeffter T,
Koehler T, Eggers H, & Doessel O. An optimal radial profile order based on
the Golden Ratio for time-resolved MRI. IEEE Trans Med Imaging, 2007;26(1),
68-76.
4.
Payne A, Merrill R, Minalga E, Vyas U, de Bever J, Todd N, Hadley R, Dumont E,
Neumayer L, Christensen D, Roemer R, Parker D. Design and characterization of a
laterally mounted phased-array transducer breast-specific MRgHIFU device with
integrated 11-channel receiver array. Medical physics 2012;39(3):1552-1560.
5. Baron P, Ries M, Deckers R, de
Greef M, et al. In vivo T2 -based MR thermometry in adipose tissue layers for
high-intensity focused ultrasound near-field monitoring. Magn Reson Med,
2014;72(4), 1057-1064.
6. Rieke V, Butts Pauly K. MR
thermometry. J Magn Reson Imaging 2008;27(2):376-390.