Benjamin N Conrad1,2, Satoshi Maki1, Jennifer M Watchmaker3, Bailey A Box1, Robert L Barry4,5, Seth A Smith1,3, and John C Gore1,3
1Institute of Imaging Science, Vanderbilt University, Nashville, TN, United States, 2Neuroscience, Vanderbilt University, Nashville, TN, United States, 3Radiology and Radiological Sciences, Vanderbilt University Medical Center, Nashville, TN, United States, 4Radiology, Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, MA, United States, 5Radiology, Harvard Medical School, Boston, MA, United States
Synopsis
A hypercapnic gas challenge was used to demonstrate blood
oxygenation level dependent (BOLD) signal changes in the cervical spinal cord,
and the sensitivities of two functional acquisition sequences (standard single
shot (GE-EPI) and 3D multishot (3D-EPI) gradient echo EPI) were compared.
Results indicated that both acquisitions were able to detect signal changes of
about 1% in gray matter and higher values in white matter confirming that BOLD
effects in the cord are reliable. The 3D multishot sequence exhibited higher temporal
SNR and reduced susceptibility distortions, making it an attractive option for BOLD
fMRI in the spinal cord.
Purpose
Several reports of reliable task-induced activity, and demonstrations
of resting state connectivity within and across segments in the spinal cord,
highlight the need for better characterization and validation of spinal cord blood
oxygenation level dependent (BOLD) signals.1,2 A hypercapnic gas
challenge induces iso-metabolic vasodilation and an increase in BOLD signal in
the brain and one previous study has employed such a challenge in spinal cord.3
We sought to measure BOLD signal changes in spinal cord tissues with hypercapnia
and to compare two acquisitions schemes, single-shot gradient echo EPI (GE-EPI)
and 3D multi-shot gradient echo EPI (3D-EPI), for their abilities to detect and
quantify BOLD signal changes. Methods
After
signed, informed consent, seven subjects participated in the study (26.4 ± 4.9 years, 4 female). Scans
were acquired using a Philips Achieva 3T with dual-channel transmit and
16-channel neurovascular coil for reception. After anatomical imaging, two 10-minute gas challenge fMRI runs
were acquired. The acquisition parameters are described in Figure 1. Both were
acquired axially at the same resolution of 1x1x5mm with 8 slices, a FOV of
150x150x40mm, and centered in the middle of C4 vertebral level. A hypercapnic
normoxia gas mixture (5%CO2,21%O2,74%N2) was used and the gas presentation
paradigm is described in Figure 2. Respiration, cardiac, and end tidal CO2 (etCO2) signals were recorded.
Anatomical
masks of gray matter (GM), white matter (WM), and cerebrospinal fluid (CSF)
were manually defined. A fourth tissue mask for the spinal cord venous plexus
(SVP) was created by dilating the border of cord white matter by two voxels which was intended to encompass the spinal veins running along the cord's periphery. The
processing pipeline for each fMRI run involved motion correction focused on the
cord/canal, RETROICOR (correction using the respiratory/cardiac traces), and
coregistration/resampling to the high-resolution anatomical (T2*w mFFE) space (acquired
with the same geometry). An initial step of slice timing correction was applied
to the single shot EPI data. No smoothing was applied. While relatively few distortions
or signal dropout were present in the 3D-EPI data, significant susceptibility
artifacts and tissue distortion in the single shot EPI data were accounted for
by editing the tissue masks prior to statistical analysis.Analysis
A general
linear model (GLM) was applied using a gamma function convolved with the
stimulus time course and delayed by 20 seconds from the onset of gas blocks,
which most closely matched our measured average etCO2 time course (Fig. 2). Percent
signal change time series were calculated at the voxel level by dividing the
signal at each time point by the mean signal of the voxel across the four
baseline periods. To compare the magnitude of signal change between the two
acquisition schemes, for each subject, voxels of each tissue class were sorted
according to the t-values estimated from GLM analysis and the top half of
voxels for each subject were used in the group average signal time course (Fig.
3). Additionally, the numbers of voxels demonstrating a significant BOLD effect
(i.e. “active” voxels) for each acquisition scheme were assessed and compared.
Results
In all subjects
a reliable increase in etCO2 of 4-5mmHg was observed during gas blocks. Mean
tSNR in gray matter was calculated to be 18.3±2.9 for 3D-EPI and 13.6±1.9 for GE-EPI, which was significantly greater in a paired t-test
across subjects (p=0.00006). Figure
3 demonstrates that both sequences detected robust BOLD signal change in the
tissues of interest. The average % peak signal change in GM/WM was 1.20/1.54
for GE-EPI and 0.90/1.05 for 3D-EPI. GE-EPI demonstrated higher average peak
signal change for all tissue classes. The signal change values in CSF and SVP
were highly variable in GE-EPI, suggesting that this sequence is more sensitive
to flow effects and CSF pulsation compared to 3D-EPI. In Figure 4, none of the
paired t-tests were significant. Similar to Cohen-Adad et al.3, we
found significant voxels at the edge of the cord (SVP) and outside the cord in
CSF, suggesting there may be BOLD effects in these regions from vasculature
along the cord and nerve rootlets.Discussion and Conclusion
Both GE-EPI and 3D-EPI are sensitive BOLD signal changes in
the cervical spinal cord. While our GE-EPI provided slightly higher detection
and peak signal change of BOLD response in the tissues of interest, 3D-EPI
provided similar levels of detection, despite a much shorter echo time and
different TR and flip angle. Interestingly, BOLD effects were larger in WM than
in GM, unlike the brain. The merits of 3D-EPI, including higher tSNR and less
susceptibility artifact, make it an attractive option for BOLD fMRI in the
spinal cord.Acknowledgements
This research was
supported by: 4T32EB014841-04 (Gore), W81XWH-13-1-0073 (Smith), NIH/NINDS
R21NS087465, National MS Society, and NIH/NIBIB R00EB016689 (Barry)References
1. Cohen-Adad J & Wheeler-Kingshot C (Eds.), Quantitative MRI of the Spinal Cord, Elsevier (2014).
2. Barry RL, et al. NeuroImage 2015; 133:31-40.
3. Cohen-Adad J, et al. NeuroImage 2010; 50:1074-1084.