Yong Zhang1, Zhongping Zhang1, Kang Wang2, and Zhenyu Zhou1
1MR Research China, GE Healthcare, Shanghai, People's Republic of China, 2GE Healthcare, Milwaukee, WI, United States
Synopsis
Quantitative measurement of oxygen delivery
and consumption in the brain can provide insight into neurovascular and
metabolic coupling. A single shot asymmetric spin echo (ASE) EPI sequence was implemented
for quantitative oxygen extraction fraction (OEF) measurement in healthy
volunteers for the feasibility study. OEF results were in good agreement with
the normal range of 30%-40% reported in the previous literature. ASE EPI shows the potential to provide quantitative OEF maps with
good brain coverage and without the need of gas challenges. Further investigation
is required to evaluate the sensitivity of OEF measurement in the disease
situation, such as stroke.
Purpose
Quantitative measurement of oxygen delivery and
consumption in the brain can provide insight into neurovascular and metabolic
coupling in the healthy and disease situations.
The oxygen consumption by brain tissue can be measured with oxygen extraction
fraction (OEF), also a potential indicator of the occurrence of stroke.1
Positron emission tomography (PET) is regarded as the golden standard for OEF
quantification, which involves the radiation induced side effect. Some other
methods require hypercapnic and hypoxic gas challenges, which are not very practical
in a clinical setting.2 Recently, a single shot asymmetric spin echo
(ASE) EPI approach was reported by H. An and W. Lin for quantitative OEF
measurement.3 The transverse MR signal decay in the presence of a
blood vessel network was modeled to explore the sensitivity of the reversible
transverse relaxation rate to the OEF, as proposed by Haccke and Yablonskiy.4
In this study, we developed an ASE EPI sequence for measuring the OEF in the
brain and performed the feasibility study on healthy volunteers.Methods
Five healthy volunteers (age 20-43 years old, one female)
gave written informed consent to participate the study. All MRI scans were
performed on a 3.0-T MR750 scanner (GE Healthcare, Milwaukee, USA) using an
8-channel phase array head coil. The ASE EPI sequence was a variation of a
single-shot spin echo EPI sequence. The 180° refocusing pulse was shifted relative
to the original position (time shift τ) while the time of echo (TE) was kept
constant (Fig. 1). The ASE EPI sequence used a fixed TE to achieve the same R2
weighting but various time shift τ to introduce different R2’ and R2* weighting
on the acquired images. Scan parameters were as follows: TR/TE=2000/80ms, FOV=240mm,
slice thickness/gap=5/1mm, acquisition matrix=64×64, number of excitations
(NEX) = 3. The ASE EPI sequence was repeated 19 times to acquire 19 asymmetric
spin echoes with time shift τ corresponding to -18, -16, …, 2, 0, 2, …, 16, 18ms.
The total acquisition time was 4min to acquire 28 slices. The ASE EPI images
were used to estimate two parameters (reciprocal of reversible transverse
relaxation rate R2’ and deoxygenated blood volume λ) by fitting them to the
model proposed by Haacke and Yablonskiy.4 Then δω map could be calculated
using: πΏπ=π
2′/λ (Eq.1). After
that, OEF value was estimated using the following equation: ππΈπΉ=πΏπ/(πΎβ4/3πββχβπ»ππ‘βπ΅0) (Eq.2),
where γ was the gyromagnetic ratio, Δχ represented the susceptibility
difference between fully deoxygenated blood and fully oxygenated blood with the
value of 0.18ppm. Hct was the hematocrit value, which was assumed to be 0.42.Results
Fig.2 shows the MR signal on the ASE EPI images sampled
at multiple refocusing pulse offsets.
The MR signal first increased when the 180° refocusing pulse approached
to the middle position between the excitation and readout pulses and then
decreased when the refocusing pulse passed the symmetric position. Fig. 3 shows
the representative ASE EPI image (left) and the corresponding OEF map (right) of one volunteer. Results were in good agreement with the normal
range of 30%-40% reported in the literature.3Discussion and Conclusion
ASE EPI has the potential to provide quantitative OEF maps
with good brain coverage and without the need of gas challenges. These results
hold promise for some clinical uses, such as stroke, showing increased OEF in
the cortex areas with decreased blood supply. Further investigation is required
to assess the sensitivity of OEF measurement in the disease situation.Acknowledgements
No acknowledgement found.References
1. Derdeyn CP, Yundt KD, Videen TO, et al.
Increased oxygen extraction fraction is associated with prior ischemic events
in patients with carotid occlusion. Stroke. 1998; 29:754-758.
2. Ni WW, Christen T, Zaharchuk G. Assessing reproducibility
and changes in oxygenation with R2’ during clinical hypercapnic and hypoxic gas
challenges. Proc Intl Soc Mag Reson Med. 2015;1450.
3. An H, Lin W. Impact of intravascular signal
on quantitative measures of cerebral oxygen extraction and blood volume under normo-
and hypercapnic conditions using an asymmetric spin echo approach. Magn Reson
Med. 2003;50:708–716.
4. Yablonskiy DA, Haacke EM. Theory of NMR
signal behavior in magnetically inhomogeneous tissues: the static dephasing
regime. Magn Reson Med. 1994;32:749–763.