Laurentius Huber 1, Jun Hua 2, Valentin G Kemper3, Sean Marrett4, Benedikt A Poser3, and Peter A Bandettini1
1SFIM, NIMH, Bethesda, MD, United States, 2FM Kirby Research Center, Johns Hopkins, United States, 3MBIC, Maastricht University, Netherlands, 4NIMH, United States
Synopsis
fMRI at ultra-high field strengths (≥7T) allow
submillimeter voxels across different cortical layers. A big challenge to infer
layer-dependent activity information, however, is to find an fMRI contrast that
has the best combination of local specificity to microvascular responses within
cortical layers and sensitivity to detect activity changes. Here we compare
contrast mechanisms that have been proposed to be applicable to layer-dependent
fMRI, including blood volume fMRI with VASO, SE-BOLD, T2/T1ρ-prep-BOLD,
diffusion weighted T2-prep-BOLD. We find that CBV-weighted VASO and T2-prep
methods have a favorable compromise between sensitivity and specificity. Hence,
we believe that these fMRI methods might be better suited for layer-dependent
applications than conventionally used GE-BOLD and SE-EPI.
Purpose
High resolution,
cortical depth-dependent (layer-specific) fMRI has the potential to address
questions about top-down and bottom-up driven connectivity. There are two major
limitations for the application of neuroscientific cortical depth dependent
studies. A) the limited contrast-to-noise-ratio (CNR) with decreasing voxel volume,
and B) the unwanted functional sensitivity to large locally-unspecific blood
vessels distant of the activated cortical layer. Until today there is no
consensus which kind of fMRI modality would provide the best compromise between
CNR and local specificity to investigate depth-dependent activity signals. The
purpose of this study is to compare a variety of fMRI methods including, GE-BOLD,
SE-BOLD, VASO-CBV, T2-prep, and diffusion-weighted T2-prep. We do this
comparison with the focus to layer-dependent activity in the human motor cortex
during a finger tapping task.Methods
Experiments were performed on a 7T Siemens scanner
with a 32-channel NOVA head-coil. Four volunteers participated in 18
experiments.
Simultaneous CBV-VASO (Fig. 1A) and GE-BOLD (Fig.
1B) data are acquired analogously to [1] (TE = 22ms). The SE-BOLD (Fig. 1C) was
acquired with a C2P sequence is from (https://www.cmrr.umn.edu/multiband/) (TE
= 50 ms) [2]. For functional T2-weighting, T1ρ-prep [3] was implemented with a
spin locking pulse on resonance, where T1ρ ≈ T2 [4]. We included a
custom-designed adiabatic spin-locking pulse that is optimized for 7T. The
magnetization is kept in the transverse plain (on-resonance) for TE = 50 ms (Fig.
1D). An alternative T2-prep (Fig. 1E) contrast was generated by re-implementing
the Phillips sequence from [5] into the SIEMENS environment (prep TE = 50 ms).
In order to minimize unwanted unspecific intravascular T2-BOLD signal within
large pial veins, we included small diffusion weighting crusher gradients [6]. All experiments were conducted with 0.7 mm x 0.7 mm
in-plane nominal resolution. 1.5 mm thick slices are aligned to be
perpendicular to the cortical surface of M1 (Fig. 2A). All scans used GRAPPA 2
acceleration (left-right). A 12-min finger-tapping paradigm was used here to
investigate the layer-specificity of the different methods. Layering analysis
was done in the respective functional datasets analogous to [1].Results
High resolution tapping induced activity is depicted
in Fig. 2. GE-BOLD signal has the highest CNR but the strongest contamination
of large veins. T2-prep sequences and VASO seem to be less dominated from locally unspecific big blood vessels above the cortical surface, but they are more limited by CNR. CNR is
estimated here as the signal change times the tSNR, normalized by the square
root of the repetitions per unit of time.
VASO, T1ρ/T2-prep, are able to
reveal distinct layer-dependent activity in deeper cortical depths (Fig. 2 D/G/H
and Fig. 3A). GE-BOLD and SE-BOLD fail to reveal layer-dependent peaks
presumably due to their limited specificity, while diffusion-weighted T2-prep
seems to fail to reveal layer-dependent peaks presumably due to its limited
sensitivity.Discussion
The fact that SE-BOLD with EPI-readout has large
unwanted sensitivity to pial veins might be coming from T2*-contaminations in
off-center k-space lines as shown in [7,8]. This T2* component can be minimized
in a T1ρ-prep method using a TurboFLAH readout. When diffusion gradients are
applied, the additionally unwanted sensitivity to intravascular blood in large
pial veins is reduced. This, however, is accompanied with a substantial loss of
activation detectability. The limited SNR in the diffusion weighted T2-prep might
also come from incomplete refocusing in the 90º-180º-180º-90º module, when B1+-inhomogeneities
are present. Future comparisons with alternative promising functional
T2-weighted imaging modalities, such as 3D-GRASE [8], RARE/TSE [9], SSFP [10] could
make this comparison study more complete.Conclusion
Fig. 3B/C shows that in a
simplified picture, those functional contrasts that have a high CNR have in
turn a low specificity and vice versa. Dependent on the expected functional
contrast change expected VASO might provide a good compromise between CNR and
specificity, as is an outlier in the negative correlation between sensitivity
and specificity in Fig. 3C.Acknowledgements
The research was supported by the NIMH
Intramural Research Program (#ZIA-MH002783).References
[1] Huber et al., ISMRM, 2016, #948;
[2] Auerbach et al., MRM, 2013, 69:1261-1267;
[3] Rane et al., MRI, 2014, 32:813-818;
[4] Taheri and Sood, MRI, 2006, 24:51-59;
[5] Hua et al., MRM, 2014, 72:1530-1540;
[6] Boxermann et al., MRM, 1995, 34:4-10;
[7] Goense et al., MRI, 2006, 24:381-392;
[8] Kemper et al., Front Neurosci, 2015, 9:163;
[9]
Poser et al., MAGMA, 2007, 20:11-17;
[10] Gao et al., Front Neurosci, 2014, 8:49.