Myung-Ho In1, Shinho Cho1, Yunhong Shu2, Hoon-Ki Min1,2,3, Matthew A. Bernstein2,3, Oliver Speck4,5,6,7, Kendall H. Lee1,3, and Hang Joon Jo1
1Departments of Neurologic Surgery, Mayo Clinic, Rochester, MN, United States, 2Departments of Radiology, Mayo Clinic, Rochester, MN, United States, 3Departments of Physiology and Biomedical Engineering, Mayo Clinic, Rochester, MN, United States, 4Department of Biomedical Magnetic Resonance, Institute for Experimental Physics, Otto-von-Guericke University Magdeburg, Magdeburg, Germany, 5German Centre for Neurodegenerative Diseases (DZNE), Magdeburg, Germany, 6Leibniz Institute for Neurobiology, Magdeburg, Germany, 7Center for Behavioral Brain Sciences, Magdeburg, Germany
Synopsis
Deep brain stimulation (DBS) fMRI has been considered as an emerging
tool in investigating the DBS mechanisms and corresponding clinical outcomes,
but suffers from severe susceptibility artifacts near metallic electrodes and
tissue/air boundaries. A recent study showed that point spread function (PSF)
mapping-based reverse gradient approach has a potential to correct distortions even
in gradient-echo echo-planar imaging (GE-EPI) images with opposite phase-encoding polarity using
a PSF dataset. To minimize the susceptibility artifacts, in this study, we
apply the PSF approach for DBS-fMRI in swine. The results demonstrate that this
approach can be beneficial for improving the reliability of DBS-fMRI.
PURPOSE
As deep brain
stimulation (DBS) has been widely used for the treatment of many disorders, it is
important to understand the underlying mechanisms. fMRI is an emerging tool in
investigating the relationship between the DBS mechanisms and clinical
outcomes. In general, DBS-fMRI suffers from severe susceptibility artifacts
near metallic electrodes for electrical stimulation as well as tissue/air boundaries
of the brain, which result in strong intensity and geometric distortions along
the phase-encoding (PE) direction in gradient echo
echo-planar imaging (GE-EPI). These
distortions present a major challenge to perform a faithful data analysis and interpretation.
A recent study1 showed that the point spread function (PSF)
mapping-based reverse gradient approach can correct distortions in both forward
and reverse phase-encoded GE-EPIs with opposite distortions using a single 3D
GE-PSF reference dataset acquired with only one PE direction. In this study, we
apply the proposed approach for DBS-fMRI in swine to minimize the susceptibility
artifacts and investigate the efficiency for improving the reliability of DBS-fMRI
studies.METHODS
Eleven healthy
pigs underwent unilateral nucleus accumbens DBS-fMRI on a 3T Signa Excite MRI
scanner (16.0M4, GE Medical Systems, Milwaukee, WI) using a homemade 6-channel receive-only
surface coil. The details of the experimental preparation including DBS surgery
are described in previous studies2,3.
A unilateral DBS lead was externalized and connected to an external pulse
generator for electrical stimulations (5V), which was synchronized to the start
time of DBS-fMRI scan. A block paradigm consisted of 5 rest (30 sec.) and
electrical stimulation (6 sec.) cycles and ended with the rest period. The PE
polarity in GE-EPI was altered, which resulted in a pair of GE-EPIs with
opposite geometric distortions, and
measured in an interleaved order during DBS-fMRI. The imaging protocol was: TR/TE=1500/40 ms,
no partial echo, GRAPPA factor=2, 19 slices, readout bandwidth=651Hz/pixel,
FOV=160×160, matrix size=96×96, voxel resolution (x,y,z)=1.7×1.7×2.4 mm3. To calculate kernels for distortion
correction of the image pair, PSF reference data were acquired in prior scan using
identical imaging parameters with a shorter TE (30 ms) to minimize T2*
dephasing effects. The PE polarity yielding dominantly stretched distortions
around the electrodes was chosen for the PSF scan. Motion correction and voxel
smoothing (using 3 mm Gaussian kernel) were performed using SPM24 in
different variants of the EPI series including forward and reverse phase-encoded
EPIs, with and without distortion correction and the weighted average of the
distortion-corrected EPI pair. Finally, BOLD contrasts for DBS stimulation effects were calculated for
comparison. In addition, the number of voxels with signal dropouts at the
metallic electrode was counted in the distortion-corrected and the combined
images.RESULTS AND DISCUSSION
Strongly
stretched (yellow arrows) and compressed regions (green arrows) are present in
the forward and reverse EPI, respectively due to metal-induced strong field
inhomogeneities (Fig. 1). With
the proposed distortion correction, the geometric differences between the
forward and reverse EPIs were substantially reduced.
The anatomy of the distortion-corrected images aligned with each other and the
reference image well (see dashed line in Fig. 1), which enabled a reliable
combination of the distortion-corrected images. Since severe intra-voxel dephasing effects resulting in
strong signal dropouts in GE-EPI, strongly compressed regions manifest in opposite direction in the corresponding GE-EPI
pair with opposite PE polarity. We observed some image intensities mismatch even
after the distortion correction. Therefore, losses of BOLD
contrasts caused by signal dropouts were only partially recovered in the
distortions (Fig. 2). Overall a stronger BOLD activation was obtained due to
the improvements of the signal-to-noise ratio (SNR) of BOLD sensitivity in the weighted
average image series (Fig. 2), but at the cost of temporal resolution reduced by
a factor of 2. The number of signal dropout voxels around the metallic
electrode (Fig. 3A) was reduced by up to 40% in the combined image (Fig. 3B). Less intensity distortions in either forward or reverse
phase-encoded EPIs were observed from 5 and 6 subjects, respectively among the 11
subjects. In addition, this
approach resolved the intensity distortions near the tissue-air interfaces
(Fig. 4), especially near the frontal lobes, where strong BOLD activation was previously
observed3.CONCLUSION
The
results demonstrate that the proposed approach could correct geometric distortions
of GE-EPI data with reversed PE directions efficiently and could minimize
strong susceptibility artifacts round the metallic electrodes as well as in the
regions near the tissue/air boundaries in the brain. Therefore, the proposed
approach can be a good method to improve BOLD contrast for DBS-fMRI, which is
beneficial in investigating the mechanisms of DBS. To improve the temporal
resolution of this approach, multi-echo EPI with opposite PE polarity will be
considered in future.Acknowledgements
This work was supported by The Grainger Foundation.References
1. In
MH, Posnansky O, Beall EB, Lowe MJ, Speck O. Distortion correction in EPI using
an extended PSF method with a reversed phase gradient approach. PLoS ONE.
2015;10(2):e0116320. doi: 10.1371/journal.pone.0116320.
2. Min H-K, Hwang S-C, Marsh MP, Kim I,
Knight E, Striemer B, et al. Deep brain stimulation induces BOLD activation in
motor and non-motor networks: an fMRI comparison study of STN and EN/GPi DBS in
large animals. Neuroimage. 2012;63(3):1408-20.
3. Knight EJ, Min H-K, Hwang S-C, Marsh MP, Paek S, Kim I, et al. Nucleus
accumbens deep brain stimulation results in insula and prefrontal activation: a
large animal FMRI study. PloS one. 2013;8(2):e56640.
4. http://www.fil.ion.ucl.ac.uk/spm/