Christian Anderson1, Charlie Wang1, Yuning Gu1, Yun Jiang1, Dan Ma2, Mark Griswold1,2, Xin Yu1,2,3, and Chris Flask1,2,4
1Biomedical Engineering, Case Western Reserve University, Cleveland, OH, United States, 2Radiology, University Hospitals Cleveland Medical Center, Cleveland, OH, United States, 3Physiology and Biophysics, Case Western Reserve University, Cleveland, OH, United States, 4Pediatrics, University Hospitals Cleveland Medical Center, Cleveland, OH, United States
Synopsis
Magnetic Resonance Fingerprinting allows for rapid,
simultaneous multi-parametric quantification. The use of a FISP imaging kernel
is necessary due to field inhomogeneity in human body and preclinical imaging.
We have observed that increases in the applied gradient moment lead to gradient
moment size dependent changes in T2 accuracy. By examining different applied
gradient moments we illustrate this loss of T2 accuracy while T1 measurement
accuracy is maintained. This has implications for FISP MRF design and the
implementation of unbalanced gradient moments for quantification.
Purpose
Magnetic Resonance Fingerprinting has demonstrated a powerful ability to
perform simultaneous multi-parametric quantification in vivo on high field, preclinical MRI scanners (≥7T)1,2.
Similar to human applications in the body3,4, use of MRF at high
field strengths requires a FISP imaging kernel with an unbalanced gradient
moment to eliminate banding artifacts seen in TrueFISP. We have observed that
the magnitude of the dephasing gradients in the FISP MRF technique has the
potential to introduce quantification errors, particularly to T2 estimation. Understanding
the importance of the applied gradient moment in quantification accuracy is a
key feature in developing accurate preclinical MRF methodologies and robust
methods. In this work, we investigate the impact of altered dephasing gradient
moment magnitude on the accuracy of MRF-based T1 and T2 measurements on a 7T
high field preclinical MRI system.Methods
Data for this work was acquired on a Bruker 7T preclinical
system with temperature controlled at 27 ºC ±1.5 ºC for in vitro data. A FISP imaging kernel served as the basis for the
MRF acquisitions with repetition time (9.5-12 ms) and flip angle (0-70 degrees)
patterns similar to previous implementations1,5,6. The FISP imaging
kernel was adapted by changing the unbalanced gradient moment for each
experiment (red gradient lobe in Figure 1). To test
different imaging conditions, 6 gradient moments were applied in separate in vitro MRF experiments: 0 mT·ms/m
(TrueFISP), 31 mT·ms/m, 62 mT·ms/m, 93 mT·ms/m,
124 mT·ms/m, and 248 mT·ms/m. MRF-based T1 and T2 relaxation time maps
were obtained by matching acquired data on a pixel by pixel basis to a simulated dictionary of possible profiles using template
matching to identify the best match. The same MRF dictionary, assuming a constant multiple of 2π dephasing within a voxel per TR, was used for each
FISP MRF experiments. A separate TrueFISP MRF dictionary was simulated for the
0 mT·ms/m gradient moment as this experiment exhibits different spin dynamics, does not have a constant dephasing moment per TR, and is sensitive to off resonance. The resulting in vitro T1 and T2 maps were compared across all of the gradient
moments. In vivo mouse kidney experiments
were also performed (TrueFISP MRF vs 2 x FISP MRF
acquisitions) to evaluate the impact of the FISP MRF gradient dephasing moment.Results
In
vitro MRF-based T1 and T2 maps for 4 MnCl-doped saline phantoms
are shown in Figure 2. Mean values for
each phantom from each MRF acquisition were plotted against gold-standard spin
echo results in Figure 3. Representative profiles from each phantom for each
gradient moment are plotted in Figure 4. MRF-based mean T1 values showed good
agreement with gold-standard spin echo results across all gradient
moments. Mean T2 values measured with
MRF exhibited gradient moment size dependent deviations from corresponding spin
echo results. This same effect was observed qualitatively in the in vivo kidney MRI results (Figure 5). The
FISP MRF results with the higher gradient dephasing moment (248 mT·ms/m) exhibited
a narrowed dynamic range in T2 measurements in comparison to the lower gradient
moment (62 mT·ms/m) as evidence by a loss of T2 contrast between the renal
cortex and medulla. TrueFISP in vivo
maps suffered from banding artifacts that were eliminated in the FISP MRF maps.Discussion
In this study, we have demonstrated the gradient moment
dependence of the T2 relaxation time maps for the FISP MRF technique on high
field preclinical MRI scanners. These initial in vitro and in vivo MRI
results show that FISP MRF-based T2 relaxation time estimates are significantly
reduced in comparison to gold-standard spin echo MRI assessments for larger
gradient dephasing moments. Therefore, the amount of dephasing moment applied
for a FISP MRF acquisition should be implemented carefully to ensure the accuracy
for both in vitro and in vivo studies. These results suggest
that there is a progressive reduction in the MRF signal coherence as the
dephasing gradient moment is increased resulting in significant
under-estimation of the T2 values.Conclusion
We have identified a potential error source in T2 relaxation time
estimates for FISP MRF implementations on high field preclinical MRI
scanners. The sensitivity of these T2 estimates
to the magnitude of the dephasing gradient moment demonstrates the need to
carefully design FISP based MRF acquisitions. This study can help advise future
MRF experiments in order to maintain T2 accuracy with a properly
designed MRF imaging kernels.Acknowledgements
NIH T32EB007509, NIH TL1TR000441, NHLBI R21 HL130839, Cystic Fibrosis FoundationReferences
1. Gao Y, Chen Y, Ma D, et al. Preclinical MR fingerprinting (MRF) at 7T: effective quantitative imaging for rodent disease models. NMR Biomed. 2015; 28: 384-394.
2. Buonincontri G, Sawiak S J. MR Fingerprinting with simultaneous B1 estimation. Mag Res Med. 2015. 76(4): 1127-1135.
3. Chen Y, Jiang Y, Pahwa S, et al. MR Fingerprinting for rapid quantitative abdominal imaging. Radiology. 2016. 279(1): 278-286.
4. Hamilton J, Jiang Y, Chen Y, et al. MR Fingerprinting for rapid quantification of myocardial T1, T2, and proton spin density. Mag Res Med. 2016 Early View.
5. Jiang Y, Ma D, Seiberlich N, et al. MR Fingerprinting using fast imaging with steady state precession (FISP) with spiral readout. Mag Res Med. 2015. 74(6): 1621-1631.
6. Ma D, Gulani V, Seiberlich N, et al. Magnetic resonance fingerprinting. 2013. Nature. 495: 187-192.