Slice-GRAPPA (SG) is often used to reconstruct blipped-CAIPI simultaneous multi-slice EPI data, in particular in conjunction with the dual even-odd kernel approach (SG-DK) to mitigate ghosting related reconstruction artifacts. To achieve good performance in blipped-CAIPI acquisition, the CAIPI shift factor should be optimized in a case by case manner to minimize g-factor penalty. The g-factor is influenced by the reconstruction approach and so far a fast analytical g-factor calculation has been developed only for standard SG reconstruction but not SG-DK, where time-consuming Monte Carlo simulation is still needed. Here we propose an analytical g-factor calculation for SG-DK and demonstrate that the proposed method is fast and accurate. Our simulation and experimental results also highlight the superior performance of SG-DK over SG reconstruction, both in mitigating image artifacts and noise penalty.
The standard analytical g-factor calculation 5 for acquisitions that contains both slice and in-plane accelerations (MB x Rinplane), reconstructed using SG and in-plane GRAPPA is shown as equation 1 in table 1. Here, p represents the coil sensitivities, and W and V the in-plane and slice GRAPPA weights in image domain. Equation 2 shows the image domain reconstruction equation for kth coil image (Ik) for the case where SG-DK and in-plane GRAPPA are utilized. Here, V1 and V2 represent modified image domain slice-GRAPPA weights, synthesized from the odd and even GRAPPA kernel weights (Veven and Vodd) according to equations 3 and 4. Equation 5 shows the analytical g-factor formulation for the combined SG-DK and in-plane GRAPPA reconstruction, where the slice weight V is replaced by Ṽ that is a hybrid slice weight combined from the odd and even kernels, Veven and Vodd (as per equation 6). Here, V1,j,l and V2,j,l are separately computed as the sum of odd-even weights and the difference of odd-even weights in image domain; Σ² is the noise covariance matrix; Λ² is the augmented noise covariance matrix from Σ².
In vivo human brain data were acquired using the blipped-CAIPI SMS-EPI sequence (TE/TR: 74/3600ms; matrix: 297×147×91; voxel size: 1.5mm iso; FOV: 220mm; MB × Rinplane = 3×3; FOV/2 CAIPI shift) on a SIEMENS 3T Prisma scanner (SIEMENS, Erlangen, Germany) with commercial 64-channel head/neck coil (12 neck channels were not used during acquisition). Fully sampled dataset with matching image distortion was also obtained as references scan, using multi-shot EPI acquisition. This fully sampled reference dataset was used for g-factor calculations to compare the Monte Carlo and the proposed analytical method. Two acceleration factors were simulated at MB × Rinplane = 3×2 and 3×3, with the later matching that of the acquired accelerated dataset. To carefully replicate the situation in the slice-accelerated acquisitions, only “slice-averaged” ghost correction is performed on the simulated accelerated data prior to slice unaliasing via SG/SG-DK. Additional g-factor simulations and image reconstructions were also performed for an exaggerated ghost case (6 times of actual ghost level) to mimic acquisition scenario that contain high ghost levels. Here, ghost was added to the dataset prior to SG/SG-DK reconstruction. All image reconstruction and analysis were performed offline in MATLAB.
In the simulations of Figure 1, the reconstructed images from SG and SG-DK were compared at different accelerations (MB × Rinplane = 3 × 2 and 3 × 3) and different EPI ghost levels. The results show SG-DK can provide improved reconstruction and lower noise penalty when compared to SG reconstruction, particularly when the ghost level is high. In Figure 2 and 3, the simulation (MB × Rinplane = 3 × 2; FOV/2 CAIPI shifting) and in vivo experiment (MB × Rinplane = 3 × 3; FOV/2 CAIPI shifting) illustrate the proposed g-factor calculation provide similar results to the traditional Monte Carlo method (100 replicas). Moreover, the calculation time of the analytical method was 50× less than the traditional method. Figure 4 shows the analytical method could be used to quickly analyze the SNR performance at different accelerations (MB × Rinplane = 3×1 and 3×3) and CAIPI shift (FOV/2, FOV/3 and FOV/4).
1. Setsompop K, et. al. Blipped-controlled aliasing in parallel imaging for simultaneous multislice echo planar imaging with reduced g-factor penalty. Magn Reson Med. 2012; 67(5):1210-24.
2. Breuer FA, et. al. Controlled aliasing in parallel imaging results in higher acceleration (CAIPIRINHA) for multi-slice imaging. Magn Reson Med. 2005; 53(3):684-91.
3. Setsompop K, et. al. Improving diffusion MRI using simultaneous multi-slice echo planar imaging. Neuroimage. 2012; 63(1):569-80.
4. Cauley SF. et. al. Interslice leakage artifact reduction technique for simultaneous multislice acquisitions. Magn Reson Med. 2014; 72(1):93-102.
5. Breuer FA, et. al. General formulation for quantitative G-factor calculation in GRAPPA reconstructions. Magn Reson Med. 2009; 62(3):739-46.
6. Zhu K, et. al. Analytical G-factor for Cartesian Simultaneous Multi-Slice Imaging. ISMRM 2016. Abstract: 609.
7. Robson PM, et. al. Comprehensive
quantification of signal-to-noise ratio and g-factor for image-based and
k-space-based parallel imaging reconstructions. Magn Reson Med. 2008;
60(4):895-907.