Xiaona Li1, Zhigang Peng1, Panli Zuo2, Yi Sun2, Nittka Mathias3, and Jianling Cui1
1Radiology, the Third Hospital of Hebei Medical University, Shijiazhuang, People's Republic of China, 2MR Collaboration NE Asia, Siemens Healthcare, Shanghai, People's Republic of China, 3Siemens Healthcare, Erlangen, Germany
Synopsis
To compare the
clinical detection rates in local lesions using the slice-encoding for metal
artifact correction (SEMAC) & view angle tilting (VAT) Turbo Spin Echo
(TSE) with the
standard TSE sequence in patients with titanium screws.
Following surgery, patients with reported discomfort at the operation site were
examined with both TSE and SEMAC-VAT TSE imaging. They were also examined by
digital radiography (DR) and computed tomography (CT). All images were assessed
and scored. Despite the long scan time, SEMAC-VAT TSE reduced metal-induced
artifacts and may improve the diagnostic sensitivity compared to standard TSE,
DR and CT.
Purpose
This
study was conducted in patients with titanium screws who had reported symptoms
of discomfort at the operation site. The aim was to compare the slice-encoding
for metal artifact correction (SEMAC) & view angle tilting (VAT) 1-3 Turbo Spin Echo (TSE) sequence with the standard
TSE sequence in terms of detection rates in local lesions, and discuss the
clinical value of this technique.Methods
Nineteen patients with musculoskeletal disorders (13 males and 6 females; mean
age, 41.3±6.0 years;
range, 14-56 years) with titanium screws in the spine (n=8), femur (n=8) and
tibiofibula (n=3) after their
operation, were examined because of discomfort
using a 3T MR scanner (MAGNETOM Verio,
Siemens, Erlangen, Germany). The examination included conventional TSE and a SEMAC-VAT TSE
sequence. The acquisition parameters for TSE were: FOV of 180×180 mm2 for spine and 300×300 mm2 for extremity, matrix of 320×256,
slice thickness of 3 mm, flip
angle of 120 degrees, turbo factor of 3. T1-weighted imaging was performed with TE/TR of
6.5/886 ms; T2-weighted imaging was performed with TE/TR of 70/6000ms. T2-weighted imaging with fat suppression was performed with TE/TR of 70/6000ms, TI of 150 ms. The slice-encoding steps were 15 for SEMAC-VAT TSE. All patients were also
examined by digital radiography (DR) and
computed tomography (CT). Three radiologists, each with more than 10 years experience
in musculoskeletal imaging,
independently evaluated the image quality using 5-grade scores, based on the metal
artifacts, delineation of the prosthesis, and delineation of the
peri-prosthesis structures. They then counted the detected lesions on different
images based on MR findings, needle aspiration biopsy, and pathology findings.Results
Compared
with standard TSE images, visualization
of all peri-prosthetic anatomic structures were significantly improved in SEMAC-VAT TSE (p<0.01) (Table 1).
Figure 1 shows a case of a patient with reported back pain one week after spinal surgery. The hyperintense
regions surrounding the implant were metal-induced artifacts. They were largely
suppressed in SEMAC-VAT TSE images, allowing the edema surrounding the screws
to be visualized (arrow). Figure 2 shows a case of a patient with a history of
fracture in the neck of the femur. Femoral head necrosis was confirmed on
SEMAC-VAT TSE images, but was not visible in standard TSE images. Figure 3 shows a patient with recurring
tumor around the metal implant in the middle femur. The tumor was visible on
SEMAC-VAT TSE images, but was missed on standard TSE imaging due to metal-induced artifact. 17 out of the 19 SEMAC-VAT TSE scans
accurately demonstrated lesions (detection
rate=89.5%), including
eight with local infection and surrounding edema, six with femoral head
necrosis, and three with recurring tumors. For postoperative infection with
edema in soft tissue, the
detection rate using DR was 37.5% (n=3), CT or standard TSE was 62.5% (n=5), and SEMAC-VAT TSE was 100% (n=8). For femoral head necrosis,
there were eight suspected cases revealed by DR and one by CT. However, two suspected cases were false positives
using these two methods. They were both excluded by SEMAC-VAT TSE, while
standard TSE missed one. For recurring tumors, only MR successfully revealed
the lesions, two revealed by standard TSE and three by SEMAC-VAT TSE.Discussion
TSE
imaging with SEMAC-VAT technique not only reduced metal-induced artifacts, but
also improved the diagnostic sensitivity in postoperative patients. The diagnosis rate of the local lesion was higher in SEMAC-VAT
TSE than in standard TSE, DR and CT. However,
the acquisition time was significantly longer in SEMAC-VAT than in standard TSE (T1w, 1min 6s; T1w-
SEMAC-VAT, 5mins 18s; T2w, 1mins 17s;
T2w- SEMAC-VAT, 8mins 12s).Conclusion
This
study demonstrated that SEMAC-VAT provides a higher clinical diagnosis rate than standard TSE, DR and CT in
routine musculoskeletal postoperative examination for patients reporting
discomfort near the operation site. Future advances in MR technique to reduce
scan time, such as through compressed
sensing, will benefit the use of SEMAC-VAT in clinical practice.Acknowledgements
No acknowledgement found.References
1.Ai T,
Padua A, Goemer F, et al. SEMAC-VAT and MSVAT-SPACE
sequence strategies for metal artifact reduction in 1.5T magnetic resonance
imaging. Invest Radiol. 47(5): 267-76 2012.
2. Friedrich B, Wostrack M, Ringel
F, et al. Novel Metal Artifact Reduction Techniques with Use of Slice-Encoding
Metal Artifact Correction and View-Angle Tilting MR Imaging for Improved
Visualization of Brain Tissue near Intracranial Aneurysm Clips. Clin
Neuroradiol. 26(1):31-7 2014.
3. Shun Qi, Wu AG,Mu YF, et al. SEMAC-VAT MR Imaging
Unravels Peri-instrumentation Lesions in Patients With Attendant Symptoms After
Spinal Surgery. Medicine. 95(14):e3184 2016