Maolin Qiu1, Yuqing Wan1, and R. Todd Constable1
1Yale School of Medicine, New Haven, CT, United States
Synopsis
MRI with
non-linear spatial encoding magnetic fields (SEM) can provide high quality MR
images, but there images are usually calculated using iterative optimization
procedures. Such reconstruction methods can
take a long time to converge, and the results may depend on the initial iteration
parameters. We propose a fast, non-iterative image reconstruction method for
O-space imaging based on the local K-space and local SEMs and demonstrate its effectiveness for image reconstruction of nonlinear O-space imaging
data.
Introduction
MRI with
non-linear spatial encoding magnetic fields (SEM) can provide high quality MR
images, but there images are usually calculated using iterative optimization
procedures. Such reconstruction methods can
take a long time to converge, and the results may depend on the initial iteration
parameters. We propose a fast, non-iterative image reconstruction method for
O-space imaging1 based on the local K-space and local SEMs2. Preliminary
results based on an analytical phantom are presented.Theory
In general, O-space imaging with arbitrary
k-space trajectories can be expressed as $$s(t)=\int_{(x,y)\in{I}}^{} m(x,y)e^{j2\pi\gamma\mu(x,y,t)}dxdy=\int_{(x,y)\in{I}}^{} m(x,y)e^{j2\pi[xk_x+yk_y+\gamma(x^2+y^2)G_Ot]}dxdy \space\space (1)$$ When the
readouts are radial projections: $$s(t)=\int_{(r,\theta)\in{I}}^{} m(r,\theta)e^{j2\pi\gamma[rG_{xy}+r^2G_O]t}rdrd\theta \space\space (2) $$ in which the
phase is only relevant to r. Since
the non-linear SEM is radial, we have encoding frequency: $$\omega(r)=\gamma\frac{\partial \mu}{\partial t}=\gamma(rG_{xy}+r^2G_O) \space\space (3)$$ Its
linear approximation $$\omega(r)\approx\omega(r_0)+\frac{\partial \omega}{\partial r_0}(r-r_0)=\gamma(r_0G_{xy}+r_0^2G_O)+\gamma(G_{xy}+2r_0G_O)(r-r_0) \space\space (4)$$ at r0 with the local k-space $$K(r_0), k_r|_{r=r_0}=\gamma\frac{\partial \mu}{\partial r}|_{r=r_0}$$ For simplicity, we choose the local K-space
located at r0=0, Eq 2 becomes $$s(t)=\int_{(r_L,\theta)\in{I}}^{} m(r_L,\theta)e^{j2\pi\gamma r_LG_{xy}t}r_Ldr_Ld\theta \space\space (5)$$ for which IFT methods
could be used for image reconstruction after the position and intensity
corrections described below. Non-linear correction: 1D FFT in the
radial direction will give rL so we can the solve the following equation for r that will be used for
resetting the position: $$rG_{xy}+r^2G_O=r_LG_{xy}$$ The derivative of (3), $$\frac{\partial \omega(r)}{\partial r}=\gamma(G_{xy}+2rG_O)$$ which dictates the local
stretch in r, will be used for
intensity correction. The corrected 1D
signal is then IFFT'd back to raw K-space and 2D IFT can be used for 2D image
reconstruction.Results
An analytical phantom (Fig 2a) was used to
generate the k-space data according to Eq 2 with the following parameters: Gxy=18.35mT/m,
Go=0.025mT/m, 64 projections covering the 2D and 64 readouts per projection but
the O-center only sweeps the left 180o. The 1D position and intensity correction in
one projection are illustrated in Fig 2. The k-space data were first corrected
used the proposed method, then after correction the conjugate phase method3 was
employed to reconstruct the images (Fig 2c). For comparison, if the same conjugate
phase method was applied directly on the raw data without any correction, reconstruction
results are shown in Fig 2b.Discussion and Conclusions
This work introduces and
demonstrates the effectiveness of a fast, non-iterative method for image
reconstruction of nonlinear O-space imaging data. Residual distortions and intensity
non-uniformities in the image near the edges still exist, as shown in Fig 2c.
This might be due to the approximation of Eq 2 with a linear form of Eq 5, by a
single local k-space located at r0=0. Further investigation is needed
to refine the distortion and intensity correction. This approach can also
incorporate multiple local k-space data set with a multi-channel RF parallel
acquisitions, for further reductions in image distortion. Also needed is a more
elaborate implementation of the conjugate phase method for a better evaluation
of the reconstruction efficacy.Acknowledgements
This work is supported by the NIH under grant NIH R01
EB076978.References
1.Stockmann
JP, Ciris PA, Galiana G, Tam L, Constable RT. O-space imaging: Highly efficient
parallel imaging using second-order nonlinear fields as encoding gradients with
no phase encoding. Magn Reson Med. 2010 Aug;64(2):447-56. doi:
10.1002/mrm.22425.
2. Gallichan
D, Cocosco C, Dewdney A, Schultz G, Welz A, Hennig J, Zaitsev M, Simultaneously
driven linear and nonlinear spatial encoding fields in MRI, Magn. Reson. Med.,
65 (2011), pp. 702-14.
3. Douglas
C. Noll, Jeffrey A. Fessler, Bradley P. Sutton. Conjugate Phase MRI Reconstruction
with spatially variant sample density correction. IEEE Transactions on Medical
Imaging, Vol. 24, No. 3, March 2005 pp.325-36.