Eddy Solomon1, Gilad Liberman1, Zhiyong Zhang 1, and Lucio Frydman1
1chemical Physics, Weizmann Institute of Science, Rehovot, Israel
Synopsis
This
study presents a new diffusion tool based on a novel single-shot 2D MRI method
called xSPEN. xSPEN is characterized with unusual resilience to field
heterogeneities, but extending it to DTI requires to overcome the strong intrinsic
diffusion weighting of this technique. To achieve this we formulated xSPEN’s
diffusion weighting using a novel, spatially localized b- matrix analysis, and devised
a novel diffusion-weighting scheme that overcomes xSPEN’s original limitations.
These methods were numerically validated and applied to new DTI xSPEN sequences
with which we mapped diffusion in often unreachable human head regions,
including optic nerve and olfactory bulb regions.
Motivation
xSPEN
[1] is a new sequence providing exceptional resilience to field heterogeneities
by relying on the application of a continuous gradient along the
slice-selection axis, to deliver its images. This imparts a heavy
diffusion weighting, making it hard to obtain either isotropic or tensor
diffusion information. This work overcomes this and extends xSPEN’s use to full
DTI mapping, by formulating xSPEN’s diffusion effects on the basis of a local b-matrix analysis, and then using this
insight to devise an alternative diffusion weighting scheme that enables the sampling
of a sufficiently large volume in b-space.
Using these new tools we carried out human diffusion experiments that
demonstrate xSPEN’s ability to measure diffusion in head regions, that are
usually unreachable by single-shot methods.Theory
As in other SPEN based sequences [2,3], the
sequential encoding and decoding occurring along xSPEN’s imaging axis results
in a localized b-weighting which can be described by an extension of the
Karlicek-Lowe diffusion formalism [4]:$$$\textbf{b}(t,r_0)=\gamma^2\int_{0}^{t}(\vec{\nabla}(\phi(t',r_0)\vec{\nabla}(\phi(t',r_0)^\text{T})dt'$$$, where $$$\phi(t',r_0)$$$ is the
localized phase evolved by spins until time t’ under both the imaging
and diffusion manipulations. A dominant contribution to b(t,ro) arises in xSPEN from the gradient acting along the slice-select axis,
which is on throughout the scan. Gradient strengths in clinical scanners are
insufficient for modulating further the diffusion weighting along this axis. Thus,
an alternative gradient scheme was devised whereby the low-bandwidth dimension
(y in Fig 1) is decoded twice, in two
alternate sets of experiments that swap the readout (x) and slice-select (z) axes.
As the slice-select gradient dominates the b-weighting
this means that, for each of these acquisitions, the diffusion will be
dominated by orthogonal bii
directions. Bipolar diffusion modules applied along the remaining orthogonal orientations
(see sequence in Figure 1) will further modulate the effective sampling of the b-space, leading to sets of conic shapes
rotated about bii. The
combination of two orthogonal xSPEN acquisitions then leads to a sufficiently
extensive sampling of the full b-space,
to enabling the retrieval of DTI parameters. Figure 2 illustrates this form of
sampling the reciprocal diffusion space, and compares it to a standard
spherical diffusion scheme.Methods
Sequence
validation: To
confirm the reliability of the double-cone approach, a set of 2601 sample
tissues were simulated, with each tissue having a random proton density and a
characteristic diffusion tensor with randomized directionality and eigenvalues spanning
realistic fractional anisotropy (FA) and apparent-diffusion-coefficient (ADC) values.
The signals arising from these different tissues under the action of the scheme
in Fig. 2B were simulated, and additive Gaussian noise was added to the
simulated signals so that the mean SNR for the b0 set would be 7%.
Based on these synthetic sets, FA and ADC values were then estimated, and both mean
absolute difference and r2 values were calculated against the ground
truth FA and ADC values for various test gradient schemes.
Scans:
Human
head images were acquired at 3T using a Siemens Trio scanner and a 32-channels
head coil. A volunteer was scanned with the xSPEN DTI scheme at full brain coverage
(4mm isotropic resolutions, b=1000 s/mm2, 4 repetitions), and with
restricted FOV (3mm isotropic, b=800 s/mm2, 2 repetitions). EPI-DTI
was run with the same b-values and a standard 32-direction scheme. For
anatomical reference, T1- and T2-weighted scans were included.
Results
Figure
2B examines the new gradient scheme, by evaluating two planar DTI acquisitions
performed in orthogonal orientations (gray and blue), each involving 15 bipolar
diffusion modules whose directions were uniformly spread on the orthogonal plane.
The scheme provided 0.89 and 0.91 r2 fit values for the FA and ADC respectively,
vis-à-vis 0.95 and 0.96 values obtained by the standard scheme (Figure 2A and
2B, bottom).
Figure
3 compares a 4mm isotropic xSPEN ADC map obtained with this scheme on a human head,
against comparable results arising from SE-EPI. Notice how xSPEN preserves the anatomical
features and provides diffusion information on regions challenged by strong
inhomogeneity effects (Figure3, a-f) that are distorted or completely absent in
SE-EPI.
Figure
4 shows an additional example with a 3mm isotropic zoomed-in DTI parametric
maps of the optic nerve region, where xSPEN preserves anatomical features and
provides both FA and ADC values, despite the strong susceptibility effects by air
interfaced to tissue and muscle in this region.Discussion and Conclusion
Thanks
to its high resilience to field inhomogeneities, xSPEN DTI succeeds in characterizing
key brain regions that are usually inaccessible by traditional single-shot
methods. Diffusion information from those
regions can open new opportunities in basic and clinical investigations.Acknowledgements
Funding by grants ISF 795/13, ERC-2014-PoC # 633888, Minerva Foundation #712277, and the Kimmel Institute of Magnetic Resonance (Weizmann) is acknowledged. We are also grateful to Dr. Sagit Shushan (Wolfson Medical Center), and the Weizmann MRI team (Edna Furman-Haran, Fanny Attar and Nachum Stern).References
[1] Zhang, Zhiyong,
Amir Seginer, and Lucio Frydman. "Single-scan MRI with exceptional resilience to
field heterogeneities." Magn
Reson Med; doi:10.1002/mrm.26145 (2016).
[2] Shrot, Yoav, and
Lucio Frydman. "The effects of molecular diffusion in ultrafast
two-dimensional nuclear magnetic resonance." The Journal of chemical physics 128.16 (2008): 164513.
[3] Solomon, Eddy,
Noam Shemesh, and Lucio Frydman. "Diffusion weighted MRI by spatiotemporal
encoding: analytical description and in vivo validations." Journal of Magnetic Resonance 232 (2013): 76-86.
[4] Karlicek, R. F.,
and I. J. Lowe. "A modified pulsed gradient technique for measuring
diffusion in the presence of large background gradients." Journal
of Magnetic Resonance (1969) 37.1 (1980): 75-91.