Bonnie N. Joe1, Kimberly Ray1, Amie Lee1, Vignesh Arasu1, Lisa Wilmes1, Vibhas Deshpande2, Sinyeob Ahn2, and Nola Hylton1
1Radiology and Biomedical Imaging, UCSF, San Francisco, CA, United States, 2Siemens Medical Solutions
Synopsis
In order to gain acceptance of DWI in the clinical
setting, consistent high image quality with minimal breast distortion is
required. Although time of acquisition is slightly longer for the RESOLVE
compared with the advanced diffusion sequence, the benefits of improved image
quality, particularly with respect to image distortion and phase ghosting are
preferred by clinical breast imaging radiologists based on this reader study.
The theoretical benefits of using readout-segmented diffusion imaging
technique in RESOLVE to improve image quality can be realized in a routine
clinical practice.
Background
Addition of diffusion weighted imaging (DWI) to
routine clinical breast MRI protocols provides complementary information to
DCE-MRI for lesion characterization and can potentially improve the specificity
of breast MRI. Numerous studies have
demonstrated restricted water diffusion in breast malignancies and significant
differences in ADC values of benign and malignant lesions (1-3). However, variable image quality of breast DWI
due to the frequency of inadequate fat suppression, image distortion and other
artifacts has limited clinical adoption of DWI in breast imaging. In order to gain acceptance of DWI in the
clinical setting, consistent high image quality with minimal breast distortion
is required.Purpose
The purpose of this study was to compare radiologist perception
of DWI image quality on advanced diffusion & RESOLVE diffusion sequences acquired
under routine clinical conditions.Methods
This was an IRB-approved, HIPAA-compliant study. During the month of September, 2015 we
transitioned from the Advanced (single shot) DWI sequence to the RESOLVE (epi)
DWI sequence as part of our routine clinical breast MR protocol. The first 10 consecutive clinical breast MR
exams during the month preceding the transition and during the month following
the protocol transition were identified for a total of 20 exams: 10 Advanced DWI and 10 RESOLVE DWI. Scan
parameters for the advanced diffusion sequence were: TA 5:35min, 1.8x1.8x2.4
mm, 43 transverse slices, BW: 18.8Hz, TR/TE/FL: 8000.0/ 82.4/180, NEX 2, b
value = 0, 100, 600, 800 sec/ mm^2. Scan
parameters for the RESOLVE sequence were TA: 6:03min; 1.6x1.6.x2mm, 56
transverse slices, TE/TR=80/9800 msec, Ipat(GRAPPA)2, Echo spacing 0.32msec, b
value = 0 and 800 sec/mm^2.
Two fellowship trained breast imaging radiologists independently
evaluated image quality of the diffusion weighted sequences (b0 and b800) on
anonymized images obtained at nipple level with the corresponding axial T2 TSE
and axial VIBE pre-contrast images available for anatomic reference. Readers were asked to rate image distortion,
phase ghosting, homogeneity of fat suppression, and overall image quality of
the diffusion images using a 5-point scale where 1 represents poor
(uninterpretable) image quality with significant artifact (Figure 1) and 5 represents
excellent image quality with almost no artifact. Each reader evaluated the cases in a
different randomized order, blinded to type of diffusion sequence. Reader scores were tabulated and image
quality ratings for the two sequences compared using the Wilcoxon rank sum test
with p<0.05 considered significant.
Reader agreement for each type of diffusion sequence and for overall
image quality were also evaluated.Results
RESOLVE DWI scored significantly higher
for image quality overall as compared to Advance DWI (medians [95% CI]: 3.5
[3.6-4.5] vs. 2.5 [2.6-3.6], p= 0.006). RESOLVE DWI also scored significantly
higher for image distortion and phase ghosting (Figure 2, Table 1). Both
sequences scored well in terms of fat saturation with no significant
differences (medians [95% CI]: 4.75 [4.7-5.0] vs. 5.0 [5.0-5.0], p = 0.37).
Although the observed inter-reader agreement between two readers was 84-91% and
significantly higher than the expected agreement, the Kappa quadratic score was
poor to moderate, indicating high inter-reader variability (Table 2). This lower kappa may be explained by the fact
that the reader scores clustered toward the higher end of the scale rather than
being evenly distributed from 1 to 5.
Overall, this may be a positive result because it indicates reasonable
image quality for both diffusion sequences.Conclusion
Although
time of acquisition is slightly longer for the RESOLVE compared with advanced
diffusion sequence, the benefits of improved image quality, particularly with
respect to image distortion and phase ghosting are preferred by clinical breast
imaging radiologists. Acknowledgements
No acknowledgement found.References
1. Ei
Khouli RH, Jacobs MA, Mezban SD, et al. Diffusion-weighted imaging improves the
diagnostic accuracy of conventional 3.0-T breast MR imaging. Radiology
2010;256(1):64-73.
2. Partridge SC, Rahbar H, Murthy R, et
al. Improved diagnostic accuracy of breast MRI through combined apparent
diffusion coefficients and dynamic contrast-enhanced kinetics. Magn Reson Med
2011;65(6):1759-1767.
3. Partridge SC, DeMartini WB, Kurland BF,
Eby PR, White SW, Lehman CD. Quantitative diffusion-weighted imaging as an
adjunct to conventional breast MRI for improved positive predictive value. AJR
Am J Roentgenol 2009;193(6):1716-1722.