Moritz Braig1, Axel J. Krafft1, Jochen Leupold1, Juergen Hennig1, Marius Menza1, and Dominik von Elverfeldt1
1Medical Physics, University Medical Center Freiburg, 79106, Germany
Synopsis
There is growing interest in preclinical imaging and analysis
of complex flow patterns. For example, 4D flow imaging of mouse models with vascular
plaques or aortic constriction could provide insights into the development and pathogenesis
of cardiovascular diseases. Unfortunately, 4D-flow MRI is often compromised by
a trade-off of reasonable acquisition durations and achievable spatial
resolution. Our work evaluates an undersampling strategy, namely UNFOLD, for
preclinical 4D flow MRI, which has the potential to decrease measurement time
by 50%. Artifacts emerging from the undersampling are removed by unaliasing in
the temporal domain using a lowpass filter after fourier transformation.
Purpose
Preclinical 4D-flow imaging is compromised by very
long acquisition times due to the required 3D spatial and velocity encoding.
Parallel imaging techniques are less advantageous in preclinical applications than
in human MRI, as the number of available receiver coils is typically smaller
which limits the achievable acceleration factors. In our study, we quantitatively
compare fully sampled datasets to retrospectively undersampled datasets in
combination with the UNFOLD reconstruction technique1 to evaluate the
performance of the UNFOLD method for preclinical phase contrast MRI. This
method has the potential to reduce measurement time by 50%, while preserving high
quality magnitude image and velocity data. Methods
Preclinical MR imaging was
done on a Bruker Biospec 70/20 USR equipped with a two channel transmit/receive
cryogenic (26 K helium cooled, Bruker) mouse head surface coil. Six C57BL/6 mice were
measured under the approval of the ethics committee (Ref: G-14-91). A respiration
and ECG gated balanced four point cartesian phase contrast sequence2 was used (FOV = 20x16x14 mm³, TE/TR = 2/5; Matrix = 60x50x50
(interpolated to 125x100x88), VENC = 150-170 cm/s; FA = 15°, acquisition time ~
45 min). Image reconstruction was done offline using Matlab 2015b. To evaluate the
UNFOLD technique1, fully sampled
datasets were retrospectively undersampled by using every other phase encoding line
in an alternating fashion: for odd timeframes all odd phase encoding lines and
for even time frames all even encoding lines were maintained, while the
remaining lines were discarded as shown in figure 1. A fast fourier transform (FFT)
was then applied along the time direction in the undersampled k-space domain
and undersampling artifacts were removed with a rectangular filter. The rectangular
filter was defined so that the six central frames around the zero frequency after
FFT were left unchanged whereas the outer ones are set to zero. Afterwards an
inverse FFT was applied in the temporal domain and the unfolded data was conventionally
reconstructed.
Flow values and peak velocities were evaluated in four
planes which were placed perpendicular to the aorta as shown in figure 4 using
Ensight (CEI, USA) and in-house built Matlab tools3,4. In
each plane, the aorta was manually segmented for every time point. Flow values
were analysed via Bland-Altman plot.Results
Aliasing artifacts from undersampling are substantially
reduced in the reconstructed UNFOLD images (figure 5). Residual artifacts can
be observed in the first and last timeframe of the magnitude images. The
comparison of velocity data between fully sampled and undersampled dataset
shows minor deviations as seen from the Bland-Altman plot (figure 2). Higher
velocities seem to be slightly underestimated whereas at low velocities UNFOLD values
appear systematically higher. The highest discrepancy is observed at the first
acquired timepoint of the cardiac cycle. The streamlines of fully sampled and undersampled
datasets agree very well and reveal the same complex flow pattern (figure 5). Exemplary
flow curves over the cardiac cycle of one animal are illustrated in figure 4. Mean
peak velocities at the four different evaluation planes for each dataset are
shown in figure 3.Discussion
The undersampling artifacts in the magnitude images
were strongly reduced and only apparent in the first and last timepoint. Residual
artifacts originate mainly from the chest wall that has a very high signal due
to its high fat content and its proximity to the surface coil. There is a good
agreement between peak velocities and flow curves, the slight underestimation
of the peak velocities, as seen in the Bland-Altmann plot, may arise from a
temporal filtering effect. Here, improvements could be achieved using tailored
filters5
or an increased temporal sampling rate with a higher number of cardiac frames. Nevertheless,
the comparison of fully sampled and unfolded data shows consistent flow values and
even streamlines of the complex flow pattern in the aorta agree well. Therefore,
the implementation of UNFOLD undersampling would enable a shortened measurement
time without substantially compromising velocity information. Shorter
measurement time may be beneficial to ensure a more stable heart rate during
the acquisition, as heart rate changes are more likely to occur over longer
timespans6.
Additionally shorter measurement times allow higher scanner throughput and
reduce animal stress. Conclusion
Measurement time could be decreased by 50% using the
UNFOLD technique while preserving the complex flow pattern of the murine aorta.
The described technique allows combining 4D-flow imaging together with other
imaging protocols, or increase the resolution while maintaining reasonable total
scan times. Improvements could be made with a more sophisticated temporal filter
design.Acknowledgements
No acknowledgement found.References
1. Madore,
B., Glover, G. H. & Pelc, N. J. Unaliasing by fourier-encoding the overlaps
using the temporal dimension (UNFOLD), applied to cardiac imaging and fMRI. Magn.
Reson. Med. 42, 813–828 (1999).
2. Braig, M. et al. High Quality
Preclinical 4D-Flow Phase Contrast Imaging. in (2016).
3. Bock, J., Kreher, W., Hennig, J. &
Markl, M. Optimized pre-processing of time-resolved 2d and 3d phase contrast
MRI data. in In Proceedings of 15th Scientific Meeting of International Society
for Magnetic Resonance in Medicine, (2007).
4. Stalder, A. F. et al.
Quantitative 2D and 3D phase contrast MRI: optimized analysis of blood flow and
vessel wall parameters. Magn. Reson. Med. 60, 1218–1231 (2008).
5. Kellman, P., Sorger, J. M., Epstein, F.
H. & McVeigh, E. R. Low Latency Temporal Filter Design for Real-Time MRI
Using UNFOLD. Magn. Reson. Med. Off. J. Soc. Magn. Reson. Med. Soc. Magn.
Reson. Med. 44, 933–939 (2000).
6. Constantinides, C., Mean, R. &
Janssen, B. J. Effects of Isoflurane Anesthesia on the Cardiovascular Function
of the C57BL/6 Mouse. ILAR J. Natl. Res. Counc. Inst. Lab. Anim. Resour.
52, e21–e31 (2011).