Courtney K. Morrison1, Jacob M. Johnson2, Yuji Iwadate3, Kevin King4, James H. Holmes2, Frank R. Korosec1,2, Roberta M. Strigel1,2,5, and Kang Wang6
1Department of Medical Physics, University of Wisconsin, Madison, WI, United States, 2Department of Radiology, University of Wisconsin, Madison, WI, United States, 3Global MR Applications and Workflow, GE Healthcare, Hino, Japan, 4Global MR Applications and Workflow, GE Healthcare, Waukesha, WI, United States, 5Carbone Cancer Center, University of Wisconsin, Madison, WI, United States, 6Global MR Applications and Workflow, GE Healthcare, Madison, WI, United States
Synopsis
Spatial resolution has typically been prioritized at
the expense of temporal resolution in the setting of dynamic contrast-enhanced breast
MRI. In this work, we evaluated the use of compressed sensing (CS) with
intermittent fat suppression for improved temporal resolution by comparing quality
of fat suppression and overall image quality between a sequence accelerated
using CS to one without CS.
Introduction
Spatial and temporal resolution present competing
demands in MRI. In clinical dynamic contrast-enhanced (DCE) breast MRI, spatial
resolution is typically prioritized over temporal resolution due to the
importance of assessing lesion morphology.1 However, recent advances
in MRI may help improve the temporal resolution of DCE breast MRI while
maintaining the same high spatial resolution. One such advancement, compressed
sensing (CS), allows for the reconstruction of MR images from undersampled
k-space data, thus improving the temporal resolution.2-4 However,
this technique requires incoherent sampling, which can be particularly challenging
when combined with intermittent fat saturation. In this work, we present a
technique that uses compressed sensing in order to accelerate the acquisition
while also enabling intermittent fat suppression.Methods
Nineteen patients undergoing a clinical breast MRI
exam consented to this IRB-approved, HIPAA-compliant study. Patients were
scanned on a 3.0 T scanner (Discovery MR 750w, GE Healthcare, Waukesha, WI)
using a 16-channel breast coil (Sentinelle, Invivo, Gainesville, FL). Single
phases of two 3D T1-weighted spoiled gradient echo based acquisitions, one with
CS and one without CS (non-CS), were acquired at the end of the clinical breast
MRI exam, approximately ten minutes after the injection of contrast agent. The order
of the two research sequences was alternated to account for the difference in post-contrast
timing.
To reduce scan time, data-driven parallel imaging
was used to uniformly undersample outer k-space. For the CS scans, outer
k-space was further undersampled pseudo-randomly using a Gaussian probability
function, which is well-suited for compressed sensing reconstruction.5
A spectrally-selective adiabatic fat suppression pulse was used for both
research sequences. However, in order to achieve acceptable fat suppression
with CS, the timing of the intermittent fat suppression pulses was modified to account
for the incoherent sampling required for CS. The spatial resolution of both scans
was 0.7 x 0.7 x 1.4 mm3, while the temporal resolutions for the CS
and non-CS scans were 85 and 109 seconds, respectively.
Two readers specializing in breast MRI performed a
blinded review of images during two sessions, one to assess the quality of fat
suppression and one to assess overall image quality. Images were scored as: 1) CS
much better than non-CS (clinically significant), 2) CS slightly better than
non-CS (not clinically significant), 3) CS and non-CS equivalent, 4) non-CS
slightly better than CS (not clinically significant), and 5) non-CS much better
than CS (clinically significant). During each session, images were reviewed
side-by-side, and the position of the images on the screen (left or right) was
randomized. During each session, the order of the patients was also randomized.
Readers were blinded to the sequence type and results of the other reader.
Results
CS reduced the scan time by 24 seconds (~22%) for a single phase
compared to the non-CS sequence. Most cases were scored as having equivalent
fat suppression (16/19 and 17/19 for each reader). Of the cases that did not
have equivalent fat suppression, the difference in fat suppression quality was never
considered clinically significant (Table
1). Similarly, most cases were scored as having equivalent image quality
(13/19 for both readers). One reader scored a single case as a clinically
significant difference, with the non-CS images having better image quality than
the CS images. While no obvious motion was present during the acquisition of
the non-CS scans, evidence of patient motion was visible during the acquisition
of the CS scans and likely contributed to the difference in image quality (Figure 1). All other differences in
image quality were not clinically significant. A representative case is shown
in Figure 2.Discussion and Conclusion
Combining the acquisition with CS undersampling and intermittent fat
suppression provided a 22% improvement in temporal resolution while maintaining
the quality of fat suppression and overall image quality compared to the
sequence without CS. Although one case was considered to have a clinically significant
difference in overall image quality, patient motion during the CS acquisition likely
contributed to the decreased image quality. This study evaluated a single phase
post-contrast; however, a full DCE exam would likely assist in identifying the presence
of motion. Further work is needed to demonstrate the clinical utility of this
technique as well as to evaluate this technique in patients with malignant
lesions.Acknowledgements
We would like to thank Fred Kelcz, MD, JulieAnn
Stover, MD, and Shane
Rassman, MD, for participating in the reader studies. We appreciate support
from the RSNA, GE Healthcare, NIH/NCI P30 CA014520, and the Department of Radiology and Cancer Center at
the authors’ institution.References
1. Kuhl CK, Schild HH, Morakkabati N. Dynamic bilateral
contrast-enhanced MR imaging of the breast: Trade-off between spatial and
temporal resolution. Radiology 2005;236(3):789-800.
2. Lustig M, Donoho D, Pauly JM. Sparse MRI: The application of
compressed sensing for rapid MR imaging. Magn Reson Med 2007;58(6):1182-1195.
3. Gamper U, Boesiger P, Kozerke S. Compressed sensing in dynamic MRI.
Magn Reson Med 2008;59(2):365-373.
4. Hargreaves BA, Saranathan M, Sung K, Daniel BL. Accelerated breast
MRI with compressed sensing. Eur J Radiol 2012;81S1:S54-55.
5. King K, Xu D, Brau AC, Lai P, Beatty PJ, Marinelli L. A new
combination of compressed sensing and data driven parallel imaging. Proceedings
of the Annual Meeting of ISMRM. Stockholm, 2010. (abstract 4881).