Takao Goto1, Miki Araki1, and Kenji Asano1
1MR Engineering, GE Healthcare, Hino-shi, Japan
Synopsis
Accurate
placement of a 2D plane across the aorta arch while examining scout images is a
complex task that makes the operator’s workflow difficult when bolus tracking in
DCE-MRI. We present a novel method for automated scan prescription for an oblique
plane delineating the entire aortic arch used to monitor bolus arrival. The
oblique plane was prescribed automatically by selecting the optimal oblique
angle using regression forests. A dataset with 31 volunteers was tested, and
all cases depicted the cross section of the aortic arch clearly. This
automation will assist the operator and decrease the total examination time.
PURPOSE
In bolus tracking, the operator prescribes an oblique plane through the
aorta. This prescription is a complex process, and its accuracy is highly
dependent on operator skill. Previously, we proposed an aorta detection method
(ADM)1, which automated oblique plane positioning through the
abdominal aorta. Since the aortic arch is also often used to detect bolus
arrival, we propose an automated method for prescribing a 2D oblique plane
through the aortic arch, thereby allowing us to offer a suite of workflow automation
techniques for abdominal bolus tracking.METHODS
2D SSFSE scout images were analyzed so as to not prolong the total
examination time. Since the aortic arch is close to the heart, it is
susceptible to motion effects due to heartbeats, resulting in flow artifacts
around the aorta. Furthermore, multiple vessels often cause false positives.
Therefore, we adopted a different approach from the ADM. Fig. 1 shows the algorithm
flow. First, the location of the liver dome peak in the superior/inferior (S/I)
direction is detected by an automated navigator tracker2 (Fig. 1a)
and the location of the abdominal aorta in the right/left (R/L) and
anterior/posterior (A/P) directions are detected by the ADM (Fig. 1b). This
permits the selection of an axial plane containing both the aorta and the liver
dome peak from a 3D abdominal dataset. In addition to the axial plane, a
reference point called the “anchor aorta” is detected from the 3D abdominal
dataset (Fig. 1c). Using the selected axial plane and anchor aorta as inputs,
an oblique plane crossing through the anchor aorta is reformatted by selecting
an optimal oblique angle using a machine learning technique called regression
forests (RF) 3 (Fig. 1d). Fig. 2a shows the definition of the
oblique angle in the axial plane. Our
training dataset was comprised of 31 volunteers whose optimal oblique angles were
evaluated by two trained experts, an MR
application specialist and scientist, to
generate the angle of ground truth (GT angle) in the oblique image. RF inputs were
created as follows. Each oblique plane was reformatted from a random selection
of 11 oblique angles between -30 and +30 degrees, with zero centered at the GT
angle. Then, a small image patch (41×53 pixels) was cropped from the oblique
plane, and its histogram of oriented gradients (HOG) feature vector was treated
as an input for the RF. This yielded a total of 3641 oblique angle and HOG
feature vector pairs. In the RF training step, the parameters of decision trees
in the RF were trained. We used 100 trees and 10 depth parameters. A
leave-one-out method was used for cross-validation. In the testing step, the
oblique plane was reformatted from an angle randomly selected from the range of
-30 to +30 degrees, with zero centered on the average of the GT angle, and followed
by HOG feature calculation. The RF outputs the value closest to zero as the
optimal angle of the oblique plane. Fig. 3 shows an example of how the contents
of oblique plane images (a-d) change as the oblique angle rotates about the
anchor aorta. In the case of Fig. 3, plane b is the most appropriate plane
position, delineating the entire aortic arch. The oblique plane matrix size was
set to 256×132, 380×270 mm field of view, and 10 mm slice thickness. Volunteer datasets
were obtained with informed consent and received institutional review and
approval.RESULTS
Our algorithm worked well for all 31 volunteers. The
distribution of the GT angles measured
from the 31 volunteers is shown in Fig. 2b. Here, the angle error is the
difference between the GT angle and the angle that the algorithm outputted. The
statistical values of the angle error and GT angle are given in Table 1. The
average and standard deviation of the GT angle were 113.4 and 8.7 degree, respectively.
Fig. 4 shows the comparison of the oblique images, GT angle (a,d,g), the average of the GT angles (b,e,h) and the algorithm results (c,f,i). The resulting oblique images
from the proposed algorithm were almost the same as the GT angle images.DISCUSSION
Although
it was difficult to obtain the sufficient delineation of the aortic arch by one
average GT angle, our proposed method is essential to cover the wide
range of possible GT angles. The results
were satisfactory in healthy volunteers.
However, an actual patient dataset may
include cases with deformation of the aorta and other organs. Therefore, it is
critical to make the algorithm robust against variations and other outlier
cases.CONCLUSION
We propose a novel method for the automated
prescription of oblique planes passing through the aortic arch, critical for
bolus tracking. This automation will decrease operator burden and shorten the total
exam time per patient.Acknowledgements
No acknowledgement found.References
[1]. Goto T. Araki M. ISMRM, 2016, Singapore, 248
[2]. Goto T. Kabasawa H. Magnetic Resonance
Imaging, 2015;33:63–71
[3]. Criminisi A. et. al. In: MICCAI 2010
workshop on medical computer vision: recognition techniques and applications in
medical imaging.