Daniel V Litwiller1, Valentina Taviani2, Lloyd Estkowski2, and Ersin Bayram3
1Global MR Applications & Workflow, GE Healthcare, New York, NY, United States, 2Global MR Applications & Workflow, GE Healthcare, Menlo Park, CA, United States, 3Global MR Applications & Workflow, GE Healthcare, Houston, TX, United States
Synopsis
Here we present a single-shot fast spin echo
sequence optimized for fast, motion-robust T1-weighted body imaging through the
use of inversion recovery preparation, and the introduction of variable
refocusing flip angle configured for centric, partial Fourier encoding.
Introduction
Single-Shot Fast Spin Echo (SSFSE or HASTE) is revered
for its speed, inherent motion robustness, and its excellent T2-weighted
contrast. T1-weighted imaging, of interest for routine body imaging, and specialized
applications such as whole-body imaging, is typically performed with
conventional interleaved multi-shot FSE, which tends to be relatively slow and
prone to motion artifacts, in spite of its otherwise exceptional image quality
(in terms of signal-to-noise and resolution).
For SSFSE, T1-weighting can be introduced with the addition of an
inversion recovery preparation pulse (SSFSE-IR) and a centric encoding scheme
to minimize the echo time (TE), the main drawback of this approach being the effective
doubling of T2-blur due to the interleaving of low-order phase encoding steps1. Here, we attempt to mitigate some of this
additional T2-blur for a T1w SSFSE-IR sequence with the introduction of
variable refocusing flip angle and partial Fourier to an inversion-recovery
prepared SSFSE pulse sequence (vrfSSFSE-IR) with centric encoding.Methods
A vrfSSFSE-IR pulse sequence was modified to
support centric, partial Fourier encoding and a complimentary refocusing flip angle
schedule, characterized by four flip targets: finit 130, fmin
60, fcen 100, and fend 130 degrees2. The centric partial Fourier encoding scheme
consisted of two regions, including an early, interleaved segment of low-order
phase encoding steps, and a late, linear train of high-order phase encoding
steps. The central flip target, fcen,
was located at the echo index corresponding to the end of the early, interleaved
portion, with the goal of minimizing T2-modulation across this early,
interleaved portion of the echo train. Extended
phase graph (EPG) simulation results are summarized in Figure 13. Following informed consent, several volunteers
were scanned on a 60-cm 3.0T MRI scanner (Discovery MR750, GE Healthcare,
Waukesha, WI). For the purposes of
demonstration, two body protocols comparing conventional T1w FSE and
vrfSSFSE-IR were performed, summarized in Table 1, including an axial acquisition
with whole-pelvis coverage, and a coronal 2-station acquisition with
liver-to-knees coverage. In the coronal,
multi-station acquisition, outer volume suppression in the L/R phase direction was
used to eliminate phase wrap from the subject’s arms4.Results
EPG simulation results in
Figure 1 for assumed T1/T2 values of 60/1500 ms demonstrate the effects of the
IR pulse on the resulting signal amplitude, and the subsequent effects of the
variable refocusing flip angles, including a leveling of transverse
magnetization across the interleaved portion of the echo train, and a
corresponding narrowing of the resulting point spread function in image
space. Initial in vivo results are
summarized in Figures 2 and 3, showing comparisons of conventional T1-weighted
FSE and vrfSSFSE-IR in the axial and coronal scan planes. Scan time for the vrfSSFSE-IR sequence was
reduced by a factor of 3 to 5 over conventional FSE, with the exact scan time
advantage conferred dependent on slice coverage, acceleration, phase wrap
factor and other factors summarized in Table 1. Discussion
These results demonstrate the promise of a fast,
motion-robust T1w SSFSE-based pulse sequence.
In general, the vrfSSFSE-IR sequence produces image contrast comparable
to conventional T1w FSE in substantially less time with fewer motion artifacts,
at the modest expense of lower signal-to-noise ratio. T2-mediated blurring is limited with vrf, and effective resolution is increased in the abdomen and other moving anatomical structures for vrfSSFSE-IR
due to its inherent ability to freeze motion. In
addition, owing to the short TE utilized in T1w imaging, and the short
corresponding signal coherence pathway (from excitation to TE), this T1w
version of vrfSSFSE-IR is not nearly as susceptible to signal loss and image
shading due to cardiac motion as its (long-TE) T2w vrf counterpart, potentially enabling
the use of lower minimum refocusing flip angles (fmin) to reduce specific
absorption rate and/or to further mitigate T2-blurring5. Further optimization and evaluation of the
clinical utility of this technique remains as future work.Conclusion
We have introduced an enhanced version of the
SSFSE-IR pulse sequence that utilizes variable refocusing flip angles and centric,
partial Fourier encoding to quickly acquire, motion-robust T1-weighted images
in the body. We believe this approach
may offer valuable workflow advantages without substantial compromises in image
quality.Acknowledgements
No acknowledgement found.References
1. Malamateniou C, McGuinness
AK, Allsop JM, O'Regan DP, Rutherford MA, Hajnal JV. Snapshot inversion
recovery: an optimized single-shot T1-weighted inversion-recovery sequence for
improved fetal brain anatomic delineation. Radiology. 2011 Jan;258(1):229-35.
2. Busse RF, Hariharan H, Vu
A, Brittain JH. Fast spin echo sequences with very long echo trains: design of
variable refocusing flip angle schedules and generation of clinical T2
contrast. Magn Reson Med 2006 May;55(5):1030-7.
3. Hennig,
J. Echoes—how to generate, recognize, use or avoid them in MR-imaging sequences.
Part II: Echoes in imaging sequences. Concepts Magn. Reson. 1991,3: 179–192.
4.
Taviani T, Litwiller D, Loening AM, Manojkumar S, Hargreaves BA, Vasanwala
SS. Single-Shot Fast Spin Echo of
Targeted Regions with Variable Refocusing Flip Angles and Quadratic Phase
Pulses for Outer Volume Suppression.
Proceedings of the ISMRM 2015.
5. Litwiller DV, Holmes JH, Saranathan M,
Loening AM, Glockner JF, Vasanawala SS, Bayram E. Sensitivity of modulated
refocusing flip angle single-shot fast spin echo to impulsive cardiac-like
motion. Proceedings of the ISMRM 2014.