Jacob Macdonald1, Philip Corrado1, Sydney Nguyen2, Kevin M Johnson1,3, Christopher J Francois3, Ian Bird2, Dinesh Shah4, Thaddeus G Golos2,4,5, and Oliver Wieben1,3
1Medical Physics, University of Wisconsin - Madison, Madison, WI, United States, 2Comparative Biosciences, University of Wisconsin - Madison, Madison, WI, United States, 3Radiology, University of Wisconsin - Madison, Madison, WI, United States, 4Obstetrics and Gynecology, University of Wisconsin - Madison, Madison, WI, United States, 5Wisconsin National Primate Research Center, University of Wisconsin - Madison, Madison, WI, United States
Synopsis
4D
flow measurements were performed with PC-VIPR in rhesus macaque monkeys
undergoing healthy pregnancies to determine the feasibility of flow
measurements in the fetal vasculature and umbilical cord. Flow measures
appeared to be viable for the larger fetal vessels (aorta and IVC), but more
variable in smaller vessels with slower flow (umbilical vessels). Image quality
improved for later gestational ages as a result of increased vessel area.
Purpose
Assessment
of blood flow to the placenta and to and from the fetus is of clinical and
research interest to evaluate fetal well-being. We recently demonstrated the
feasibility of 4D flow MRI for the assessment of flow in the utero-placental
vessels of the pregnant rhesus macaque1,2 In this pilot study, we
investigated the feasibility of 4D flow MRI for vessel visualization and the
integrity of subsequent quantitative flow measures in the rhesus macaque fetus.Methods
Eight
healthy, pregnant rhesus macaques were imaged in a right-lateral position on a
3.0T scanner (Discovery MR750, GE Healthcare) with a 32-channel cardiac or torso
coil. All procedures were approved by our institution’s animal care and use
committee (IACUC). The gestational ages of the monkeys ranged from early 2nd
trimester to early 3rd trimester. 4D flow imaging of the abdomen was
performed with a radially-undersampled sequence (PC-VIPR3, 5-point
velocity encoding, TR/TE=6.1/2.6ms; FA=8°; VENC=60cm/s; FOV=16x16x16cm3;
matrix=192x192x192; isotropic voxel size=0.83 mm, scan duration=610s). All
animals were sedated with isoflurane during imaging. Retrospective respiratory
gating was used to include only data during expiration (50% efficiency) for
image reconstruction. A time-averaged PC angiogram (PC MRA) was generated by
combining the magnitude and velocity data similar to complex difference
processing. Segmentation of the umbilical cord and fetal vessels was performed
on PC MRA images with MIMICS (Materialize). Time-averaged flow was measured in
the ascending aorta and inferior vena cava (IVC) of the fetus and the proximal
and distal ends of the umbilical cord within EnSight (CEI Inc.) by integrating
through-plane velocity components over the segmented vessel areas. A conservation
of mass approach was used to assess the internal consistency of PC-VIPR
measures by comparing flow in the ascending aorta against the IVC, and flow in
the umbilical vessels at both ends of the umbilical cord. Results
Vessel
visualization and flow measurements in the fetal aorta and IVC were feasible in
five of the monkeys, while measurements in the umbilical cord were possible in
four. The three animals that did not have sufficient SNR for detection of the
fetal vasculature were all early-to-mid 2nd trimester. Fig. 1 shows the segmented umbilical cord and fetal vasculature in
one rhesus. In all monkeys, only one of two
umbilical arteries was visualized within the umbilical cord (Fig 2). Fig. 3 shows a comparison of flow, velocity, and
diameter measures in the ascending aorta and IVC for each rhesus. The largest
difference in flow between these two vessels was only 12%. Figures 4 and 5 list
mean velocity, flow, and mean diameter measures in the umbilical artery and
vein, respectively. These measurements demonstrated less consistency, with an
average change in flow of approximately 20% between the two ends of the vessels.Discussion
Our
pilot study shows that 4D flow acquisitions in fetal and umbilical vessels in
the macaque are feasible in the later stages of gestation, from late 2nd
trimester onwards. Vessel area and flow were likely not high enough at earlier
stages to provide sufficient SNR to resolve vessels above the noise floor. Even
in early 3rd trimester, smaller vessels such as the second umbilical
artery were not separately resolved in the umbilical cord. Limits in spatial
resolution are likely the source of these difficulties, as measurements in
resolved vessels showed mean vessel diameters (approx. 4-5 mm) not much larger
than our spatial resolution (0.83mm). The fetal aorta and IVC
showed internal consistency of flow measurements, suggesting PC-VIPR is a
viable tool for measurements in these vessels. Measurements in the umbilical
vessels, however, were less reliable and not possible in four monkeys.
Partial-volume effects were likely enhanced in these vessels, as a larger
fraction of the voxels covering the vessel would have been affected given the
smaller vessel diameter. The sedation of the mother also sedated the fetus,
which was likely an important factor in the image quality of these
acquisitions, as it eliminated fetal motion during scanning. The challenge of
fetal motion will need to be addressed in future studies with human subjects
who are not sedated.Conclusion
This
feasibility study suggests that high resolution 4D Flow MRI acquisitions using PC-VIPR
are feasible in the aorta and IVC of the rhesus macaque fetus from late 2nd
trimester onwards. Assessment of the umbilical cord is more challenging and
might require improved spatial resolution. Image quality likely benefited from
the sedation of the mother and fetus. Acknowledgements
The authors acknowledge the support of the NIH Human Placenta Project
(NICHD U01HD087216) and NIH grant number P51 OD011106 to the Wisconsin
National Primate Research Center. We also thank GE Healthcare for their
support.References
1.
Macdonald J, Skopos S, Johnson K, Ludwig K, et al. Magnetic Resonance Imaging
of Utero-Placental Vascular Flow and Tissue Perfusion in Pregnant Rhesus
Macaques. Placenta 2016;85(6): 85.
2.
Macdonald J, Skopos S, Johnson K, Francois C, et al. 4D Flow Imaging of the
Placenta and Umbilical Cord in the Rhesus Macaque – Initial Experience. 28th
Society for Magnetic Resonance Angiography (SMRA). 2016.
3.
Johnson KM, Lum DP, Turski PA, Block WF, et al. Improved 3D phase contrast MRI
with off-resonance corrected dual echo VIPR. MRM. 2008;60(6):1329-1336.