Thomas Martin1,2, Dapeng Liu2, Teresa Chanlaw3, Sherin U. Devaskar3, Carla Janzen4, Tess Armstrong1,2, Yutaka Natsuaki5, Daniel Margolis2, Rinat Masamed2, Holden Wu1,2, and Kyunghyun Sung1,2
1Biomedical Physics, University of California, Los Angeles, Los Angeles, CA, United States, 2Radiological Sciences, University of California, Los Angeles, Los Angeles, CA, United States, 3Department of Pediatrics, University of California, Los Angeles, Los Angeles, CA, United States, 4Department of Obstetrics and Gynecology, University of California, Los Angeles, Los Angeles, CA, United States, 5Siemens Healthcare, Los Angeles, CA, United States
Synopsis
Proper
placental function is essential for normal fetal development. MRI can easily assess the growth and
development of the placenta throughout gestation due to its capabilities in
functional imaging. However,
uterine contractions, and fetal and maternal motion can lead to inaccuracies in
the quantitative assessments. In this we assessed the impact of
extraneous motion on placental imaging by using a 3D multi-echo golden angle
radial sequence to generated dynamic images of the placenta.
Introduction
Proper
placental function is essential for normal fetal development. The placenta is
known to transfer nutrients to the fetus with its high blood volume1. MRI can easily assess the growth and
development of the placenta throughout gestation due to its capabilities in
functional imaging2. However,
uterine contractions, and fetal and maternal motion can lead to inaccuracies in
the quantitative assessments. It is not well documented on how much these
motion characteristics weigh in and need attention in the final assessment. In this study we aim to assess the impact of
extraneous motion on placental imaging by using a 3D multi-echo golden angle
radial sequence to generated dynamic images of the placenta.Methods
To
assess the effect of motion on the placenta 7 healthy pregnant subjects were
scanned at two different times in their gestation (first scan 14-16 weeks,
second scan 19-22 weeks) for a total of 14 scans
on
a 3T scanner (Prisma, Siemens Healthcare, Erlangen, Germany). A 3D multi-echo golden angle (GA) radial GRE
sequence was used to acquire the raw data, which was then reconstructed offline
using k-space weighted image contrast (KWIC)3 to create high spatial
resolution dynamic image series without much streaking. There were two main protocols used for
imaging and are listed in table 1. The
multi-echo imaging was used for R2* quantification, but
only the first echo that imparted the highest signal was used for
reconstruction. Motion was
semi-quantitatively assessed by assessing the absolute difference between the
first image of the timed series and all other subsequent images in the time
frame of scanning (see figure 1). Two regions-of-interests (ROI) were drawn to segment the placenta
and the outer uterus on three orthogonal slices. A T2-HASTE image was used as
reference for the placental location. To
reduce the effect of noise and streaking artifacts the subtraction image was
filtered using a window leveling so that majority of the remaining non-zero
pixels correlated with motion. The number of pixels were then counted and
normalized to derive the percent change in the # of pixels across time and then
averaged across the entire scan time frame. Results
Figure
1b shows that there is visible motion between the first and last images of the
dynamic series of a subject at 21 weeks gestation, and figure 1c shows that the
motion timed series is in agreement with the subtraction image. The sharp rise
near t = 50s is due to noise build up between the reference time point and
later timed images, and not necessarily motion. Figure 2 shows a different subject
at 21 weeks in whom the uterus expands and then contracts during the middle of
the scan across a 2min interval. It
shows that the motion is not abrupt, but smooth over a long period of time. Figure 3 shows the mean percent change of
pixels for the total scan time (4-6min) as well as half of the total scan time
(2-3min). There is a noticeable decrease
in the mean percent pixel change for scan time of 2-3min compared to longer
scan time. There is also an overall slight trend of decreased motion throughout
the scan as the gestation weeks increase. The largest amount of pixel change
was about 22%.Discussion
In
this study we have shown a semi-quantitative analysis of uterine and placental motion
throughout gestation. The average range
across the total scan time of the percent pixel change was about 5-15%. However, due to the noise accumulation there
was a sharp rise in the percent pixel change for every case around t=50s. This could be improved by removing the noise
bias. Also by reducing the scan time the
amount of uterine motion is decreased. However,
there was still a significant amount of motion detected that could be
problematic in interpretations.Conclusion
We
have demonstrated that there is placental and uterine motion in the first half
of gestation under long scan times (>4min).
Decreasing the scan time can reduce the motion within the scan
time. However, for perfusion imaging
with multiple scans, motion correction should be considered.Acknowledgements
Funding for project is provided by NICHD U01-HD087221.References
1.
Rossant et al., NRG 2001
2.
P. Gowland, Seminars in Fetal & Neonatal Medicine 2005
3.
Song et al., MRM 2004