Ping Wang1, Henry Zhu1, and John C. Gore1
1Radiology and Radiological Sciences, Vanderbilt University Institute of Imaging Science, Nashville, TN, United States
Synopsis
Previous studies have indicated that R1ρ (= 1/T1ρ)
dispersion may be dominated by chemical exchange processes at higher fields, and
the dispersion may be used to quantify exchange processes between water and
labile protons, mainly amides and hydroxyls.
At 3T some biological tissues with high macromolecular protein content may
demonstrate a considerable dispersion, but fat and water rich tissues typically
have a negligible R1ρ dispersion.
In this study, we observed that the degree of R1ρ dispersion
in white matter at 7T was more than three times the dispersion at 3T, confirming that
chemical exchange is a major contribution to R1ρ and suggesting R1ρ
dispersion at higher fields may help to characterize tissue physicochemical
properties.
Introduction
R1ρ (= 1/T1ρ) is sensitive to slow macromolecular
interactions but also may vary with the spin-locking fields used. At higher magnetic fields (3T and beyond),
this dispersion may be dominated by chemical exchange processes so that the
dispersion may be used to characterize tissue composition and physicochemical
properties.1-3 The degree of R1ρ dispersion is
tissue specific and depends on the composition and exchange rates of
constituent molecules that exhibit chemical exchange between water and labile
protons (mainly amides and hydroxyls). If
chemical exchange is significant the magnitude of the dispersion should depend strongly
on the magnetic field. In this study, we
compared the R1ρ dispersion in human brain at 3T and 7T scanners. At 3T there is a small dispersion in both
white matter and gray matter in the brain, but this dispersion increased significantly
(by more than three fold) in white matter at 7T, whereas R1ρ
dispersion in gray matter remains similar at both magnetic fields. This observation suggests that R1ρ
dispersion at 7T (ΔR1ρ, the difference of R1ρ between low
and high locking fields) is dominated by chemical exchange processes in white
matter, which may be sensitive to subtle changes in composition or pH.Methods
Six healthy volunteers (aged
24 to 37 with a median age of 29.5) were recruited in this study. Two subjects
participated in 3T imaging and the other four were imaged at 7T. Both the scanners were Philips Achieva
systems (Philips Healthcare, Best, the Netherlands) and the same strategy was
used for T1ρ data acquisition (T1ρ pre-pulse4 followed by a TSE readout). The 3T T1ρ imaging was acquired
using a Philips 8-channel head coil, with FOV=220 x 220mm2, pixel
size = 2 x 2mm2, thickness = 3mm, TR/TE = 5500ms/10ms, NSA = 1. Five spin-locking times (TSL) [2ms, 22ms,
42ms, 62ms, 82ms] were combined into a single scan for T1ρ
calculations, resulting in a scan time of 3min 18sec. The T1ρ experiment was repeated at
different spin-locking fields (FSL) [0Hz, 50Hz, 100Hz, 300Hz, 500Hz] to
construct the dispersion curve. The 7T T1ρ
imaging followed the same procedures but used a 32-channel NOVA head coil. Other parameters were: FOV = 240 x 240mm2,
pixel size = 1 x 1mm2, thickness = 4 mm, TR/TE = 8000ms/55ms, NSA = 1,
each T1ρ acquisition took 2min 48sec (with SENSE acceleration). After acquisition, an R1ρ map at
each spin-locking frequency was calculated by fitting the signal intensity vs TSL to a three-parameter
mono-exponential model on a pixel-wise basis.
The R1ρ dispersion was assessed on both white matter and gray
matter regions. To quantify the degree
of a dispersion curve, a relative dispersion coefficient (RDC) was defined as:
RDC = (R1ρ,max - R1ρ,min)/(R1ρ,max + R1ρ,min),
where R1ρ,max and R1ρ,min represent the maximum and
minimum of the R1ρ values on the dispersion curve.Results
Figure 1 shows the
comparison of R1ρ maps between 3T and 7T at each FSL, it is seen
that the overall R1ρ value at 7T is great than 3T. Also because the 7T scanner is more susceptible
to field inhomogeneities, the R1ρ map at 7T appears less uniform than
3T. Figure 2 indicates the selected
regions of interest (ROIs) from white matter and gray matter, and the
dispersion curves for these regions are plotted in Figure 3. It is clear that more pronounced R1ρ
dispersion is observed in white matter at 7T over 3T, whereas the dispersion in
gray matter remains similar between the two magnetic fields. Through quantitative analyses (shown in Table
1), the relative dispersion coefficient in white matter at 7T is more than three
times that at 3T (average RDC of 0.4 vs
0.1175), while they are close in the gray matter region (average RDC of 0.16 at
7T vs 0.132 at 3T).Discussions
Biological tissues containing high concentrations of amides, amines,
hydroxyls or other protons that exchange at appropriate rates may demonstrate a
considerable dispersion in R1ρ, but fat and water rich tissues
typically have a negligible R1ρ dispersion. It has been shown in this study that the R1ρ
dispersion in white matter at 7T is much greater than 3T, while the dispersion
at the two fields remains similar in gray matter region, confirming that
exchange is significant in white matter.
Because gray matter contains more water and less myelin (and the
associated hydroxyl groups of e.g. cholesterol), chemical exchange processes may
be less relevant compared to white matter.
For tissues with appropriate composition of exchangeable protons, R1ρ
dispersion at higher field is expected to better characterize the tissue
composition and physicochemical properties. Acknowledgements
No acknowledgement found.References
1. Cobb J, Xie J, Gore J. Contributions of chemical exchange to T1rho dispersion in a tissue model. MRM. 2011; 66(6):1563-1571.
2. Cobb J, Xie J, Li K, et al. Exchange-mediated contrast agents for spin-lock imaging. Magn Reson Med. 2012; 67(5):1427-1433.
3. Wang P, Block J, Gore J. Chemical Exchange in Knee Cartilage Assessed by R1ρ (1/T1ρ) Dispersion at 3T. MRI. 2015; 33(1):38-42.
4. Witschey W, Borthakur A, Elliott M, et al. Artifacts in T1ρ-Weighted Imaging: Compensation for B1 and B0 Field Imperfections. JMR. 2007; 186(1):75-85.