Chengcheng Zhu1, Bing Tian2, Sinyeob Ahn3, Esther Raithel4, Gerhard Laub3, and David Saloner1
1Radiology, University of California, San Francisco, San Francisco, CA, United States, 2Radiology, Changhai Hospital, Shanghai, People's Republic of China, 3Siemens Healthcare, CA, United States, 4Siemens Healthcare, Erlangen, Germany
Synopsis
Current
3D black-blood high-resolution MRI (0.4-0.6mm isotropic) of intracranial vessel
wall is limited by long scan times (~10 minutes). This study implemented a
compressed sensing SPACE (CS-SPACE) sequence to reduce the scan time. The scan
and reconstruction parameters were optimized in volunteers and then validated
in patients. The optimized CS-SPACE protocol achieved good image quality and
reliable vessel area measurements compared with SPACE, with a 37% time
reduction. 0.5mm isotropic resolution can be achieved in <7 minutes, and
0.6mm3 is possible in 4 minutes. This fast intracranial vessel wall
technique has potential for use in a clinical setting.
Purpose
3D
high resolution black blood MRI has been widely used for imaging of vessel wall
pathology of intracranial vascular diseases, including atherosclerotic plaques,
aneurysms and vasculitis. High isotropic resolution (0.4mm to 0.6mm) has been
achieved with good blood suppression and image quality. However, these
protocols often need a long scan time of around 10 minutes, which is poorly
tolerated by patients 1-2. Acceleration techniques that significantly
reduce the scan time while maintaining image quality would be very valuable.
This study aims to evaluate and optimize compressed sensing methods for fast
intracranial vessel wall imaging.Methods
Sequence: A compressed sensing
fast-spin-echo sequence (CS-SPACE) was implemented on a Siemens 3T scanner. A
Poisson-disk k-space pattern and symmetric radial k-space sampling trajectory were
used (Figure 1). Online
reconstruction was implemented and took ~2 minutes (20 iterations) for current
CS-SPACE protocols. The scan and reconstruction parameters were first optimized
in 5 healthy volunteers, and then the optimized protocol was evaluated on 5
patients with cerebral vascular disease (3 intracranial aneurysms, 1 vertebral
artery dissection, and 1 vertebral artery occlusion.) Patients were scanned with
SPACE and CS-SPACE both pre- and post-Gd contrast injection. The original SPACE
has whole brain coverage and 0.5mm isotropic resolution (Table 1). CS-SPACE scanning parameter optimization:
Echo train length from 40 to 60, and under-sampling factors from 14% to 20%
were tested. Reconstruction parameter optimization: Iteration times from 10
to 40 and regularization factor from 0.0005 to 0.004 were tested. Image
analysis: For volunteers, three
major intracranial arteries (bilateral middle cerebral and internal carotid
arteries, and basilar artery) were included in analysis. Qualitative scores were
graded by an experienced radiologist (0-4 scale). Quantitative area and
contrast ratio (wall/lumen) were measured. Image scores were evaluated in both
sagittal and axial planes. For patients, the qualitative image quality and characterization
of contrast enhancement were evaluated. Results
The quantitative
comparison is shown in Figure 2 and
qualitative image scores are shown in Figure
3. Generally there is good agreement for area
measurements between CS-SPACE and SPACE, and there is no difference in
wall-lumen contrast ratio. The highest performing scanning protocol has ETL 60
and 20% under-sampling factor. Reconstruction tests showed 20 iterations times
and regularization factor of 0.002 provided the best image quality. The
optimized protocol achieved comparable image quality to the fully sampled SPACE
with a scan time reduction of 37% (6:48 vs. 10:50). Sample patient images are shown
in Figure 4. There is no significant
difference in the image quality between CS-SPACE and SPACE. SPACE and CS-SPACE
had perfect agreement on identifying wall enhancement (3 of 5 patients demonstrated
enhancement). We also tested a 0.6mm isotropic protocol which can be done in ~4
minutes with good image quality. Discussion
To our
knowledge, this is the first study to apply compressed sensing to 3D black
blood intracranial vessel wall imaging. Our optimized CS-SPACE protocol shows
good image quality and reliable area quantification compared to SPACE, in a
scan time of less than 7 minutes (37% reduction in scan time). 0.6mm isotropic
imaging is possible with CS-SPACE in as little as 4 minutes. Fast imaging of
intracranial vessel wall using CS-SPACE is a promising tool to evaluate
patients with intracranial vascular disease in the clinical setting, and has
the potential to improve risk stratification in these patients.Conclusion
Compressed
Sensing SPACE can achieve high isotropic resolution whole brain intracranial
vessel wall imaging with comparable image quality to SPACE, while significantly
reducing the scan time. This technique has the potential
to be used in a clinical setting for the evaluation of patients with cerebral vascular
diseases. Acknowledgements
This study is supported by National Institute of Health (NIH) grants R01HL114118 and R01NS059944References
1. Fan
Z et al, Whole-brain intracranial vessel wall imaging at 3 Tesla using
cerebrospinal fluid-attenuated T1-weighted 3D turbo spin echo. MRM 2016
2. Zhu C et al, High resolution imaging of
the intracranial vessel wall at 3 and 7 T using 3D fast spin echo MRI. MAGMA
2016