Katsuhiro Inoue1, Maki Umino2, Ryohei Nakayama3, Shinichi Takase1, Tsunehiro Yamahata1, Makoto Obara4, Masayuki Maeda5, and Hajime Sakuma6
1Department of Radiology, Mie University Hospital, Mie, Japan, 2Department of Radiology, Mie University School of Medicine, Mie, Japan, 3Department of electronic and Computer Engineering, Ritsumeikan University, Shiga, 4Philips Electronics Japan, Tokyo, Japan, 5Department of Advanced Diagnostic Imaging, Mie University School of Medicine, Mie, Japan, 6Department of Radiology, Mie University School of Medicine, Tsu, Mie, Japan
Synopsis
Evaluation of parotid
blood flow (PBF) may provide an insight into the pathophysiology of parotid
diseases. We evaluated and compared the blood flow of normal parotid glands in
six volunteers using 2D-echo-planar (EP) and 3D-turbo-spin-echo (TSE) pseudo-continuous
ASL (pCASL) sequences. Visual assessment indicated that 3D-TSE pCASL maps were
significantly better for delineating normal parotid glands. 3D-TSE pCASL showed larger PBF values than 2D-EP
pCASL. Fewer measurement errors were observed in 3D-TSE pCASL data. We concluded
that 3D-TSE pCASL is more suitable for the evaluation of PBF than 2D-EP pCASL.
INTRODUCTION
Arterial spin labeling (ASL)
facilitates the noninvasive quantitative assessment of tissue perfusion without
a contrast agent and is now used in clinical settings. However, the use of ASL
to assess perfusion in normal and diseased parotid glands is limited [1, 2]. In
previous studies, researchers have used different ASL sequences such as two-dimensional
echo-planar (2D-EP) and three-dimensional turbo spin-echo (3D-TSE) [1, 2]. However, the literature has not yet identified
which ASL sequence is better for evaluating parotid gland perfusion.PURPOSE
The purpose of this
study is to evaluate the blood flow in normal parotid glands using two types of
pseudo-continuous ASL (pCASL) sequences, 2D-EP and 3D-TSE, and determine which
is better for evaluating parotid perfusion in clinical settings.MATERIALS and METHODS
For this study, we used a
3T MRI (Ingenia, Philips Healthcare, Best, Netherlands) with a dS
HeadNeckSpine coil. We measured the blood flow of normal parotid glands in six
volunteers (aged 24 ±
2.9 years, three men and three women) using 2D-EP and 3D-TSE pCASL. The 2D-EP pCASL
parameters were as follows: TR/TE = 4500/14, number of excitations = 36, flip
angle = 70º, matrix = 80 × 80, field of view (FOV) = 240 mm,
label type = parallel slab, label gap = 73.36 mm, label duration = 1,650 ms,
post-labeling delay = 1,800 ms, and scan time = 5 min 33 s for 22 slices. The 3D-TSE
parameters were as follows: TR/TE = 6000/40, number of excitations = 3, flip
angle = 90º, matrix = 80 × 80, FOV = 240 mm, label type =
parallel slab, label gap = 73.36 mm, label duration = 1,650 ms, post-labeling
delay = 1,800 ms, and scan time = 5 min 36 s for 44 slices.
Labeling
and control-dynamic data were transferred to a personal computer, processed,
and analyzed for parotid blood flow (PBF) calculation using quantitative
measurement software (C++ programming
language, Philips Healthcare Korea, Republic of Korea). A region of interest
(13 pixels) was identified within the normal parotid gland. Standard deviation
within the parotid glands was used for the analysis of measurement errors.
Two reviewers reached a
consensus on the visual assessment of the PBF image for each pCASL map. A four-point
scoring system was used: excellent (4), good
(3),
fair
(2),
or poor
(1).
We compared PBF quantitative data and the images obtained using PBF
between 3D-TSE pCASL and 2D-EP pCASL. We used a Wilcoxon signed-rank test for
statistical analysis.RESULTS
Visual assessment
indicated that 3D-TSE pCASL maps were significantly better for delineating
normal parotid glands (p < 0.01, Figs. 1, 2). 3D-TSE pCASL showed larger PBF
values than 2D-EP pCASL (50.3 ± 8.5 vs. 37.7 ± 5.9 ml/min/100 mg; p < 0.01,
Fig. 3). Standard deviation was significantly smaller in 3D-TSE pCASL compared
with that in 2D-EP pCASL (13.6 vs. 25.6; p < 0.01, Fig. 4), indicating fewer
measurement errors in 3D-TSE pCASL data.DISCUSSION
Blood-flow measurements
in normal parotid glands were significantly higher in 3D-TSE pCASL than those in
2D-EP pCASL. This was probably due to the differences in readout and dimensions
between the two techniques. 3D-TSE pCASL showed better delineation of the
parotid gland and fewer measurement errors, which was likely due to a higher
signal-to-noise ratio and less image distortion on 3D-TSE pCASL maps. Therefore,
we conclude that 3D-TSE pCASL is more suitable for the evaluation of PBF than 2D-EP
pCASL.Acknowledgements
No acknowledgement found.References
1. Kato
H, et al. Perfusion
imaging of parotid gland
tumours: usefulness of arterial spin labeling for differentiating Warthin's
tumours. Eur Radiol. 2015;25(11):3247-54.
2. Kami
YN, et al. Arterial spin
labeling imaging for the parotid
glands of patients with Sjögren's syndrome. PLoS One.
2016;9;11(3):e0150680