Ana L Gomes1, Paul Kinchesh1, Stuart Gilchrist1, Alex Gordon-Weeks1, Ruth J Muschel1, and Sean C Smart1
1CRUK/MRC Oxford Institute for Radiation Oncology, University of Oxford, Oxford, United Kingdom
Synopsis
Prospective gating
and automatic reacquisition of data corrupted by respiration motion were
implemented in 3D balanced steady-state free precession (bSSFP) to provide
contrast that enables unambiguous detection of liver tumours in the whole mouse
liver with 200 µm isotropic resolution and in scan times that are routinely
less than 7 minutes. The method was used for orthotopic tumour burden
quantification in 8 female
C57BL/6 mice
at days
7, 9, and 11 post intra-hepatic injection of MC38-GFP cells, and enabled measurement
of tumour volumes less than 1 mm3.
Purpose
The balanced steady-state free precession (bSSFP) scan mode
is of particular interest because it is fast and the inherent ‘T2’ contrast causes many pathologies
to appear bright. Cardio-respiratory (CR) synchronisation, k-space segmentation
and short constant TR have recently been applied to 3D bSSFP to reduce
imaging times below what is otherwise achievable with standard techniques such
as retrospective gating.1 This abstract demonstrates that the method enables robust imaging
of liver tumours in the mouse with 200 µm isotropic resolution in less than 7 minutes
and is therefore a suitable tool for the routine screening of disease
development and monitoring of in vivo tumour response to therapy.Methods
Anaesthesia was
induced with 4% isoflurane in air and maintained with 1-3% isoflurane in a 1:5
O2:air mixture for surgery and MRI. 1x104 to 5x105
MC38-GFP tumour cells were injected directly into the liver of 8 female C57Bl/6
mice to generate orthotopic tumours.2 MRI was performed at days 7,
9, and 11 post inoculation at 7.0 T (Varian VNMRS), using a 30 mm long 25 mm ID
quadrature birdcage (Rapid Biomedical). Respiration was monitored using a
pressure balloon. ECG needles were placed subcutaneously in the chest. CR-synchronised
bSSFP scans were performed volumetrically with automatic and immediate
reacquisition of data corrupted by respiration motion. Scan parameters were TR
2.8 ms, TE 1.4 ms, RF hard pulse 16 µs, FA 30º, FOV 51.2×25.6×25.6 mm3,
matrix 256×128×128, and 32 k-lines per R-wave to give a single 3D data set with
200 µm isotropic resolution in less than 105 s. bSSFP banding artefacts were
robustly removed with combination of four phase-cycled images acquired in less
than 7 minutes in total using an elliptical signal model.3 CR-synchronised
RF and gradient spoiled 3D gradient echo scans were acquired with identical
scan parameters apart from TE 1.15 ms, FA 8º, and the acquisition of 4
identical repeats to give a comparative scan of essentially the same duration.
The bSSFP scan is marginally longer to enable each phase-cycled image to reach
steady-state before the start of acquisition. Liver tumour volumes were
calculated using the threshold-based automatic segmentation provided by ITK-SNAP.4Results
Examples of
orthogonal views through the tumour are shown for mouse 4 at 9 days post
intra-hepatic injection in Figures 1 and 2 for the bSSFP and gradient echo scan
modes respectively. Figure 3 shows the tumour volumes of 8 mice which varied
from 0.824 mm3 to 160.7mm3. Tumour growth was observed
between successive observations in all 8 mice. Figure 4 shows slices through
the tumour in bSSFP scans of mouse 4 at 7, 9 and 11 days post intra-hepatic
injection. Photographs of the excised liver confirm the presence of the tumour
and the bottom view replicates the tumour shape observed in the MRI slices.Discussion
In both scan modes the
automatic and immediate reacquisition of data corrupted by respiration motion
robustly eliminates respiration motion artefact. Contrast is generated over a
timescale of up to several T1
but maintained with a timescale of TR such that synchrony with the cardiac
R-wave can always be achieved to within one TR which is 2.8 ms in this instance.
It is evident that application of the bSSFP scan mode enables unambiguous
delineation of tumour within the liver. Although the gradient echo scan is more
robust with respect to movement of spins since phase coherence of magnetisation
is only required during one TE, pure T1
contrast is not able to distinguish between tumour and, for example, gall
bladder. Due to the lack of accurate and fast in vivo orthotopic tumour volume
quantification methods, most therapy response studies are done in subcutaneous
models, which do not entirely mimic the normal tumour microenvironment. It has
been demonstrated that the bSSFP scan mode allows rapid assessment of
orthotopic liver tumour volumes. Furthermore the method would enable quantification
of response to anti-cancer agents and/or other therapies in vivo, and could be applied
to a variety of other disorders.Conclusion
Cardio-respiratory
gated 3D bSSFP enables unambiguous detection of orthotopic liver tumours in the
mouse with 200 µm isotropic resolution and in scan times that are routinely
less than 7 minutes.Acknowledgements
The work presented was
supported financially by Cancer Research UK (CRUK grants C5255/A12678,
C2522/A10339), the Engineering and Physical Sciences Research Council (EPSRC
grant C2522/A10339) and the Medical Research Council Unit Grant for the Oxford
Institute for Radiation Oncology.References
1. Kinchesh P,
Gilchrist S, Gomes AL, Kersemans V, Beech J, Allen D, Smart S. Accelerated
imaging of the mouse body using k-space segmentation, cardio-respiratory
synchronisation and short, constant TR: Application to b-SSFP. Proc Intl Soc Mag Reson Med 2016;24:1825.
2. Rao Q, You A, Guo Z, Zuo B, Gao X, Zhang T, Du Z, Wu C, Yin H. Intrahepatic tissue implantation
represents a favorable approach for establishing orthotopic transplantation hepatocellular
carcinoma mouse models. Plos One 2016;11:e0148263.
3. Xiang Q-S, Hoff MN. Banding artifact removal for bSSFP imaging with an
elliptical signal model. Magn Reson Med 2014;71:927-933.
4: Yushkevich PA,
Piven J, Hazlett HC, Smith RG, Ho S, Gee JC, Gerig G. User-guided 3D active
contour segmentation of anatomical structures: Significantly improved
efficiency and reliability. Neuroimage
2006;31:1116-1128.