Arne Berneking1, Adam Mehina, and N. Jon Shah
1Institute of Neuroscience and Medicine, Forschungszentrum Juelich GmbH, Juelich, Germany
Synopsis
The goal of this study is to demonstrate hybrid MR/PET scanner design
optimization methods from an RF coil perspective with a focus on UHF MRI. Here,
coil performances are investigated depending on different shielding distances
to present simulation and measurement methods to include RF coil performances
into the hybrid MR/PET scanner design. Moreover, integration of PET detector
shielding and RF coil shield is investigated. The results of this study clarify
that a trade of between a compact design, PET and RF coil performances is
necessary and that RF coil performances can be optimized by an integrated
shielding with an optimized configuration.
Purpose
In hybrid MR/PET scanner design an RF shield is commonly installed
between the RF coils and the PET data acquisition systems [1-2]. This prevents
electromagnetic (EM) coupling between the two parts by shielding the interior
of the cassette from the EM field generated by the RF coils. In whole-body hybrid
MR/PET scanners the space available for a PET ring is significantly smaller. As
such, a compact design is essential and is driven by the fact that the already
high price of a hybrid scanner can be reduced with a reduction in the bore size.
The bore diameter is directly proportional to the magnet price, which is the
dominant cost in an MRI system. This aspect is even more critical for whole
body, ultra-high field (UHF) hybrid scanners, such as the 7 T and 9.4 T
variations. Furthermore, a compact PET ring close to the subject increases PET
scanner sensitivity and reduces the amount of required detectors. When the PET detectors
with its electronics are positioned as close as possible to the object to be
imaged, the RF shielding of the PET electronics will be in close proximity to the
RF coil. To reduce the total required volume, the PET electronics shielding and
the RF coil shield can be integrated into a single component. However, the
shielding design has a significant influence on the RF coil performance. The
aim of this study is to investigate the influences of the shielding on typical
RF coil performance for hybrid scanners and to compare the results from a 4 T
scanner with those from an UHF 9.4 T scanner.
Methods
To evaluate coil performance for different shielding configurations, the
B1+ field, specific absorption rate (SAR), and quality factor (Q value) were
simulated and measured. First, CST microwave studio simulations investigated
the dependence of B1+ field strength and SAR in the human head on the distance from
the shielding to a surface coil. Simulations for both 168 and 400 MHz were
performed. The simulation setup consisting of a human head, surface coil, and
planar copper shield is demonstrated in Figure 1. The same simulation was then
repeated with the planar copper shield replaced by the PET cassettes which are covered
by copper plating. The measurement setup for analysing changes in the Q value
is demonstrated in Figure 2. Q value measurements were performed with tuned and
matched conditions for a surface coil. The Q value measurements were taken for
both unloaded and loaded coils. After every adjustment to the shielding
configuration, the coil was retuned and rematched to the new environment.Results
Figures 3 shows the normalized B1+ field of the coil depending on the
shielding distance as well as the normalized B1+ field if PET cassettes are
used as an RF coil shield. Figure 4 demonstrates the dependency of the SAR on
the cassette distance. In Figure 5 the unloaded and loaded Q value is
demonstrated for 4 T and 9.4 T.
Discussion
As demonstrated in Figure 3, the B1+ field profile inside the human head
changes with an increase of the frequency. The less homogeneous B1+ field at
400 MHz versus at 168 MHz is due to a decrease in wavelength - which is already
in the range of the observed object. While at 4 T the highest B1+ can be
achieved without a shield, where the RF coil is not coupling to a shield this
behaviour was found to change at 9.4 T. At 400 MHz radiation losses have a more
significant impact. It can be derived from Figure 3 that a shielding distance
of 5 cm shows the highest b1+ field as well as a deeper RF penetration depth. Replacing
the RF shield by PET cassettes, the B1+ field distribution changes. In Figure
4, it can be seen that for 4 of the 6 distances simulated the highest peak SAR
increased in comparison to the no shield case. As observed in Figure 5, the
shield distance also influences the Q value of the coil. The optimal distance
changes with frequency. At 4 T, a shorter distance shows the highest Q value.
At 9.4 T, the optimal distance for highest Q value is increased.Conclusion
The work presented in this study shows the influence of shielding
configurations on RF coils in hybrid MR/PET scanners. A consequence for compact
MR/PET scanner design is that size reduction and PET performance such as sensitivity
and spatial resolution detection cannot be optimized separately from the
performance of the RF coils of the MRI scanner.Acknowledgements
No acknowledgement found.References
[1] A. Berneking, R. Trinchero, P. Cerello, C. W. Lerche, and N. J. Shah
“Low-Pass Shielding Design for MRI Applications Optimized for Strong RF
Shielding Effectiveness”, presented at the ESMRMB conference, Vienna, Germany,
September 29–October 1, 2016.
[2] A. Berneking, R. Trinchero, N. J. Shah, P. Cerello, and C. W.
Lerche, “Design and Characterization of a Frequency Selective RF Shield for PET
Detector Modules in Hybrid MR-PET Imaging”, presented at the PSMR conference,
Cologne, Germany, May 23–25, 2016.
[3] H. Herzog, et al., “High resolution BrainPET combined with
simultaneous MRI”, Nuklearmedizin, vol. 50, pp.74–82, Feb. 2011.