Irena Zivkovic1, Iliya Tolstikhin2, Bernhard Schoelkopf2, and Klaus Scheffler1,3
1High Field MRI, Max Planck Institute for Biological Cybernetics, Tuebingen, Germany, 2Empirical Inference, Max Planck Institute for Intelligent System, Tuebingen, Germany, 3Dept. for Biomedical Magnetic Resonance, University of Tuebingen, Tuebingen, Germany
Synopsis
At high magnetic fields B0
inhomogeneities are more pronounced and B0 shimming becomes very important.
Dynamic or slice based shimming always provides better results than global or
shimming in a volume. For shimming in a small volume, we propose a setup composed
of irregularly shaped coils. Performances of the proposed setup are compared to
performances of the configuration composed of only loop elements. The setup with
irregular coils shows improved performances over the setup containing only loop
elements. Future work should validate performances of the proposed setup on
different volume sizes and different brain’s B0 maps.
Introduction
B0
shimming is very important especially at high magnetic fields. The
susceptibility differences between tissues and air cavities increase for
increasing main magnetic field which cause different image artifacts. A method to
reduce B0 inhomogeneities has been proposed recently1,2 in the form of a close fitting array of
conductive loops placed between the RF coil and head. In some recent
publications3,4 optimizations of the individual coil’s geometry have
been proposed in order to further improve B0 shimming performance. Dynamic or
slice-based shimming always provides better results than global shimming or
shimming on a certain volume. In this work we present numerical studies of
the shimming performance in a small volume of a setup composed of irregularly
shaped coils and compare it to configurations composed of only loop elements.Methods
We synthesized a 16 channel configuration
with 15 irregularly shaped coil elements and one circular loop (loop was placed
around shim cylinder) by optimizing jointly independent parameters describing the
geometry and position of every single coil (for a loop only its position was
used as a parameter). Coils were placed on a cylinder which length was 300mm
and diameter 360mm. The optimization was performed on three slices
simultaneously (Figure 1) based on a B0 field map acquired within a brain of a
volunteer. The resulting setup is shown in Figure 2. The second 16 channel
setup contained only loop elements. The loops were evenly distributed in two
rows and the diameter of each loop was 120mm (Figure 2). The dimensions of a
cylinder containing only loop elements were the same as dimensions of cylinder
with irregularly shaped coils.
The 30mm thick volume that
includes slices 1, 2 and 3, was shimmed with both setups. The standard
deviation on slices 1, 2 and 3 was then calculated (Table 1, Figure 3). The
slices 1, 2 and 3 were also shimmed individually with both setups and results
are in Table 2.Results and discussion
Three representative slices were observed for both volume and
slice based shimming. In volume shimming, shim performance of the setup with
irregularly shaped elements was improved compared to the setup with only loop
elements (Table 1, Figure 3). Improvement was around 15% on slices 1 and 3 and
60% for slice 2.
In slice based shimming, both
setups performed similarly (Table 2). Setup containing only loop elements
performed much better when shimming on a single slice than on a volume. Setup
containing irregularly shaped elements performed similarly for both volume and
single slice shimming. Conclusions
In high field MRI, global
shimming and shimming in a certain volume does not produce satisfying results. In this preliminary study, we presented an optimized
setup composed of irregularly shaped elements designed to improve shimming
performance on a chosen volume. The performance of the proposed setup is improved
compared to the configuration with only loop elements. Future
work will validate performances of the proposed setup with irregularly shaped
elements on different volume sizes as well as on different brain’s B0 maps.Acknowledgements
No acknowledgement found.References
1. Juchem
C et al. Dynamic multi-coil shimming of the human brain at 7T. J Magn Reson
2011; 212:280-288.
2. Juchem
C et al. Multi-coil magnetic field modeling. J Magn Reson 2013; 236:95–104.
3. Zivkovic
I et al. B0 shimming at 9.4T using a multicoil approach: coil design with
genetic algorithm, In Proc. 24th ISMRM 2016, Singapore, 1152.
4. Zivkovic
I et al. B0 matrix shim array design – optimization of the position, geometry
and the number of segments of individual coil elements, In Proc. 33rd
ESMRMB 2016, Vienna, Austria, 29(Supplement1)S36.