A novel parallel transmit array design method is proposed that integrates both Maxwell and Bloch equations. The method is demonstrated to be better than traditional coils in dynamic multiband shimming and reduced field-of-view imaging scenarios to achieve better excitation accuracy and lower RF energy deposition, as well as robustness across multiple subjects and excitation schemes.
Figure 2 shows the excitation of the multiband shimming arrays. The coil pruning and combination strategy with joint-5-subject design showed better excitation accuracy than the standard 8-ch array and the quadrature combination method.
Figure 3 summarize the transmit NRMSE and the maximum 10g SAR of the multiband excitation pulses in the brain. The coil combination weights and pruning strategy obtained from the joint-5-subject design showed the best overall performance.
Figure 4 shows (a) the coil element arrangement of the proposed 8-loop-2-channel array, designed by jointly considering all 5 subjects. It achieved roughly the same excitation accuracy as the standard 2-channel array for all 5 subjects, with (b) the excitation patterns and errors of the center slices of each subject shown.
Figure 5 shows the excitation accuracy and maximum 10g SAR of different array design methods. The proposed 8-loop-2-channel array shows large reduction of 10g SAR as much as 50% than the standard 2-coil array. According to IEC regulations (11), the latter would exceed the maximum 10g SAR limit of 10W/kg, while the former would not. Again, the joint-5-subject design showed overall better performance than the subject-1-only design, signifies the importance of using a population of models. Coil combination weights and pruning strategy was also found to be robust on different excitation k-space trajectories.
[1] Mooiweer, R., Sbrizzi, A., Raaijmakers, A. J.E., van den Berg, C. A.T., Luijten, P. R. and Hoogduin, H. (2016), Combining a reduced field of excitation with SENSE-based parallel imaging for maximum imaging efficiency. Magn. Reson. Med.. doi:10.1002/mrm.26346
[2] Lattanzi, R., Sodickson, D. K., Grant, A. K. and Zhu, Y. (2009), Electrodynamic constraints on homogeneity and radiofrequency power deposition in multiple coil excitations. Magn. Reson. Med., 61: 315–334. doi:10.1002/mrm.21782
[3] Cao, Z., Yan, X. and Grissom, W. A. (2016), Array-compressed parallel transmit pulse design. Magn. Reson. Med., 76: 1158–1169. doi:10.1002/mrm.26020
[4] Cao, Z., Yan, X. and Grissom, W. A. Proceedings of ISMRM 2016 in Singapore (Poster 2155).
[5] Christ A, Kainz W, Hahn EG, Honnegger K, Zefferer M, Neufeld E, Rascher W, Janka R, Bautz W, Chen J, Kiefer B, Schmitt P, Hollenbach HP, Shen J, Oberle M, Szczerba D, Kam A, Guag JW, Kustler N. The virtual family development of surface-based anatomical models of two adults and two children for dosimetric simulations. Phys Med Biol 2010; 55: N23–N38.
[6] Eichfelder, G. and Gebhardt, M. (2011), Local specific absorption rate control for parallel transmission by virtual observation points. Magn. Reson. Med., 66: 1468–1476. doi:10.1002/mrm.22927.
[7] Marques, J.P. and Bowtell, R. (2005), Application of a Fourier-based method for rapid calculation of field inhomogeneity due to spatial variation of magnetic susceptibility. Concepts Magn. Reson., 25B: 65–78. doi:10.1002/cmr.b.20034
[8] Malik, S. J., Keihaninejad, S., Hammers, A. and Hajnal, J. V. (2012), Tailored excitation in 3D with spiral nonselective (SPINS) RF pulses. Magn. Reson. Med., 67: 1303–1315. doi:10.1002/mrm.23118
[9] Cloos MA, Boulant N, Luong M, Ferrand G, Giacomini E, Hang MF, Wiggins CJ, Le Bihan D, Amadon A. Parallel-transmission-enabled magnetization-prepared rapid gradient-echo T1-weighted imaging of the human brain at 7 T. Neuroimage 2012;62:2140–2150.
[10] Carluccio G, Erricolo D, Oh S, Collins CM. An approach to rapid calculation of temperature change in tissue using spatial filters to approximate the effects of thermal conduction. IEEE Trans Biomed Eng 2013;60:1735–1741.
[11] European Committee for Electrotechnical Standardization. Particular requirements for the safety of magnetic resonance equipment for medical diagnosis (IEC 60601-2-33:2008). Technical report, European Committee for Electrotechnical Standardization, Brussels, Belgium; 2002.