Esben Søvsø Szocska Hansen1,2, Kasper Wigh Lipsø3, Rasmus Stilling Tougaard1,4, Christoffer Laustsen1, and Jan Henrik Ardenkjær-Larsen3,5
1The MR Research Centre, Aarhus University, 8200, Denmark, 2Danish Diabetes Academy, Odense, Denmark, 3Department of Electrical Engineering, Technical University of Denmark, 4Department of Cardiology, Aarhus University Hospital, Denmark, 5GE Healthcare, Denmark
Synopsis
Dynamic
Contrast-Enhanced MR (DCE-MR) perfusion assessment with gadolinium contrast
agents is currently the most widely used cerebral perfusion MR method. Hyperpolarized
water has recently been shown to succeed 13C probes as angiography probe. In
this study, we demonstrate the feasibility of hyperpolarized water for visualizing
the brain vasculature of a large animal in a clinically relevant setting. In
detail, reference perfusion values were obtained and large to small arteries
could be identified.
Introduction
Dynamic
Contrast-Enhanced MR (DCE-MR) perfusion assessment with gadolinium contrast
agents is currently the most widely used cerebral perfusion MR method
1. The method relies on the
acquisition of a dynamic series of T1-weighted images during administration of
the contrast agent, and as such, the contrast is converted to an apparent
concentration, which is then used to model the cerebral perfusion
2. However, increasing awareness
on the deposition of gadolinium contrast in brain tissue supports the pursuit
of novel perfusion measures such as arterial spin labelling (ASL). ASL is,
however, inherently hampered by the low signal to noise ratio. In contrary, hyperpolarized
MR is the signal source in itself and thus simplifies the quantification
3. A limitation, however, is the need for
13C-molecular contrast agents, which limits spatial and temporal resolutions
significantly. Recently, hyperpolarized water has been demonstrated in rodents
4,5 and in the porcine kidney
6, showing superior spatial resolution compared to
13C angiographies. Here we investigated the potential for utilizing
hyperpolarized water in the porcine brain, in conjunction with a model-free
deconvolution, to map the cerebral perfusion
7.
Methods
A
healthy female Danish domestic pig weighing 31 kg was anaesthetized via continuous intravenous infusion of
both Propofol and Fentanyl. Catheterizations were performed in the left
femoral vein for blood sampling and the right and left femoral arteries for
administration of hyperpolarized water, Percutaneous Intervention (PCI) -procedure
and invasive blood pressure monitoring. 15
mL hyperpolarized water was injected over 5 s, initiated approximately 22 s
after dissolution.
A sample
of 1 mL 30 mM TEMPO (2,2,6,6-Tetramethylpiperidine 1-oxyl, 98%) in H2O/glycerol
1:1 (w/w) was prepared. Dissolution medium consisted of D2O with 1
mM calcium disodium ethylenediaminetetraacetic acid (EDTA) and 9 g/L NaCl (all
from Sigma Aldrich, Denmark). The sample vial was rapidly transferred
from the liquid nitrogen bath to the 5 T magnet in the polarizer (SPINlab, GE
Healthcare, Denmark). The sample was irradiated with app. 50 mW
microwaves at 139.923 GHz for one hour. After dissolution, the radical was
removed by extraction in 25 mL heptane (Sigma Aldrich, Denmark) in a separatory
funnel.
MRI imaging
was performed on a 3 T GE HDx with an 8-channel cardiac array receiver coil (GE
Healthcare, USA). Axial and coronal images were acquired to
cover the brain to allow for planning of the angiography slab for
hyperpolarized water imaging. The sequence was as follows: a 3D T1
weighted sequence with a standard SSFP (TE =
1.1 ms, TR = 2.7 ms, FA = 35°, matrix 256x256, FOV = 340x340 mm2,
in-plane resolution of 1.3 mm and slice thickness of 3 mm). Angiographies were
acquired in a coronal plane covering the brain using a gradient echo sequence: FA = 5°, slice thickness of 40 mm, TR = 3.4 ms, TE = 0.984 ms, matrix 256x256 and FOV = 140x140 mm2. The acquisition time for each frame
was 870 ms and 60 frames were acquired.
Brain perfusion maps
were calculated in OsiriX (Pixmeo, Geneva, Switzerland) using an open source
OsiriX plug-in for T1-DCE-MRI perfusion analysis8. A pixel-by-pixel
model-free, fast deconvolution was applied, assuming a direct signal
enhancement of the hyperpolarized water, a hematocrit of 0.45 and a
regularization of 0.15. Signal-to-Noise-Ratio (SNR) was measure as peak
signal versus baseline noise in the input ROI.Results
Injection
of hyperpolarized water into the brain allowed for visualization of both large
and small vessel as well as for acquisition of perfusion values from brain
tissue. Figure 1 shows the acquired frames and presents the amount of details
from vasculature we have obtained with this setup.
The
calculated perfusion and MTT maps are shown in Figure 2. The results from the perfusion ROI analysis are
(ROI mean value): Anterior,
255 (mL/100mL/min); Posterior, 220 (mL/100mL/min); Circle of Willis, 1895 (mL/100mL/min);
Input, 2645 (mL/100mL/min). The
results from the MTT ROI analysis are (ROI mean value): Anterior, 1.09 (s); Posterior, 1.77 (s); Circle of
Willis, 0.80 (s); Input, 1.00 (s).
Figure
3 shows the dynamic evolution of hyperpolarized water entering the four ROIs
and the difference between peak signal from input to anterior and posterior
part of the brain is clearly observed. SNR was measured to be 273.Conclusion
In this study, we demonstrate the feasibility of
hyperpolarized water for visualizing the brain vasculature of a large animal in
a clinically relevant setting. Furthermore, perfusion values within normal
physiological range are obtained from the dynamic acquisition. We believe this
to be a promising case for further sequence and method optimization. We expect
that the polarization of the hyperpolarized water can be increased by a factor
5-10 relative to the current study.Acknowledgements
Funded
by The Danish Diabetes Academy supported by the Novo Nordisk Foundation.References
1. Essig M, Shiroishi MS, Nguyen TB, et al. Perfusion MRI: The Five Most Frequently Asked Technical
Questions. AJR. American journal of
roentgenology. 2013;200(1):24-34.
2. Khalifa F,
Soliman A, El-Baz A, et al. Models and methods for analyzing DCE-MRI: A review. Medical Physics. 2014;41(12):124301.
3. Johansson E,
Mansson S, Wirestam R, et al.
Cerebral perfusion assessment by bolus tracking using hyperpolarized 13C. Magnetic resonance in medicine : official
journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic
Resonance in Medicine. 2004;51(3):464-472.
4. Lingwood MD, Siaw
TA, Sailasuta N, et al.
Hyperpolarized Water as an MR Imaging Contrast Agent: Feasibility of in Vivo
Imaging in a Rat Model. Radiology. 2012;265(2):418-425.
5. Ardenkjaer-Larsen
JH, Laustsen C, Bowen S, Rizi R. Hyperpolarized H2O MR angiography. Magnetic Resonance in Medicine. 2014;71(1):50-56.
6. Wigh Lipso K,
Hansen ES, Tougaard RS, et al. Renal MR angiography
and perfusion in the pig using hyperpolarized water. Magnetic resonance in medicine : official journal of the Society of
Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 2016.
7. Østergaard L,
Sorensen AG, Kwong KK, et al. High resolution
measurement of cerebral blood flow using intravascular tracer bolus passages.
Part II: Experimental comparison and preliminary results. Magnetic Resonance in Medicine. 1996;36(5):726-736.
8. Zollner FG,
Weisser G, Reich M, et al.
UMMPerfusion: an open source software tool towards quantitative MRI perfusion
analysis in clinical routine. J Digit
Imaging. 2013;26(2):344-352.