Patrick Liebig1,2, Robin Martin Heidemann2, Bernhard Hensel1, and David Porter3
1University of Erlangen-Nuremberg, Erlangen, Germany, 2Siemens Healthcare GmbH, 3Fraunhofer MEVIS
Synopsis
Echo-Planar-Imaging1
(EPI) with trapezoidal Readout (RO) and blipped Phase-Encoding (PE) gradients
gives a high level of acoustic noise2. To reduce acoustic noise we
suggest prolonging the duration of the blipped PE gradient in combination with
a sinusoidal or trapezoidal RO gradient and continuous data sampling throughout
the whole RO train. This results in a variable density sampling along PE (less
dense at the edges, denser in the centre of k-space) with a non-Cartesian
trajectory, where we use ESPIRiT3 to reconstruct the data. The
efficiency is improved due to continuous data sampling, resulting also in a
reduced echo time.
Purpose:
In
standard Echo-Planar Imaging (EPI) a trapezoidal Readout (RO) gradient is
combined with a blipped Phase-Encoding (PE) gradient. During the blips no data
are acquired, resulting in a lower efficiency. Furthermore, this sampling
scheme comes with a high level of acoustic noise2. The current gold
standard in reducing acoustic noise is using a sinusoidal RO gradient with a
constant PE gradient (constant PE-EPI)4. This has beneficial effects
on fMRI studies5. Data can be acquired throughout the whole RO train
except for the edges of k-space (about 5 %) resulting in a higher efficiency
compared to blipped PE gradient. This EPI acquisition results in a
non-Cartesian trajectory (compare Fig.1). It would be necessary to use
non-Cartesian parallel imaging methods for accelerated acquisitions. It was
shown that a variable density trajectory is beneficial for non-Cartesian
parallel imaging6. We optimize the trajectory to be less dense
sampled in PE direction at the edges of k-space, while it is oversampled in PE
direction in the centre of k-space. This method is named Variable-Blipped EPI
(VB-EPI).Method:
To achieve
the above proposed trajectory, we suggest to prolong the duration of the
blipped PE gradients. This results in a lower amplitude and slew rate of the PE
gradient blips. As a consequence, the acoustic noise is reduced (see Fig. 2).
In our method, data can be sampled during the complete RO train. We compare our
method for different blip durations against constant PE-EPI with sinusoidal RO
gradient.
Simulations
have been performed in IDEA (Siemens sequence development environment) to
investigate the efficiency gain (see Fig. 3). The gain in efficiency is
determined by comparing the smallest possible FOV for VB-EPI and blipped EPI.
This is given here in % difference of the smallest possible FOV relative to
blipped EPI sampling for different echo-spacings and with different
acceleration factors. For continuous data acquisition throughout the whole
RO-train, the trajectory in k-space will not be Cartesian anymore (compare Fig.
1). Consequently, the data is reconstructed by using non-Cartesian parallel
imaging methods like ESPIRiT3. We hereby used the BART
Reconstruction toolbox7. To assess the image reconstruction quality,
the Root-Mean-Square-Error (RMSE) of our VB-EPI sequence and in addition a constant
PE-EPI against a blipped EPI was determined in Matlab (Mathworks, Nattick,
USA). This was done for different blip durations and acceleration factors
(compare Fig. 4). All measurements were performed with a prototype VB-EPI
sequence on a standard 3T Skyra XQ system with a 20 channel Head-Neck Coil
(Siemens Healthcare, Erlangen, Germany) whereby only the 16 head elements were
used.Results:
Obviously,
the acoustic noise (see Fig. 2) decreases with increased blip duration. For
blip durations of about ½ echo-spacing, the acoustic noise level is in the
range of constant PE-EPI. On the other hand, the reconstruction quality
decreases with increasing blip duration and acceleration factor (compare Fig.
4). However, the RMSE error is much lower for VB-EPI compared to constant PE
EPI. It can be seen in Fig. 3 that the gain in efficiency increases for lower
echo-pacing, higher acceleration factor and for RO segmentation8. In
Fig. 5 sinusoidal blipped-EPI is compared to our VB-EPI sequence for a blip
duration of ½ echo-spacing.Discussion:
For
the VB-EPI sequence there is a compromise to be made between extending the duration
of the blip to reduce acoustic noise and the level of image quality, which
deteriorates with increasing blip duration. As seen from the results in Figs. 2
and 4, a good compromise can be reached by using a blip duration equal to half
the echo spacing. This provides a sampling scheme with the low acoustic noise
properties of constant PE-EPI and the high image quality of blipped-PE EPI.
Indeed, as shown in Fig. 5, this sampling scheme results in an image quality
that is visually comparable to that of blipped-PE EPI. The higher sampling
efficiency benefits at short echo spacing make the VB-EPI sampling scheme
particularly well suited to RO-segmented-EPI, which uses a substantially
shorter echo spacing than that of single-shot EPI. Conclusion:
We
have shown that the new method VB-EPI successfully reduces acoustic noise in
EPI with minimal loss in image quality. This is of high importance for fMRI
studies. The combination of non-Cartesian parallel imaging strategies, such as
ESPIRiT, and variable density acquisitions is beneficial for higher
acceleration factors. Especially in this regime, the gain in efficiency and the
reduction of acoustic noise is most pronounced. Currently we are implementing
this approach in 3D-EPI/SMS-imaging, conducting fMRI studies and acquiring
RO-segmented diffusion data. Acknowledgements
No acknowledgement found.References
1. Mansfield P. Multi-planar image formation using NMR
spin echoes. J. Phys. C: Solid State Phys. 1977;10(3):L55
2. Ravicz ME, Melcher JR, Kiang NYS, Acoustic noise
during functional magnetic resonance imaging. The Journal of the Acoustical
Society of America, 2000;108(4):1683-1696.
3. Uecker M, Lai P, Murphy MJ, Virtue P, Elad M, Pauly
JM, Vasanawala SS, Lustig M. ESPIRiT--an eigenvalue approach to autocalibrating
parallel MRI: where SENSE meets GRAPPA. Magn Reson Med. 2014;71(3):990-1001
4. Schmitter
S, Diesch E, Amann M, Kroll A, Moayer M, Schad LR. Silent
echo-planar imaging for auditory FMRI. Magma, 2008;21(5):317-25.
5. Peelle
JE, Eason RJ, Schmitter S, Schwarzbauer C, Davis MH. Evaluating an
acoustically quiet EPI sequence for use in fMRI studies of speech and auditory
processing. Neuroimage.
2010;52(4):1419-9.
6. Heidemann
RM, Griswold MA, Seiberlich N, Nittka M, Kannengiesser S, Kiefer B, Jakob PM. A
Fast Method for 1D Non-Cartesian Parallel Imaging Using GRAPPA. Magn Reson Med.
2007 Jun; 57(6):1037-1046.
7. Uecker M, Ong F, Tamir JI, Bahri D, Virtue P, Cheng
JY, Zhang T, Lustig M, Berkeley Advanced Reconstruction Toolbox, Annual Meeting
ISMRM, Toronto 2015 In: Proc Intl Soc
Mag Reson Med 23:2486
8. Porter DA, Heidemann RM. High resolution
diffusion-weighted imaging using readout-segmented echo-planar imaging,
parallel imaging and a two-dimensional navigator-based reacquisition. Magn Reson Med. 2009;62(2):468-75.