Ayana Setoi1 and Katsumi Kose1
1University of Tsukuba, Tsukuba, Japan
Synopsis
Ultrashort echo time (UTE) imaging with cone trajectories was installed
to a 1.5T compact MRI system using an unshielded insertable gradient coil. K-trajectories
of the 3D cone trajectory acquisition were measured using a small capillary
phantom and used for image reconstruction. A LEGO block sample with T2* of
about 0.6 ms was successfully imaged with echo time of 0.05 ms to 0.6 ms. This result
demonstrated that UTE imaging sequences with cone trajectories were
successfully installed to our system.
Introduction
Ultrashort echo time (UTE) imaging is widely used for MRI of T2
or T2* short materials such as cortical bones, tendons, and lung
tissues. Because UTE imaging requires short and intense RF pulses, and fast and
intense field gradients, its implementation to existing whole body scanners has
several difficulties. In this study, we solved these problems using an
insertable gradient coil and cone trajectories with k-space trajectory
correction.Materials and methods
We used a home-built compact MRI system consisting of a 1.5 T and 280 mm
diameter horizontal bore superconducting magnet (JMBT-1.5/280/SS, JASTEC, Kobe,
Japan), a second-order room temperature shim coil set (diameter = 135 mm), an
unshielded gradient coil set designed with the target field method [1] (diameter
= 105 mm, efficiencies: 4.48, 4.73, and 4.95 mT/m/A for Gx, Gy, and Gz), an
eight-rung birdcage coil (diameter = 64 mm, length = 64 mm), and a digital MRI
console (MRTechnology, Tsukuba, Japan). Figure 1 shows samples used for UTE imaging
and k-space trajectory correction. Cone trajectories used for 3D image
acquisition were developed by combining spiral trajectories designed by the
Glover’s approach [2] and a radial sampling along the other direction (e.g. z
direction). The polar angle from z direction (0~π) was equally divided into
128. The number of shots for one cone was 8, 16, 32, and 64. K-space
trajectories were measured using a water phantom in a glass capillary (inner
diameter = 1.0 mm, water length = 15mm). The k-trajectories were measured in
the xy-plane and in the xz plane by placing the capillary phantom along z and y
directions, respectively. The trajectories were calculated from the phase of
the NMR signal of the capillary phantom. The SNR of the NMR signal was improved
by 1,024 times signal accumulation. UTE images of the samples were acquired
with cone trajectories some delay times after a 60° RF pulse (40 µs width) or a 90° RF pulse (120 µs width). The repetition times
of the sequences were 400 ms.Results
Figures
2(a) and (b) show temporal changes of (a) kx and ky and
(b) kx and kz, and k-trajectories in (a) xy and (b) xz
plane measured using the capillary phantom. When the time t is smaller than
about 0.5 ms, considerable distortion of the k-trajectories was observed. Figure
3 shows 2D cross-sections selected from 3D image datasets of the cylindrical
phantom acquired with the cone-trajectories. The field of view was 64 mm cube
and image matrix was 128×128×128. Figures 3(a) and (b) were acquired when the
cylindrical phantom was placed parallel to the xy plane and parallel to the xz
plane and reconstructed using theoretical and measured trajectories.
Figures
4(a) and (b) show 2D cross-sections selected from 3D image datasets of the LEGO
block sample acquired with (a) 256-shot radial trajectories for one polar angle
and (b) 64-shot cone trajectories for one polar angle. The FOV and image matrix
were identical to those of Fig.3. The spatial resolution is comparable between
(a) and (b) but the acquisition time of (b) was about 1/4 of that of (a) except
the overhead time of the sequence updates.
Figure
5 shows 2D cross-sections selected from 3D image datasets of the LEGO block
sample acquired with cone trajectories when (a) a 120 µs width square 90° RF excitation pulse and (b) a
40 µs
width square 60° RF
excitation pulse were applied. The image intensity of the LEGO block sample monotonously
decayed with the echo time, which reflected T2* (~0.6ms) of the
sample.Discussion
As shown in Fig.2, considerable distortion was observed for the
k-trajectories when the time t was smaller than 0.5 ms. We think that this
distortion was caused by contamination from NMR signals of solid-like materials
such as epoxy resin to seal the water in the capillary and some materials used
to fix the capillary phantom. This distortion was not harmful for k-trajectory
correction because the theoretical trajectories can be used for image
reconstruction after overall fitting of the trajectories. Figure 3 demonstrates
that k-trajectory correction worked well. Figure 4 shows the 3D UTE
imaging using cone trajectories (64 shots for one cone angle) was three to four
times faster than that using radial trajectories with similar spatial
resolution. Because the T2* of the LEGO block protons were about
0.6 ms, Figure 5 shows the 3D UTE imaging using cone trajectories were
successfully installed to our system. In conclusion, 3D UTE imaging using an
insertable gradient coil can be a useful approach to 3D MRI of solid-like materials
in whole body MRI systems.Acknowledgements
No acknowledgement found.References
[1]
Turner R. A target field approach to optimal coil design. J. Phys. 1986; D19:
L147-151
[2]
Glover G H. Simple analytic spiral k-space algorithm. Magn Reson Med 1999; 42:
412.