Anirban Sengupta1, Rakesh Kumar Gupta2, Sumeet Agarwal3, and Anup Singh4
1CBME, IIT Delhi, New Delhi, India, 2Radiology Department, FORTIS hospital,Gurgaon, India, 3Electrical Engineering, IIT Delhi, India, 4Centre for Biomedical Engineering, IIT Delhi; AIIMS Delhi, New Delhi, India
Synopsis
The purpose of this study is to propose a fast T1 estimation
method with improved accuracy over existing approaches in a Multiple Flip Angle
setting. A supervised machine learning based approach has been proposed that
can be used to predict additional Flip Angle data using limited available Flip
Angle data, thereby producing more accurate T1 estimation in reduced scan time.
Both experimental as well as simulation results are shown to illustrate the
efficacy of this approach. The accuracy of T1 estimation depends on the choice
of Flip Angle data to be predicted.
Purpose
T1 estimation based upon multiple flip angles(MFA) is widely used clinically due
to shorter data acquisition time1. Accuracy of T1 map depends upon
appropriate numbers of flip angle(FA). Currently, there is no accepted FA set,
that is used to minimize the uncertainty over T1 estimation2 .
However, MFA is preferred when imaging over large T1 range3 as at
low or high T1 values dual FA reduces the efficiency1 due to high
noise bias. Moreover, errors in T1 estimation propagates to kinetic parameter
estimation4. The objective of this study was to propose a Neural
Network(NN) based supervised machine learning approach that predict additional
FA data using limited available FAs data, thereby producing improved T1
estimation in reduced scan time.
Method
This study was done for
experimental as well as simulation data on FA set[30,60,100,150],
used routinely for clinical purpose in a local hospital. MRI study was
conducted on brain of 15 tumor patients at 3T(Ingenia,PhilipsHealthcare)
scanner with TR=6ms. T1 estimation was done using a non-linear
least-square fitting of the pixel-wise image intensities at MFAs to the
theoretical equation for SPGR signal intensity4. B1-inhomogeneity
correction was done using saturated double-angle method5.
Different
Case studies were done. In Case-1, all four FA[30,60,100,150]
were used to compute T1 map. In Case-2, a NN was used to predict data for FA-30 from FA[60,100,150]
and T1 map was computed using acquired FA[60,100,150] data and predicted FA-30 . The NN parameters were
optimized to 1 hidden layer with 3 nodes by iteratively training. In Case-3, T1
map was computed using three FA[60,100,150] data. In Case-4, we used a NN, with 1 hidden
layer having 4 nodes, to predict FA data at [30,100]
from data at [60,150].
In this case, T1 map was computed using acquired FA[60,150]
data and predicted FA[30,100]
data. In Case-5, T1 was computed using only two FA[60,150]
data. Relative percentage error(RPE) in T1 computation was calculated at each
voxel for all Cases with respect to Case-1. NN was trained and optimized using data of 10
patients and the efficacy of the model was tested on remaining 5 patients. In
simulated data T1 values[800,1200,1800,2400 ms] corresponding to 4 tissues in a
tumor brain(GM,WM,Tumor,CSF) was used. Noise upto 30 % was incorporated in the
data. A NN, was trained on 15000 data points for each T1 value. The efficiency
of the NN model was tested at different noise levels, to mimic real data.
Results
Fig-1 results show that
T1 map obtained in Case-2 is similar to that of Case-1 whereas T1 map
obtained in Case-3 is a significantly underestimated. Table-1 shows that Case 2
has a nominal RPE(~5-12 %) compared to a high RPE in Case 3 (~66-70 %) for 5 patients. Fig-2(A) shows the simulated
signal intensity curves corresponding to 4 different T1 values. Fig-2(B) shows
Case-3, has much higher RPE compared to Case-2, at different noise levels. Fig-3(A)
shows that in experimental data the RPE in Case-5 is higher than Case-4, for
all 5 patients except 1. Fig-3(B) shows that in simulated data, the difference
in RPE of estimated T1 map is comparatively much less in Case-4 compared to Case-5.
Simulation result, also shows, that RPE is proportional to noise in data. Discussion
The results show that
the proposed method for T1 estimation produces significantly better
accuracy(Case-2 and Case-4) compared to their counterparts(Case-3 and Case-5)
in both experimental and simulation data. In real life scenario, where data for
one of the FA data can get corrupted due to motion or other artifacts, proposed
approach can recover this data. Accuracy can depend on the importance of FA
that is being predicted. It was found that T1 maps estimated using FA[30,60,100]
produces similar accuracy as that obtained using, known FA[30,60,100]
data and predicted FA-150data, thus highlighting the relatively less
importance of FA-150 in fitting data to the standard equation. A
careful trade-off has to be done before choosing which FA data should be
predicted. The accuracy of NN model can be further increased by training it
with a larger number of patients so that data with noise can be modelled. Since
the NN tries to predict data governed by standard equation, it remains to be
seen if modelling NN with a known relation between input and output, produces
better accuracy or not.Conclusion
The results shows the advantage of using a machine
learning approach to predict additional FA data which can be used to estimate
T1 map. This method reduces scan time as well as increases the accuracy of T1
estimation compared to lesser FA dataset.Acknowledgements
The authors acknowledge internal grant from
IIT-Delhi and technical support of Philips India Limited in MRI data
acquisition. We thank Dr. Indrajit Saha for providing technical support for MRI
data acquisition; Dr Pradeep Gupta in data acquisition.References
1. Cheng HL, Wright GA. Rapid
high-resolution T1 mapping by variable flip angles:
Accurate and precise measurements in the presence of radiofrequency field
inhomogeneity. Magnetic resonance in medicine. 2006 Mar 1; 55(3):566-74.
2. Liberman G, Louzoun Y, Ben
Bashat D. T1 mapping using variable flip angle SPGR data with flip angle
correction. Journal of Magnetic Resonance Imaging. 2014 Jul 1; 40(1):171
3. Deoni SC, Peters TM, Rutt
BK. Determination of optimal angles for variable nutation proton magnetic
spin-lattice, T1, and spin-spin,
T2, relaxation times measurement. Magnetic resonance in medicine. 2004 Jan
1; 51(1):194-199.
4. Yuan J, Chow SK, Yeung DK, Ahuja
AT, King AD. Quantitative evaluation of dual-flip-angle T1 mapping on DCE-MRI
kinetic parameter estimation in head and neck. Quantitative imaging in medicine
and surgery. 2012 Dec; 2(4):24
5.
Cunningham CH, Pauly JM,
Nayak KS. Saturated double-angle method for rapid B1+
mapping. Magnetic Resonance in Medicine. 2006 Jun 1; 55(6):1326-33.