3851

Improved CAIPIRINHA for Simultaneous Multislice Imaging Using Total Variation Regularization
Song Gao1 and Yajun Ma1

1Medical Physics Department, Peking University, Beijing, People's Republic of China

Synopsis

The CAIPIRINHA (Controlled Aliasing in Parallel Imaging Results in Higher Acceleration), proposed for simultaneous multislice imaging, combines the multiband excitation and phase cycling techniques to reduce scan time and improve the subsequent imaging reconstruction.The total variation (TV) regularization uses an edge-preserving prior which imposes a relationship between neighboring pixels for image reconstruction. In this work, the TV regularization method is applied to further improve the CAIPIRINHA. In vivo data results demonstrate that the TV regularization is very powerful in suppressing noise, maintaining fine imaging details and reducing aliasing artifacts.

Introduction

Simultaneous multislice imaging (SMS) became popular since the method of the Controlled Aliasing in Parallel Imaging Results in Higher Acceleration (CAIPIRINHA)was proposed1. Recently, a conjugate gradient (CG) iteration reconstruction framework has been developed for the CAIPIRINHA technique2. However, the images produced by the CG method without regularization still contain notable aliasing artifacts and noise contamination. Total variation (TV) regularization has been widely used in image-domain reconstruction, such as SENSE, because it is an image edge-preserving regularization method by imposing a relationship between neighboring pixels in images3,4. In this work, TV regularization method is applied to further improve the CAIPIRINHA.

Methods

For the purpose of identifying the effects of the TV regularization to the CAIPIRNHA, in vivo data were acquired from human brain on a Philips 3T MRI scanner using a normal spin-echo sequence with imaging parameters as follows: TR/TE = 490 ms/11.6 ms, matrix size = 256 196, field of view = 230 198 mm2, number of slices = 18, slice thickness = 4 mm, slice gap = 1mm. The imaging data was acquired with an eight-channel head coil. A simultaneously excited data was derived from the summation of four slices (these slices were chosen from the acquired 18 slices: the 1st, 3rd, 8th and 10th slices) with different phase cycles: 0, π/2, π, 3π/2. Fig. 1 shows the two of the channel images. Coil sensitivity profiles of the excited four slices were obtained by the self-calibration method using 48 phase-encoding lines at the center of their k-space5. Then, a fixed-point iteration algorithm was used to solve the non-linear minimization problem of equation:

$${{f}_{TV}}=\underset{f}{\mathop{\arg }}\,\ \min \left\{ \ \left\| Ef-d \right\|_{2}^{2}+\lambda \sum{\sqrt{{{\left| {{\nabla }_{x}}f \right|}^{2}}+{{\left| {{\nabla }_{y}}f \right|}^{2}}}} \right\} [1]$$

where f is the unfolded images , d is the acquired k-space data from all the channels, and denote the gradient along x and y direction in the 2D image, respectively, and | | denotes the complex modulus. The λ is the TV regularization parameter. E is the spatial encoding matrix combined with gradient and coil sensitivity encodings and the matrix element is as follows:$${{E}_{r,c}}={{C}_{ns,j}}(x,y-\Delta y)exp(in\Delta {{k}_{y}}\Delta y)\exp (-im\Delta {{k}_{x}}x)\exp (-in\Delta {{k}_{y}}y) [2]$$

In addition, CG iteration reconstruction was also implemented for comparison. For quantitative comparison, the normalized mean squared error (NMSE) is defined as the normalized square of the difference between the reconstructed image (Irec) and the reference standard image (ISOS: it was obtained by sum-of-square (SOS) method for each single acquired slice). And the NMSE is expressed as follows:$$NMSE=(\sum\limits_{(x,y)}{{{\left| \left| {{I}_{rec}}(x,y) \right|-\left| {{I}_{sos}}(x,y) \right| \right|}^{2}})}/(\sum\limits_{(x,y)}{{{\left| {{I}_{sos}}(x,y) \right|}^{2}}})) [3]$$

Results

The first three rows in Fig. 2 shows the obtained four slices produced by SOS, CG and TV reconstruction methods. The same regions in all the images are amplified for better comparison. Noise can be seen in the images by CG reconstruction and the images using TV based reconstruction almost have the same image quality as the standard SOS images.

The error maps calculated by the two equations: $$$Er{{r}_{CG}}=\left| {{I}_{CG}}-{{I}_{SOS}} \right|$$$ and $$$Er{{r}_{TV}}=\left| {{I}_{TV}}-{{I}_{SOS}} \right|$$$ are shown in the fourth and fifth rows in Fig. 2. They were normalized and displayed with a same color scale. Both the intensities of noise and artifacts are smaller and lower in the TV images than the CG images. The NMSE values are shown in Fig. 3 which presented the total errors of the TV and CG images. It is can be seen that the NMSE values of TV images are always lower than the CG images.

Discussion

In this work, we find that the image quality of the TV regularized reconstruction for CAIPIRNHA is much better than the CG reconstruction. However, solving the Eq. [1] is a non-linear minimization problem and it needs much more computation time than the CG method. To realize the real time calculation in our later work, the advanced tools can be used such as CUDA programing based on the GPU. In addition, how to select a proper TV regularization parameter λ still needs more investigation. Here, we employed the automatic search method to find the best TV regularization parameter, which is accurate but time consuming. To further accelerate the MR data acquisition, the CAIPIRNHA technique can be combined with the typical parallel imaging methods. The proposed TV regularization can also be applied in this case.

Conclusion

In this study, the TV regularization method is applied for CAIPIRNHA reconstruction. In vivo data results demonstrate that the TV regularization is powerful in suppressing noise, maintaining fine imaging details and reducing aliasing artifacts.

Acknowledgements

This work was supported by the National Natural Science Foundation of China ( Grant No.61671026), Beijing Natural Science Foundation (Grant No. 7162112)

References

1. D. J. Larkman, J. V. Hajnal, A. H. Herlihy,et al. Use of multicoil arrays for separation of signal from multiple slices simultaneously excited.J Magn Reson Imaging. 2001;13(2): 313-317.

2. S. R. Yutzy, N. Seiberlich, J. L. Duerk, et al. Improvements in multislice parallel imaging using radial CAIPIRINHA. Magn Reson Med. 2011;65(4):1630-1637.

3. K. P. Pruessmann, M. Weiger, B. Scheidegger, et al. SENSE: sensitivity encoding for fast MRI. Magn Reson Med.1999;42(2): 952–962.

4. K. T. Block, M. Uecker, J. Frahm. Undersampled radial MRI using multiple coils: iterative image reconstruction using a total variation constraint. Magn Reson Med. 2007;57(6):1086–1098.

5. C. A. McKenzie, E. N. Yeh, Ohliger MA,et al. Self calibrating parallel imaging with automatic coil sensitivity extraction. Magn Reson Med.2002;47(3):529–538.

Figures

Figure 1. Two channel images of the simultaneously excited image-domain data with the acceleration factor R = 4.

Figure 2. The imaging results produced by SOS, CG and TV reconstruction methods are shown in the first, second and third rows, respectively. The error maps of CG and TV reconstructed images are shown in the fourth and fifth rows, and their values are normalized and displayed with a same color scale.

Figure 3. The NMSE lines of CG and TV methods.

Proc. Intl. Soc. Mag. Reson. Med. 25 (2017)
3851