Mihir Pendse1 and Brian K Rutt1
1Stanford University, Stanford, CA, United States
Synopsis
We describe a procedure for obtaining minimum duration
parallel transmit pulses for simultaneous multislice imaging. The strategy
involves a combination of the IMPULSE parallel transmit pulse design algorithm
for mitigating local SAR hotspots and enforcing flip angle homogeneity; an
optimal control algorithm for intelligent design of slice selective subpulse
shape with minimum power; and time-optimal VERSE for reducing peak power with
minimum increase in pulse duration. Results indicate a reduction in pulse
duration by a factor of 4.9 compared to a pTx-SMS pulse designed with
conventional techniques.
Introduction
Simultaneous multislice (SMS) imaging is an effective
technique for reducing acquisition time or increasing slice coverage. The
design of SMS RF pulses to excite multiple slices simultaneously is challenging
because these pulses have high peak power [1]. At high fields, RF pulse design
is further complicated by B1+ inhomogeneity and increased local SAR concerns
that can be addressed through SAR-aware parallel transmission (pTx), typically
resulting in multi-spoke pulses [2,3]. In addition to having high peak power, the
duration of pTx pulses with multiple spokes can be long, making them very
sensitive to B0 inhomogeneity. Here we describe a technique for designing
minimum-duration pTx-SMS pulses that satisfy peak power, excitation
accuracy, and local SAR constraints. Our results demonstrate significant
reduction in pulse duration compared to existing pTx-SMS pulse designs.Theory
We modified the previously described IMPULSE-SMS design
algorithm [4] to find minimum-duration RF and gradient waveforms following the same
principles as minimum-SAR pTx pulse design. The design procedure is as follows.
First, IMPULSE pTx optimization is performed to find spokes locations and
channel weightings for each slice in order to minimize SAR while enforcing a specified
flip angle inhomogeneity tolerance over all excited slices. Next, the SMS pulse
to excite multiple slices is designed using optimal control theory [5] to minimize
a cost function consisting of terms corresponding to both excitation accuracy
and pulse power. The regularization term in the cost function can be chosen such
that the resulting excitation accuracy is an accordance with the specified flip
angle inhomogeneity tolerance. Finally, if the peak power limit on any channel
is exceeded, time-optimal variable rate selective excitation (VERSE) [6] is
applied to enforce the peak power constraint with minimum pulse duration by
reshaping the RF and gradient waveforms without altering the excitation profile.Methods
Small-tip pTx-SMS RF and gradient waveforms were designed to
simultaneously excite 6 coronal brain slices using a 16 channel transmit coil
and pulse constraints as described in [7]. A minimum duration pulse was
designed while satisfying all of the following constraints: peak local SAR <
10W/kg, global SAR < 3.2W/kg, and in-slice RMS flip angle inhomogeneity <
6.7%. We compared these results to those obtained with a conventional
SAR-unaware pTx-SMS design algorithm using naïve stretching of the pulse to
satisfy peak power and SAR constraints.Results
We found that it was possible to achieve an SMS factor of 6
with pulse duration of less than 5ms while meeting all constraints. The
conventional pTx-SMS pulse design resulted in a duration of 24.4 ms (Figure 1 ).
The proposed method
succeeded in reducing this duration to 4.96 ms (Figure 2) which is a factor of
4.9 improvement . As seen in Figures 3 and 4, this
results in flip angle maps that have the same degree of homogeneity while also
satisfying all local SAR, global SAR, peak RF power, and maximum gradient amplitude/slew constraints.
Discussion
The proposed approach allows for significant reduction of
the duration of pTx-SMS pulses by using a SAR-aware pTx optimization to better
satisfy safety constraints and by using time-optimal VERSE to enforce peak
power constraints in a more efficient manner than naïvely stretching the pulse.
The shorter pulses will be more robust to B0 inhomogeneity and thus can be expected to be useable in invivo imaging without resulting in major artifacts. One limitation of
this method is that VERSE is sensitive to accuracy in the gradient field and
errors in the timing between the RF and gradient waveforms but the dramatic
improvement seen here is motivation for investing in more advanced hardware
with better management of these concerns.
Acknowledgements
Research support from the NIH (P41 EB015891, 1 S10 RR026351-01A1), GE Healthcare, and Zurich MedTech AG (Sim4Science program)
Dr. William Grissom and organizers of 2015 ISMRM RF Pulse Design Challenge for motivation behind the methods developed in this work.
References
[1] Eichner
et al. Magn Reson Med. 2014
Oct;72(4):949-58
[2] M.
Pendse and B. Rutt. IMPULSE: A generalized and scalable algorithm for joint design
of minimum SAR parallel transmit RF pulses. Proceedings ISMRM 23:5008, 2015
[3] Hoyos-Idrobo,
Andres, et al. "On variant strategies to solve the magnitude least squares
optimization problem in parallel transmission pulse design and under strict SAR
and power constraints." IEEE transactions on medical imaging 33.3
(2014): 739-748.
[4] Pendse
MR and Rutt BK. IMPULSE-SMS: Local SAR and peak power optimized pTx pulse
design for simultaneous multislice imaging at high fields. Proceedings ISMRM
2016
[5] Aigner
CS, Clason C, Rund A, and Stollberger R. RF pulse design for low SAR
simultaneous multislice (SMS) excitation using optimal control. Proceedings
ISMRM 2014
[6] Lee D,
Lustig M, Grissom WA, Pauly JM. Time-optimal design for multidimensional and
parallel transmit variable-rate selective excitation. Magn Reson Med
2009;61:1471–1479.
[7]
Grissom, W. A., Setsompop, K., Hurley, S. A., Tsao, J., Velikina, J. V. and
Samsonov, A. A. (2016), Advancing RF pulse design using an open-competition
format: Report from the 2015 ISMRM challenge. Magn. Reson. Med.