Joshua S. Greer1,2, Xinzeng Wang2, Keith Hulsey2, Robert E. Lenkinski2,3, and Ananth J. Madhuranthakam2,3
1Bioengineering, UT Dallas, Richardson, TX, United States, 2Radiology, UT Southwestern Medical Center, Dallas, TX, United States, 3Advanced Imaging Research Center, UT Southwestern Medical Center, Dallas, TX, United States
Synopsis
Arterial spin labeling (ASL) is a rapidly growing area of
interest, primarily because of its ability to provide non-contrast quantitative
perfusion maps. For the technique to be adopted for clinical use, these
quantitative measurements need to be accurate and robust, which will require a
quality controlled perfusion phantom to ensure consistency for different magnet
strengths and manufacturers. In this study, we demonstrate a 3D printed
perfusion flow phantom that can be easily replicated, and used to test the
precision and repeatability of ASL perfusion measurements.
Introduction
Arterial spin labeling (ASL) is a non-contrast perfusion
imaging technique that has clinical value because it can easily provide
quantitative perfusion maps. This could be beneficial in measuring therapeutic
response in various diseases, however, for ASL perfusion imaging to be widely
adopted for clinical use, it is important to demonstrate that the technique
provides accurate and reliable quantitative perfusion measurements. In this
study, we developed a 3D printed perfusion phantom that can be used as a
quality control (QC) phantom to evaluate the precision and repeatability of arterial
spin labeled perfusion measurements.Methods
The
perfusion phantom was designed to split inflowing water into multiple branches,
mimicking the branching of arterial vessels (Figure 1). The branches terminate
at small evenly-spaced holes leading into the large chamber that can be fitted
with tissue-mimicking material, such as sponge. The small evenly-spaced holes ensure
that the entire sponge is completely perfused. The water then leaves the phantom
through a mirror image of the input branches. The phantom was submerged in a
water bath to minimize B0 inhomogeneities. A
pump outside the MR scanner room circulated water through the tubing, and was
set to different flow rates to create varied perfusion effects in the sponge.
Perfusion was measured with 2D FAIR and 2D pCASL using a single shot turbo spin
echo (SShTSE) acquisition.
Quantitative perfusion values were
calculated using the standard ASL models for each technique.1,2 pCASL labeling was applied axially
over the inflow tube for 6 seconds, followed by a 500ms delay before imaging.
The inversion time for FAIR was 3 seconds. The pump flow rate was varied from 0
to 600 mL/min to vary perfusion in the sponge (Figure 3). A proton density
weighted image was also acquired through the sponge at each flow rate for
perfusion quantification. ASL measurements were repeated using
the same pump flow rates on two different days to evaluate the repeatability of
ASL measured quantitative perfusion. Finally, the agreement between FAIR
and pCASL quantitative measurements was assessed at each flow rate.Results
Figure 2 shows axial perfusion weighted images acquired with
FAIR and pCASL at increasing pump flow rates. The improved SNR offered by the
continuous labeling technique, pCASL (fig. 2, g-l), over pulsed-ASL technique,
FAIR (fig 2, a-f), is apparent in these perfusion images. Figure 3 shows the
increasing measured perfusion as the inflow rate increases for both ASL techniques
on two different days. The delay in the observed increase in perfusion from
0-200 mL/min can likely be attributed to the later arrival of the labeled
perfusion bolus, and could potentially be corrected by using a shorter
post-labeling delay. Figure 4 demonstrates highly repeatable, quantitative
measurements between two scan dates, and figure 5 shows that FAIR and pCASL
provide the same perfusion measurements, despite the significant differences in
SNR seen in figure 2.Discussion
We demonstrated a simple 3D printed perfusion phantom that can
serve as a quality control to ensure that ASL perfusion measurements are
reliable and precise, and show that various ASL techniques can provide the same
quantitative values. Additionally, this 3D printed phantom can be easily replicated
to be used at different sites to compare the measurements made by scanners from
different manufacturers and at different field strengths. Future experiments
with this phantom will include varying the label duration and post labeling delays
to ensure that measured perfusion agrees with the quantitative perfusion model,
and defining an inflow range in which this phantom design can provide an
accurate linear increase in measured perfusion.Acknowledgements
No acknowledgement found.References
1. Buxton, Richard B., et al. "A general
kinetic model for quantitative perfusion imaging with arterial spin
labeling." Magnetic
resonance in medicine 40.3
(1998): 383-396.
2. Robson,
Philip M., et al. "Strategies for reducing respiratory motion artifacts in
renal perfusion imaging with arterial spin labeling." Magnetic Resonance in Medicine 61.6 (2009): 1374-1387.