Sebastien Bär1, Matthias Weigel2, Dominik von Elverfeldt1, Jürgen Hennig1, and Jochen Leupold1
1Dept. of Radiology, Medical Physics, University Medical Center Freiburg, Freiburg, Germany, 2Radiological Physics, University of Basel Hospital, Basel, Switzerland, Saint Helena
Synopsis
For
applications like dynamic imaging, the relatively fast decaying hyperpolarized
signal must be acquired over a higher number of sequence cycles and the optimal
refocusing flip angle needs to be determined. Optimizing the RARE refocussing flip angles allow to optimize the signal exploitation. Furthermore, optimizing the flip angles of a
RARE-based varying flip-angle TRAPS sequence enables an even longer prolongation of the signal lifetime.
Purpose
The RARE
sequence is well established in MRI routine and was already successfully used
for imaging in hyperpolarization conditions [1]. While the optimal refocusing flip angle is already known (180°) [2,3]
for applications like rapid spectroscopic imaging where the signal is acquired
over a low number of sequence cycles, for applications like dynamic imaging,
the relatively fast decaying hyperpolarized signal must be acquired over a higher
number of sequence cycles and the optimal refocusing flip angle needs to be
determined.
In this
study, we show that depending on the echo train length, the optimal RARE refocusing
flip angle allowing maximal signal exploitation can be optimized. Even more, we
show that a different approach such as the RARE-based varying flip-angle TRAPS
sequence (Transitions
between pseudo steady states) [4] can help to prolongate the signal
lifetime. Subsequent to the optimization of the minimum and maximum flip angles
(αMin, αMax) of the sequence, we show that a different
acquisition strategy may allow an even longer lasting signal.Methods
Both
the RARE and the TRAPS sequences were simulated in MATLAB based on the extended
phase graph (EPG) [5] using the following parameters: 10000 pulses with an echo
spacing of 15 ms corresponding to a total measurement time of 150 s. T1 =
25 s and T2 = 2.5 s are typical values for 13C
hyperpolarized molecules in solution. For both sequences, the 90° excitation
pulse is followed by a first refocusing pulse with flip angle 90°+α/2 that leads
to an additional signal gain [6].
Subsequent refocusing pulses were chosen to have either a constant flip angle α (RARE) or the refocusing flip angles were
varied (TRAPS) according to the flip angle pattern shown in figure 1. The flip
angle ramps from αMin to αMax and the flip angle plateaus
(constant αMin or αMax) were chosen to be 128 pulses long.
Simulations of the RARE and TRAPS sequences were performed while varying the
flip angle α and αMin, αMax with αMin < αMax
from 1° to 180° in 1° steps.Results
The maximal
possible signal for each number of pulses from 0 to 10000 pulses was
investigated and plotted together with the corresponding optimal flip angles in
figure 2. The optimal flip angles maximizing the signal after 150 s (10000
pulses) were found to be α = 9° (RARE) and αMin = 1°, αMax
= 15° (TRAPS). The signal evolution over time of both sequences was then investigated
while using the optimal flip angles (see figure 3). The TRAPS sequence shows
signal plateaus that are higher and lower than the constant decaying RARE-signal.
Images (128 x 128 pixels) using a Shepp-Logan
phantom were simulated using the RARE and TRAPS signals, while Gaussian noise was
added to each of the k-spaces. Three images at later time points (70.8 s, 109.1
s and 132 s) are shown in figure 4. The chosen ROI shows considerably higher
SNR in the TRAPS images than in the RARE images with ~47 % and ~23% more signal
after 70.8 s and 132 s, respectively.
Discussion/Conclusion
According
to our simulations, the RARE sequence is a good candidate when a long-living
signal has to be acquired (e.g. in dynamic imaging), provided the
180°-refocusing flip angle is replaced by a lower, optimized flip angle. Besides,
we have shown for typical relaxation time values of 13C
hyperpolarized molecules in solution (T1 = 25 s and T2 =
2.5 s), that a TRAPS sequence, differing from the RARE sequence only in the
flip angle scheme, allows an even longer-lasting signal exploitation. While the
here simulated TRAPS images are using only the high signal plateau, optimized
acquisition strategies of the k-space may allow increasing the image resolution
without losses of SNR. E.g. acquiring the outer and central k-space lines over
the low- and high- signal plateaus, respectively, leads to filtering k-space
with the TRAPS signal envelope function, with potential impairment on image
quality.Acknowledgements
No acknowledgement found.References
[1] Golman et al., MRM 46.1: 1-5(2001), [2]
Svensson et al., MRM 50:256-262(2003), [3] Leupold et al., Proc. ISMRM 2009, p.2426,
[4] Hennig et al., MRM 49.3: 527-535(2003), [5] Weigel et al., JMRI 41.2:
266-295(2015), [6] Hennig et al., MRM 44.6: 983–985 (2000)