Jiazheng Wang1, Alan Wright1, Richard Hesketh1, De-en Hu1, and Kevin M Brindle1,2
1Cancer Research UK Cambridge Institute, University of Cambridge, Cambridge, United Kingdom, 2Department of Biochemistry, University of Cambridge, Cambridge, United Kingdom
Synopsis
We
have developed a workflow for hyperpolarized 13C EPI phase
correction that requires no reference scan. The workflow provides ghost-free images
on phantoms with large or tight fields of view and where there are multiple
signal sources. Dynamic images acquired from hyperpolarized [1-13C]pyruvate
and [1-13C]lactate in a tumor showed comparable image quality to those
corrected using a separate 13C reference scan.
Purpose
Development
of a referenceless workflow for hyperpolarized 13C EPI. Introduction
The development of DNP has
enabled MRI studies of tissue metabolism, using hyperpolarized 13C-labelled
substrates, at relatively high spatial and temporal resolutions 1.
The transient nature of the hyperpolarization requires fast imaging. While EPI is an appealing technique because it
can acquire a 2D image after a single excitation, it suffers from Nyquist
ghosting. To correct for this a 13C reference scan can be acquired,
but this takes up one frame of the dynamic data set. Alternatively a reference
scan can be acquired using the 1H signal 2, but this can be
suboptimal for 13C phase correction. Here we describe a workflow for
hyperpolarized 13C EPI that renders reference scans unnecessary.Methods
The workflow is illustrated
in Figure 1, and is demonstrated using imaging of hyperpolarized [1-13C]pyruvate
and [1-13C]lactate in vivo. An exhaustive search of the phase error
correction coefficients is performed on the 13C EPI image to
minimize signal in the ghost-containing background, which is readily obtainable
from the proton image of tissue anatomy. As shown in Figure 2, a tumor ROI is
selected from the proton image and mapped onto the [1-13C]lactate echo
planar image, while regions outside the ROI in the phase encoding direction are treated as
ghosting-containing background. The search is performed on the [1-13C]lactate
image with the highest SNR and the correction coefficients obtained applied to
all other images and for both metabolites.Experiments
Experiments were performed
on a 7T animal scanner (Agilent). Gradient-echo
echo planar images were acquired from phantoms (diameters 7 mm and 18 mm,
filled with 8.5M and 4M [1-13C]lactate respectively at thermal
equilibrium) with a 4 cm FOV in order to examine performance of the method
under conditions where there was a regular FOV, a tight FOV, and in the
situation where there were multifocal signals. Image results from the workflow
with 1D and 2D searches for the phase correction coefficients were compared with
those corrected using a separate 13C reference scan and using 13C
reference echoes 3. The workflow was also compared with the 1H
reference scan method in the case of multifocal-signals, with a well-shimmed magnetic
field and with a background gradient of 6mT/m2 in the XY direction. Imaging
in vivo was conducted using the same parameters on female C57BL/6J EL4 lymphoma
tumor-bearing mice following injection of hyperpolarized [1-13C]pyruvate,
with alternating SpSp pulses on the [1-13C]pyruvate and [1-13C]lactate
resonances and with a time resolution of 2s per metabolite.Results
The proposed workflow
resulted in phantom images of similar quality to those obtained using a
separate 13C reference scan and using integrated 13C
reference echoes (Figures 3A). The workflow also provided similar ghost-free
images in vivo compared to those corrected with 13C reference scan
(Figure 4). Quantitative measurements of residual ghosting in both phantom and
in vivo images are shown in Figure 5A&5C. The workflow also showed an SNR
benefit over the integrated echo method because of the prolonged TE with the
latter method. The SNR was 1.7%, 3.0%, and 12.6% higher than the integrated
echo method for a regular FOV, tight FOV, and with multifocal-signals respectively
(Figure 5B). As shown in Figure 3B, the workflow gave better results than the 1H
reference scan method (residual ghosting 1.3% versus 4.4%) under well-shimmed
condition and this difference was increased in a de-shimmed condition (residual
ghosting 1.6% versus 13.7%). The in vivo
images obtained using the 1D search method and images corrected using 13C
reference data were almost identical (SSIM≈1) through the whole dynamic imaging
window for both [1-13C]pyruvate and [1-13C]lactate images
(Figures 5D&5E).Discussion and Conclusion
The
referenceless workflow resulted in ghost-free hyperpolarized 13C
echo planar images of phantoms and in vivo. The resultant image quality was
comparable to that obtained using a 13C
reference scan or integrated 13C reference echoes and was
consistently better than that obtained using 1H reference data,
especially under suboptimal conditions. It also showed an SNR benefit over the
integrated 13C reference echo method, and this benefit is expected
to be enhanced in vivo because of the shorter T2*. The 1D search showed higher efficiency than
the 2D search with similar ghost-removal capability and much reduced processing
time. The workflow is compatible with 3D EPI and parallel imaging, and is
readily translatable to clinical applications given the sparse nature of
hyperpolarized 13C metabolite data, which will nearly always provide
a ghost-containing background.Acknowledgements
No acknowledgement found.References
[1] Brindle KM. Imaging metabolism
with hyperpolarized 13C labeled cell substrates. J Am Chem Soc 2015; 137:6418-6427
[2]
Gordon, JW, Vigneron DB, Larson PEZ. Development of a symmetric echo planar
imaging framework for clinical translation of rapid dynamic hyperpolarized 13C
imaging. Magn Reson Med 2016; doi: 10.1002/mrm.26123
[3]
Miller JJ, Lau AZ, Teh I, Schneider JE, Kinchesh P, Smart S, Ball V, Sibson NR,
Tyler DJ. Robust and high resolution hyperpolarized metabolic imaging of the
rat heart at 7 T with 3D spectral-spatial EPI. Magn Reson Med 2016; 75:
1515-1524