Real-time quantification of in vivo metabolism with hyperpolarized 13C magnetic resonance spectroscopic imaging (MRSI) is currently limited by partial volume effects from intense vascular signal. Flow-sensitive, bipolar gradients are an attractive option for suppressing vascular signal due to their minimal influence on static spins. This work looks at the impact of incorporating bipolar gradients on the quantification and repeatability of hyperpolarized 13C MRSI metabolic measures of lactate-to-pyruvate area-under-the-curve ratios (AUCratio). The results suggest that incorporating bipolar gradients mitigates vascular partial voluming, increasing measured AUCratio, while reducing measurement repeatability, indicated by the larger repeatability coefficients.
To evaluate test-retest repeatability of hyperpolarized 13C MRSI of renal metabolism, four, non-fasted ICR mice were assessed four times each: twice with bipolar gradients and twice without. Imaging sessions for each mouse were performed 2-4 days apart, with each mouse receiving a unique permutation of bipolar gradient state (on/off) across all sessions. All experiments complied with institutional animal care and use committee regulations.
Imaging was performed on a 4.7T small animal scanner (Agilent, Palo Alto, CA) using a dual-tuned 1H/13C volume coil (Doty Scientific, Columbia, SC). T2-weighted, 1H FSE images (resolution=0.25×0.25mm2, THK=2mm, TR/TEeff=3500/66ms) were acquired for anatomical reference. T1-weighted, 1H SPGR images (resolution=0.25×0.25mm2, THK=1mm, TR/TE=45.9/11.7ms, FA=50°) were acquired for 12 bipolar gradient settings (Gmax=90mT·m-1, lobe duration=0-3.21ms, lobe separation=0ms, m1=0-939mT·ms2·m-1) to empirically determine which settings produced the most vascular signal suppression with the least renal signal loss for each mouse (Figure 1). For 13C imaging, bipolar gradient amplitudes were scaled by γ1H/γ13C to account for increased phase accrual of 1H vs. 13C nuclei.
For metabolic imaging, 30μL aliquots of [1-13C]pyruvic acid (Cambridge Isotope Laboratories Inc., Tewksbury, MA) doped with 15mM trityl radical (Oxford Instruments, Concord, MA) were polarized (12-20%) via DNP (HyperSense, Oxford Instruments, UK) and 10μL/g, less 100μL dead volume, was injected into the tail vein. Dynamic, hyperpolarized 13C images were acquired using a single-shot, multi-echo spiral sequence (resolution=2×2mm2, THK=3mm, TR/TE/ΔTE=110/7.5-8.4/1.19ms, echoes=5, FA=10º) either with or without bipolar gradients (Gmax=360mT·m-1, lobe duration=2.95-3.13ms, m1=3286-3689mT·ms2·m-1), and interleaved with slice-selective spectra (FA=5º) at ~5s temporal resolution. To maximize acquisition efficiency, no respiratory gating was used. Spiral data was corrected for RF excitation and readout trajectory deviations4. An iterative, least-squares estimation technique5 was used to reconstruct individual metabolite images. Parametric maps of lactate-to-pyruvate area-under-the-curve ratios (AUCratio)6 were generated after removing frames exhibiting low lactate SNR with pyruvate cross-talk. Two-way repeated measures ANOVA was used to test for differences in metabolic quantification due to bipolar gradient application. Measurement repeatability was assessed using the repeatability coefficient.7
Incorporating bipolar gradients to dephase 13C vascular spins and reduce partial voluming elicited larger AUCratio measures, consistent with previous findings.2 The lack of a statistically significant difference in the left renal AUCratio may be due to right/left differences in renal perfusion8 modulating the achievable vascular suppression by bipolar gradients. Respiratory motion may also impact right and left renal motion differently, causing variable diffusion-weighting and renal signal loss due to bipolar gradients. This could result in different AUCratio measurement variability in right and left kidneys.
Incorporation of bipolar gradients also resulted in higher AUCratio repeatability coefficients. This may be attributable to improved sensitivity to metabolic shifts in perfusion and metabolism arising from different blood-glucose concentrations9, or increased sensitivity to motion due to diffusion-weighting. Additional studies are planned to investigate the impact of respiratory gating and fasting on repeatability of measurements with and without bipolar gradients applied.
1. Ardenkjaer-Larsen J, Fridlund B, Gram A, et al. Increase in signal-to-noise ratio of >10,000 times in liquid-state NMR. Proc Natl Acad Sci. 2003;100(18):10158-10163.
2. Gordon J, Niles D, Adamson E, et al. Application of flow sensitive gradients for improved measures of metabolism using hyperpolarized 13C MRI. Magn Reson Med. 2016;75(3):1242-1248.
3. Lau A, Miller J, Robson M, et al. Cardiac perfusion imaging using hyperpolarized 13C urea using flow sensitizing gradients. Magn Reson Med. 2016;75(4):1474-1483.
4. Duyn J, Yang Y, Frank J, et al. Simple correction method for k-space trajectory deviations in MRI. J Magn Reson. 1998;132(1):150-153.
5. Reeder S, Brittain J, Grist T, et al. Least-squares chemical shift separation for 13C metabolic imaging. J Magn Reson Im. 2007;26(4):1145-1152.
6. Hill D, Orton M, Mariotti E, et al. Model free approach to kinetic analysis of real-time hyperpolarized 13C magnetic resonance spectroscopy data. PLoS One. 2013;8(9):e71996.
7. Bartlett J, and Frost C. Reliability, repeatability and reproducibility: analysis of measurement errors in continuous variables. Ultrasound Obstet Gynecol. 2008;31(4):466-475.
8. Kurklinsky A and Rooke T. Nutcracker phenomenon and nutcracker syndrome. Mayo Clin Proc. 2010;85(6):552-559.
9. Serrao E, Rodrigues T, Gallagher F, et al. Effects of fasting on serial measurements of hyperpolarized [1-13C]pyruvate metabolism in tumors. NMR Biomed. 2016;29(8):1048-1055.