Xucheng Zhu1, Jeremy W Gordon2, and Peder EZ Larson1,2
1Bioengineering, UCSF, Albany, CA, United States, 2Radiology & Biomedical Imaging, University of California - San Francisco, CA, United States
Synopsis
Due
to limited transmit coil volume, refocusing pulses spoil magnetization located
around the edge of the coil even with adiabatic pulses. This is a major problem for hyperpolarized
13C MRI, as this can destroy precious magnetization. To overcome this drawback,
we propose to use a high bandwidth slice-selective adiabatic pulse to replace
the original non-selective pulse. We designed and optimized the pulse via
simulation. Then the slice-selective double spin echo sequence was tested on
both phantom and in vivo studies.
Introduction
MRI
with hyperpolarized (HP) 13C substrates is a promising method for
non-invasive imaging of metabolism. HP [1-13C] pyruvate has been used
in many cancer diagnosis studies because its metabolic conversion into lactate
and alanine has been shown to distinguish malignant and indolent tumors [1]. For
high-field studies, sampling with T2 rather than T2* may provide increased SNR
for HP metabolites that exhibit JCH coupling, such as lactate, would
ameliorate signal loss due to the increased susceptibility at high field
strength, and would be useful for simultaneous assessment of metabolism and
metabolites compartmentalization. A double spin-echo (DSE) sequence using two
identical inverse adiabatic pulses
can provide these benefits [2].
However, for many applications, the transmit coil is smaller than the
patient or object being imaged. In this fringe field region, the
refocusing pulse would act as an
excitation pulse, destroying magnetization in that region [3]. Adding a
slice-selective gradient during the adiabatic refocusing pulse would eliminate
magnetization loss at the edge of the coil, but may exacerbate slice profile
effects [4] or or may result in unwanted sensitivity to motion. In this work, we
simulated and optimized the slice selective adiabatic pulse for hyperpolarized
dynamic imaging, and tested the sequence on both phantom and in vivo studies.Methods
The slice-selective DSE
(ss-DSE) sequences in this work are shown in Figure 1. In the sequence, the slice
selective adiabatic pulse design should fulfill some specific requirements[5].
We designed a 9kHz
bandwidth pulse with 15 mm slice selective thickness. The simulation results of comparison between non-selective
and slice selective pulse sequence are shown in Figure 2. The optimized
selective pulse
transition area is much smaller than the fringe field area at
the coil edge.
For
both phantom and in vivo study, the HP samples were prepared through following
steps. 24μL aliquots of [1-13C] pyruvate (Cambridge Isotopes,
Cambridge, Massachusetts, USA) were polarized for 60 minutes in a HyperSense
polarizer (Oxford Instruments).[6] For phantom study, a syringe containing 5mL
of hyperpolarized pyruvate was placed into the scanner. Images
from the GRE and DSE acquisition were fit to a mono-exponential decay to
quantify the decay rate induced by the adiabatic DSE pulses. For
the in vivo study, three different sequences were used: a gradient-echo EPI readout, a
DSE EPI readout with frequency selective adiabatic refocusing pulses, and a DSE
EPI readout with slice-selective adiabatic refocusing pulses. Imaging started 15s
after the start of injection. Pyruvate (θ = 10°) and lactate (θ = 30°) images were
acquired every 2s, with an 8 mm thick slice and 2 x 2 mm in-plane resolution. Dynamic
pyruvate and lactate data were fit to a mono-exponential decay to quantify the
decay rate induced by the adiabatic DSE pulses.Results and Discussion
HP
pyruvate phantom study results are shown in Figure 3. According to phantom
study, we find that ss-DSE sequence perfectly excites the central area of the
phantom with a small
transition area (red circles in Fig. 3A). From the decay
rate estimation maps (Fig. 3B), we calculated the refocused area decay rate for
quantitative comparison, mean values and standard deviations were presented in
Fig. 3C. Decay rates were almost the same for each acquisition strategies.
Thus, the selective pulse only saturated the magnetization of
transition area
without affecting magnetization outside of the
transition area.
In
vivo study data are summarized in Figure 4. According to the dynamic images in
(Fig. 4A), both HP pyruvate and lactate signal intensity with ss-DSE
acquisition are higher than with non-selective DSE (ns-DSE) acquisition. We also
chose one kidney as ROI for signal change comparison among different acquisition. To quantitatively evaluate the result, we quantified the decay rate. From Fig. 4C, decay rates of both metabolites with ss-DSE
acquisition are similar to GRE acquisition. The signal intensity of C-1
pyruvate with ns-DSE decayed to almost noise level after two dynamic
acquisitions (Fig. 4A), highlighting the impact of saturation of circulating
pyruvate from the fringe field of the coil. The decay rates of both metabolites
with ss-DSE acquisition are much smaller than with ns-DSE acquisition. Due to
the narrower
transition area, ss-DSE largely reduced the magnetization
saturation effects compared with ns-DSE sequence.
Conclusion
A
dynamic double spin-echo sequence is a promising approach for imaging with T2
weighting as opposed to T2* weighting, and simultaneously assess metabolism and
metabolite compartmentalization. However, fringe field effects at the edge of
the coil will lead to unwanted magnetization loss. Slice-selective adiabatic refocusing
pulses reduce the HP magnetization saturation caused by coil geometry without introducing
other drawbacks. Acknowledgements
This work was supported by NIH grants R01EB016741 and P41EB013598. We would like to thank Dr. Robert Bok for experiment handling.References
[1]
Albers, Mark J., et al.
"Hyperpolarized 13C lactate, pyruvate, and alanine: noninvasive biomarkers
for prostate cancer detection and grading." Cancer research 68.20
(2008): 8607-8615.
[2]
Cunningham, Charles H., et al.
"Double spin-echo sequence for rapid spectroscopic imaging of
hyperpolarized 13 C." Journal of magnetic resonance 187.2 (2007):
357-362.
[3]
Josan, Sonal, et al.
"Application of double spin echo spiral chemical shift imaging to rapid
metabolic mapping of hyperpolarized [1− 13 C]-pyruvate." Journal of
Magnetic Resonance 209.2 (2011): 332-336.
[4] Gordon, Jeremy
W., et al. "Mis-estimation
and bias of hyperpolarized apparent diffusion coefficient measurements due to
slice profile effects." Magnetic Resonance in Medicine (2016).
[5]
Conolly, Steven, Dwight Nishimura,
and Albert Macovski. "A selective adiabatic spin-echo pulse." Journal
of Magnetic Resonance (1969) 83.2 (1989): 324-334.
[6]
Gordon, JW
et al., “In Vivo
Hyperpolarized 13C Diffusion Weighted MRI Measures Lactate Efflux and Changes
in MCT4 Expression in Prostate Cancer.” ISMRM (2016)