Debra E. Horng1,2, Samir D. Sharma1, Scott B. Reeder1,2,3,4,5, and Diego Hernando1,2
1Radiology, University of Wisconsin, Madison, WI, United States, 2Medical Physics, University of Wisconsin, Madison, WI, United States, 3Biomedical Engineering, University of Wisconsin, Madison, WI, United States, 4Medicine, University of Wisconsin, Madison, WI, United States, 5Emergency Medicine, University of Wisconsin, Madison, WI, United States
Synopsis
The Poisson Estimation for Ascertaining Local fields (PEAL) kernel is a recently-introduced method for background field removal in quantitative susceptibility mapping (QSM). The PEAL kernel is determined by two parameters: radius and spatial shift. The choice of these two parameters may have a substantial effect on the accuracy of background field removal. In this work, we assessed the effect of PEAL kernel size and shift on the accuracy of background field removal and susceptibility estimation.
Introduction
Quantitative susceptibility mapping (QSM) is an
emerging phase-based magnetic resonance imaging (MRI) method for quantifying
local deposition of iron in the body. Accurate assessment of the magnetic
susceptibility distribution with QSM requires unwrapping the phase data,
removing phase contributions from outside sources (“background field”), and
solving the inverse problem that estimates susceptibility from phase data.
Current background field removal methods assume that the structures of interest
are not located close to tissue-air interfaces. Recently, a new background
field technique was proposed,1 which has the potential to accurately
remove the background field in tissue regions close to an air interface. The
purpose of this work is to characterize this background field removal method in
terms of the effect of kernel size and kernel shift on the accuracy of field
map and susceptibility estimation.Theory
Poisson kernel values within a sphere of radius R at a fixed center $$$\vec{c}$$$ are described by: $$P_{R,\vec{c}}= \frac{R^2[R^4-|\vec{r}-\vec{c}|^2|\vec{c}|^2]}{V[R^4-2R^2(\vec{r}-\vec{c})\cdot\vec{c}+|\vec{r}-\vec{c}|^2|\vec{c}|^2]^{3/2}}\text{ if }|\vec{r}|<R$$ where V is the volume of the sphere. There are many possible kernels for one radius R, depending on the choice of sphere center $$$\vec{c}$$$.
Methods
To examine the accuracy of a particular kernel, we will examine both estimated local fields, and estimated susceptibility
values.
Simulation:
A digital phantom was created of a cylinder at ‑2.2ppm
(“test tube”, diameter 1.4cm, height 6.4cm) in a rectangular prism at -9.02ppm
(“water bath”, 12.8cm×12.8cm×6.4cm). The matrix size was 512×512×128 with 1.0mm isotropic resolution and the test tubes were located at both central and
edge positions. Synthetic field maps were created2-4 for both local
and background fields.
Phantom:
A susceptibility phantom comprised of a
polypropylene test tube (diameter 1.6cm, height 7cm) was created with 0, 1, 2,
3, and 4% aqueous dilutions of gadobenate dimeglumine (MultiHance, Bracco
Diagnostics, Princeton, NJ, USA). Each test tube was submerged in a rectangular
water bath (21.5cm×14.7cm×7cm). The dilutions have calculated susceptibilities
of 0.0, 1.7, 3.4, 5.2, and 6.9 ppm respectively, relative to water at 0.0 ppm,
chosen to correspond to the range of susceptibilities for liver iron overload that
are encountered clinically.5
Phantom imaging:
Imaging was performed on a 1.5T clinical MRI
system (Signa HDxt, GE Healthcare, Waukesha, WI, USA), with an 8-channel phased
array torso coil, using a multi-echo 3D spoiled gradient echo (SGRE) sequence.
Parameters: no parallel imaging, FOV=25.6 cm, slice=1.0 mm, matrix=128×128,
TE1=1.1ms, ΔTE=1.7ms, TR=11.3ms, 6 echoes/TR (flyback readout), flip angle=5°, averages=4,
BW=±62.5kHz. Each of the test tubes was scanned at two locations, central and
next to an edge. Images were reconstructed at 1.0mm isotropic resolution. The field map
was estimated using a per-voxel linear fit of unwrapped phase vs. TE.
Background field removal:
For both simulation and phantom data, field maps
were processed with PEAL kernels of radii 6 and 12, with kernel shifts from 0-5
(radius 6) and 0-10 (radius 12). The kernel shifts were chosen to be comparable
between radius sizes.
Dipole inversion:
For both simulation and phantom data, the water
bath was assumed to be 0ppm, and one constant susceptibility was estimated for
all voxels inside the test tube.
Results
Figure 2 shows the estimated local field for different kernel shifts, and how more edge values are recovered for higher kernel shifts. Figure 3 shows horizontal profiles in the estimated local field data, which shows that larger kernel shift leads to greater accuracy in horizontal profiles of the estimated local field at the edge location. As shown in Figure 4, estimated susceptibility from phantom data is fairly constant across choices of kernel radius and kernel shift, with the exception of the larger kernel (radius=12) combined with the smaller kernel shifts (0-4 voxels). As shown previously for phantom data,6 the true susceptibility is underestimated with each concentration.Discussion and Conclusion
PEAL kernels can enable accurate background field removal near the edge of an object. Kernel size and shift affect the field map values, but do not greatly affect susceptibility estimation, with the exception that larger kernel shifts should be used for larger radius kernels near the edge of an object. This exception may be the result of eroded support area with smaller shifts. Overall, PEAL performance in susceptibility estimation will also depend on the implemented dipole inversion algorithm. In conclusion, a large range of PEAL kernel sizes and shifts are viable choices, without any decrease in susceptibility accuracy. Acknowledgements
The authors wish to acknowledge support from the NIH (UL1TR00427, R01 DK083380, R01 DK088925, R01 DK100651, and K24 DK102595), as well as GE Healthcare.References
1. Horng
D et al. ISMRM 2016; 29.
2. Salomir
R et al. Concept Magnetic Res. 2003; 19B:26-34.
3. Marques
JP et al. Concept Magn Reson B. 2005; 25B:65-78.
4. Koch
KM et al. Phys Med Biol. 2006; 51:24,6381-402.
5. Schenck
JF. Med Phys 1996; 23:815-50.
6. Zhou
D et al. MRM 2016; early view.